Comparison between gas chromatography atomic emission detection and gas chromatography mass spectrometry for the assay of propofol

被引:12
|
作者
Elbast, W
Guitton, J
Desage, M
Deruaz, D
Manchon, M
Brazier, JL
机构
[1] INST SCI PHARMACEUT & BIOL,LAB ETUD ANAL & CINET MEDICAMENT,F-69373 LYON 08,FRANCE
[2] CTR HOSP LYON SUD,SERV BIOCH TOXICOL,F-69495 PIERRE BENITE,FRANCE
来源
关键词
propofol;
D O I
10.1016/S0378-4347(96)00178-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Quantitation by gas chromatography-atomic emission detection (GC-AED) is based on the intensity of the signal measured at a wavelength characteristic of an element, after atomisation by the plasma. This response depends only on the number of atoms of this element present in the molecule under investigation, and is independent of the structure of the molecule. This technique was used for the assay of propofol, and the estimation of its two metabolites, after calibration with standard solutions of pure propofol. The results were compared with those obtained by gas chromatography-mass spectrometry (GC-MS). Propofol was quantified with higher precision and accuracy by GC-AED than by GC-MS which exhibited larger residual values. Concentration assessment for two metabolites showed a better agreement with the theoretical value by GC-AED since the response depends only on the number of carbon atoms in each molecule.
引用
收藏
页码:97 / 102
页数:6