Mixed Isotope Effects: Image Quality in Multimodality PET/SPECT Preclinical Imaging

被引:0
|
作者
Tainter, Kerrie H. [1 ]
Lokitz, Stephen J. [1 ]
Vascoe, Christopher [1 ]
Zhang, Joann [2 ]
Coco, Brent J. [2 ]
Iwata, Koji [2 ]
Wagenaar, Douglas J. [3 ]
Mathis, J. Michael [4 ,5 ]
机构
[1] Biomed Res Fdn NW Louisiana BRF, Shreveport, LA 71103 USA
[2] GE Healthcare, Preclin Imaging, Waukesha, WI USA
[3] GammaMedica Ideas, Northridge, CA USA
[4] Louisiana State Univ, Hlth Sci Ctr Shreveport, Dept Cellular Biol & Anat, Shreveport, LA 71105 USA
[5] Louisiana State Univ, Hlth Sci Ctr Shreveport, Gene Therapy Program, Shreveport, LA 71105 USA
关键词
D O I
10.1109/NSSMIC.2009.5401856
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The recent availability of pre-clinical PET/SPECT/CT tri-modality systems provides an opportunity to image animals injected with mixed modality (PET/SPECT) isotopes. However, before these procedures can be truly useful, some understanding of the effect of different modality radiotracers on image quality and the resulting noise characteristics must be investigated. Our goal in this study was to characterize effects on image resolution that might arise in one modality because of the presence of another (different modality) isotope. A phantom, consisting of a set of side-by-side micro-capillary tubes, were imaged on a trimodality (SPECT/PET/CT) pre-clinical scanner. Each tube was filled with a SPECT isotope and imaged while varying acquisition time, amount of activity, and distance between line sources. Images were evaluated to determine the visualization threshold at which objects became apparent in cross-sectional and MIP images (both that the objects were line sources and that the lines were two separate objects). These experiments were repeated in the presence of a PET isotope. After re-evaluating resolution, the signal to noise characteristics of the image as a whole were assessed and related to amounts of each isotope as well as total activity in the field of view. PET acquisition in the presence of SPECT isotopes was investigated using a contrast phantom with spheres filled with PET, SPECT, and mixed isotopes in backgrounds of mixed SPECT and PET isotopes. The noise properties of the background and spheres were measured as well as the visibility of the spheres. Finally, SPEC, PET, and CT images were acquired of a mouse after injection of both a PET and SPECT isotope. Acquisition of SPECT and PET images in the presence of different modality tracers (with careful considerations of the activity and acquisition parameters) seems feasible.
引用
收藏
页码:3675 / +
页数:2
相关论文
共 50 条
  • [31] Small-Animal Molecular Imaging for Preclinical Cancer Research: mu PET and mu SPECT
    Cuccurullo, Vincenzo
    Di Stasio, Giuseppe D.
    Schilliro, Maria L.
    Mansi, Luigi
    [J]. CURRENT RADIOPHARMACEUTICALS, 2016, 9 (02) : 103 - 113
  • [32] Effects of detector sensitivity on image quality in multidetector SPECT
    Weber, DA
    Ivanovic, M
    Loncaric, S
    Shelton, DK
    [J]. 1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1407 - 1410
  • [33] Effects of detector sensitivity on image quality in multidetector SPECT
    Weber, DA
    Ivanovic, M
    Loncaric, S
    Shelton, DK
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (06) : 2483 - 2488
  • [34] Leveraging deep neural networks to improve numerical and perceptual image quality in low-dose preclinical PET imaging
    Amirrashedi, Mahsa
    Sarkar, Saeed
    Mamizadeh, Hojjat
    Ghadiri, Hossein
    Ghafarian, Pardis
    Zaidi, Habib
    Ay, Mohammad Reza
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2021, 94
  • [35] Comparison and dual isotope imaging of 67Ga and 68Ga on a novel preclinical multi-modality SPECT/PET/CT/OI system
    de Swart, J.
    Segbers, M.
    Beekman, F.
    de Jong, M.
    Bernsen, M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S626 - S626
  • [36] Influence of reconstruction algorithms on image quality in SPECT myocardial perfusion imaging
    Davidsson, Anette
    Olsson, Eva
    Engvall, Jan
    Gustafsson, Agnetha
    [J]. CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2017, 37 (06) : 655 - 662
  • [37] IMAGE-QUALITY ASSESSMENT OF SEVERAL POSITRON EMITTERS FOR PRECLINICAL IMAGING
    Reilhac, A.
    Nezich, R.
    Price, R.
    Besmer, A.
    Sunn, N.
    Power, C.
    Zahra, D.
    Hamze, H.
    Boisson, F.
    Gregoire, M-C
    [J]. INTERNAL MEDICINE JOURNAL, 2014, 44 : 33 - 33
  • [38] Isotope specific resolution recovery image reconstruction in high resolution PET imaging
    Kotasidis, Fotis A.
    Angelis, Georgios I.
    Anton-Rodriguez, Jose
    Matthews, Julian C.
    Reader, Andrew J.
    Zaidi, Habib
    [J]. MEDICAL PHYSICS, 2014, 41 (05)
  • [39] Isotope Specific Resolution Modelling Image Reconstruction for High Resolution PET Imaging
    Kotasidis, Fotis A.
    Anton-Rodriguez, Jose M.
    Angelis, Georgios I.
    Matthews, Julian C.
    Reader, Andrew J.
    Zaidi, Habib
    [J]. 2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [40] Preclinical SPECT and PET: Joint EANM and ESMI procedure guideline for implementing an efficient quality control programme
    Vanhove, Christian
    Koole, Michel
    Costa, Pedro Fragoso
    Schottelius, Margret
    Mannheim, Julia
    Kuntner, Claudia
    Warnock, Geoff
    Mcdougald, Wendy
    Tavares, Adriana
    Bernsen, Monique
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, : 3822 - 3839