Golden beam data for proton pencil-beam scanning

被引:80
|
作者
Clasie, Benjamin [1 ,2 ]
Depauw, Nicolas [1 ,2 ]
Fransen, Maurice [1 ,2 ,3 ]
Goma, Carles [1 ,2 ]
Panahandeh, Hamid Reza [1 ,2 ]
Seco, Joao [1 ,2 ]
Flanz, Jacob B. [1 ,2 ]
Kooy, Hanne M. [1 ,2 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
[3] Eindhoven Univ Technol, Dept Appl Phys, NL-5600 MB Eindhoven, Netherlands
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 05期
关键词
SIMULATIONS; ALGORITHM; DESIGN; NOZZLE;
D O I
10.1088/0031-9155/57/5/1147
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Proton, as well as other ion, beams applied by electro-magnetic deflection in pencil-beam scanning (PBS) are minimally perturbed and thus can be quantified a priori by their fundamental interactions in a medium. This a priori quantification permits an optimal reduction of characterizing measurements on a particular PBS delivery system. The combination of a priori quantification and measurements will then suffice to fully describe the physical interactions necessary for treatment planning purposes. We consider, for proton beams, these interactions and derive a 'Golden' beam data set. The Golden beam data set quantifies the pristine Bragg peak depth-dose distribution in terms of primary, multiple Coulomb scatter, and secondary, nuclear scatter, components. The set reduces the required measurements on a PBS delivery system to the measurement of energy spread and initial phase space as a function of energy. The depth doses are described in absolute units of Gy(RBE) mm(2) Gp(-1), where Gp equals 10(9) (giga) protons, thus providing a direct mapping from treatment planning parameters to integrated beam current. We used these Golden beam data on our PBS delivery systems and demonstrated that they yield absolute dosimetry well within clinical tolerance.
引用
收藏
页码:1147 / 1158
页数:12
相关论文
共 50 条
  • [31] Beam Position Projection Algorithms in Proton Pencil Beam Scanning
    Nesteruk, Konrad P.
    Bradley, Stephen G.
    Kooy, Hanne M.
    Clasie, Benjamin M.
    [J]. CANCERS, 2024, 16 (11)
  • [32] Incorporation of the aperture thickness in proton pencil-beam dose calculations
    Slopsema, Roelf L.
    Kooy, Hanne M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (21): : 5441 - 5453
  • [33] Investigation on beam width tolerances for proton pencil beam scanning
    Ackermann, B.
    Brons, S.
    Ellerbrock, M.
    Jaekel, O.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S781 - S782
  • [34] Pencil-beam scanning catheter for intracoronary optical coherence tomography
    Kang, Jiqiang
    Zhu, Rui
    Sun, Yunxu
    Li, Jianan
    Wong, Kenneth K. Y.
    [J]. OPTO-ELECTRONIC ADVANCES, 2022, 5 (03)
  • [35] Flat-panel imager energy-dependent proton radiography for a proton pencil-beam scanning system
    Harms, Joseph
    Maloney, Luke
    Sohn, James J.
    Erickson, Anna
    Lin, Yuting
    Zhang, Rongxiao
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (14):
  • [36] A Novel Technique to Derive a Clinically-Acceptable Beam Model for Proton Pencil-Beam Scanning in a Commercial Treatment Planning System
    Scholey, J. E.
    Lin, L.
    Ainsley, C. G.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3527 - 3527
  • [37] Toward automatic beam angle selection for pencil-beam scanning proton liver treatments: A deep learning-based approach
    Kaderka, Robert
    Liu, Keng-Chi
    Liu, Lawrence
    VanderStraeten, Reynald
    Liu, Tyng-Luh
    Lee, Kuang-Min
    Tu, Yi-Chin Ethan
    MacEwan, Iain
    Simpson, Daniel
    Urbanic, James
    Chang, Chang
    [J]. MEDICAL PHYSICS, 2022, 49 (07) : 4293 - 4304
  • [38] A simulator of proton pencil beam scanning delivery
    Poulsen, P. R.
    Nystrom, H.
    Skyt, P. S.
    Jensen, M. F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S124 - S125
  • [39] IBA PROTON PENCIL BEAM SCANNING : VALIDATION
    Dessy, F.
    Linthout, N.
    Gillis, S.
    Closset, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S529 - S529
  • [40] Introduction to Pencil Beam Scanning Proton Therapy
    Zhu, M.
    Mossahebi, S.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E275 - E276