Diagnostic capability of optical coherence tomography (Stratus OCT 3) in early glaucoma

被引:91
|
作者
Parikh, Raiul S. [1 ]
Parikh, Shefali [1 ]
Sekhar, G. Chandra [1 ]
Kumar, Rajesh S. [2 ]
Prabakaran, S. [1 ]
Babu, J. Ganesh [1 ]
Thomas, Ravi [1 ]
机构
[1] LV Prasad Eye Inst, Hyderabad 50034, Andhra Pradesh, India
[2] Singapore Eye Res Inst, Singapore, Singapore
关键词
D O I
10.1016/j.ophtha.2007.03.005
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the diagnostic capability of optical coherence tomography (Stratus OCT 3) for early glaucoma in Asian Indian eyes. Design: Cross-sectional observational study. Participants: Two groups of patients (early glaucoma and normal) who satisfied the inclusion and exclusion criteria were included. Early glaucoma was diagnosed in presence of open angles, characteristic glaucomatous optic disc changes correlating with the visual field on automated perimetry (visual field defect fulfilling at least 2 of Anderson and Patella's 3 criteria with mean deviation >= -6 dB). Normals had visual acuity >= 20/30, intraocular pressure <22 mmHg with normal optic disc and fields and no ocular abnormality. Methods: All patients underwent complete ophthalmic evaluation including visual field examination (24-2/30-2 SITA standard program) and imaging with Stratus OCT 3. Main Outcome Measures: Sensitivity, specificity, positive and negative predictive values, area under the receiving operating characteristic curve and likelihood ratios were calculated for various Stratus OCT 3 parameters. Results: Seventy-two eyes (72 patients) with early glaucoma and 96 eyes (96 normal subjects) were analyzed. The inferior maximum parameter had the best combination of sensitivity and specificity, 75% (95% confidence interval [Cl], 70.2-79.8%) and 89.6% (95% Cl, 82.6-96.6%), respectively. The 6-o'clock parameter had a sensitivity of 61.1% (95% Cl, 52.3-69.9) and specificity of 99% (95% Cl, 95-100); for an assumed prevalence of 5%, the positive and negative predictive values were 75% and 98%, respectively. The positive likelihood ratio for the 6-o'clock parameter (P<5%) in early glaucoma was 61. Conclusions: Optical coherence tomography has moderate sensitivity with high specificity for the diagnosis of early glaucoma and may have a potential role in screening.
引用
收藏
页码:2238 / 2243
页数:6
相关论文
共 50 条
  • [21] A robust algorithm for retinal thickness measurements using optical coherence tomography (Stratus OCT)
    Gregori, G
    Knighton, RW
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U50 - U50
  • [22] Diagnostic accuracy of optical coherence tomography angiography (OCTA) versus spectral domain optical coherence tomography (SD-OCT) parameters in perimetric glaucoma
    McCann, Paul M.
    Hogg, Ruth E.
    Young, Ian
    Kee, Frank
    Azuara-Blanco, Augusto
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [23] The Diagnostic Value of Optical Coherence Tomography Angiography in Glaucoma
    Koutropoulou, Nikolitsa
    Panos, Georgios D.
    [J]. CURRENT MEDICAL IMAGING, 2021, 17 (10) : 1179 - 1182
  • [24] Diagnostic Performance of Optical Coherence Tomography for Pseudoexfoliation Glaucoma
    Ozcelik-Kose, Alev
    Yildiz, Merve B.
    Imamoglu, Serhat
    [J]. JOURNAL OF GLAUCOMA, 2022, 31 (08) : 651 - 658
  • [25] Diagnostic Capability of 3D Peripapillary Retinal Volume for Glaucoma Using Optical Coherence Tomography Customized Software
    Liu, Yingna
    Jassim, Firas
    Braaf, Boy
    Khoueir, Ziad
    Poon, Linda Yi-Chieh
    Ben-David, Geulah S.
    Papadogeorgou, Georgia
    Tsikata, Edem
    Simavli, Huseyin
    Que, Christian
    Lee, Ramon
    Shieh, Eric
    Vakoc, Benjamin J.
    Bouma, Brett E.
    de Boer, Johannes F.
    Chen, Teresa C.
    [J]. JOURNAL OF GLAUCOMA, 2019, 28 (08) : 708 - 717
  • [26] Reproducibility and quality assessment of stratus optical coherence tomography (OCT3) fast macular thickness scans
    Goulding, AK
    Lambert, E
    Altaweel, MM
    Hubbard, L
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [27] Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurement
    Pierre-Kahn, V
    Tadayoni, R
    Haouchine, B
    Massin, P
    Gaudric, A
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (12) : 1581 - 1585
  • [28] Meta-analysis of Stratus OCT Glaucoma Diagnostic Accuracy
    Chen, Hsin-Yi
    Chang, Yue-Cune
    [J]. OPTOMETRY AND VISION SCIENCE, 2014, 91 (09) : 1129 - 1139
  • [29] Development and comparison of automated classifiers for glaucoma diagnosis using stratus optical coherence tomography
    Huang, ML
    Chen, HY
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (11) : 4121 - 4129
  • [30] Comparison of stratus OCT III (optical coherence tomography) and Topcon 3D Fourier Domaine OCT for retinal thickness measurements
    Forsaa, V
    Wentzel-Larsen, T.
    Seland, J.
    [J]. ACTA OPHTHALMOLOGICA, 2008, 86 : 45 - 46