Design and implementation of a system for treating paediatric patients with stereotactically-guided conformal radiotherapy

被引:6
|
作者
Adams, EJ
Suter, BL
Warrington, AP
Black, P
Saran, F
Brada, M
机构
[1] Royal Marsden NHS Trust, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHS Trust, Dept Radiotherapy, Sutton SM2 5PT, Surrey, England
[3] Royal Marsden NHS Trust, Neurooncol Unit, Sutton SM2 5PT, Surrey, England
关键词
fractionated stereotactic radiotherapy; immobilisation; paediatric radiotherapy; patient fixation; conformal radiotherapy;
D O I
10.1016/S0167-8140(01)00383-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Stereotactically-guided conformal radiotherapy (SCRT) allows the delivery of highly conformal dose distributions to localised brain tumours. This is of particular importance for children, whose often excellent long-term prognosis should be accompanied by low toxicity. The commercial immobilisation system in use at our hospital for adults was felt to be too heavy for children, and precluded the use of anaesthesia, which is sometimes required for paediatric patients. This paper therefore describes the design and implementation of a system for treating children with SCRT. This system needed to be well tolerated by patients, with good access for treating typical childhood malignancies. Materials and methods: A lightweight frame was developed for immobilisation, with a shell-based alternative for patients requiring general anaesthetic. Procedures were set up to introduce the patients to the frame system in order to maximise patient co-operation and comfort. Film measurements were made to assess the impact of the frame on transmission and surface dose. The reproducibility of the systems was assessed using electronic portal images. Results: Both frame and shell systems are in clinical use. The frame weighs 0.6 kg and is well tolerated. It has a transmission of 92-96%, and fields which pass through it deliver surface doses of 58-82% of the dose at d(max), compared to 18% when no frame is present. However, the frame is constructed to maximise the availability of unobstructed beam directions. Reproducibility measurements for the frame showed a mean random error of 1.0 +/- 0.2 mm in two dimensions (2D) and 1.4 +/- 0.7 mm in 3D. The mean systematic error in 3D was 2.2 mm, and 90% of all overall 3D errors were less than 3.4 mm. For the shell system, the mean 2D random error was 1.5 +/- 0.2 mm. Conclusions: Two well-tolerated immobilisation devices have been developed for fractionated SORT treatment of paediatric patients. A lightweight frame system gives a wide range of possible unobstructed beam directions, although beams that intersect the frame are not precluded, provided that output corrections are applied. A shell system allows the use of general anaesthesia. Both systems give reproducible immobilisation to complement the high-precision treatment delivery. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:289 / 297
页数:9
相关论文
共 50 条
  • [31] Conceptual Design of PET-linac System for Molecular-guided Radiotherapy
    Ishikawa, M.
    Yamaguchi, S.
    Tanabe, S.
    Bengua, G.
    Sutherland, K.
    Suzuki, R.
    Miyamoto, N.
    Nishijima, K.
    Katoh, N.
    Shirato, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S674 - S674
  • [32] Design and Implementation of Structured Clinical Documentation Support Tools for Treating Stroke Patients
    Simon, Kelly Claire
    Munson, Richard
    Ong, Archie
    Gil, Fulvio R.
    Campanella, Franco
    Hillman, Laura
    Lai, Rebekah
    Chesis, Richard
    Tideman, Samuel
    Vazquez, Rosa Maria
    Meyers, Steven
    Frigerio, Roberta
    Maraganore, Demetrius
    JOURNAL OF STROKE & CEREBROVASCULAR DISEASES, 2019, 28 (05): : 1229 - 1235
  • [33] Applied to the shell conformal semi-active laser-guided optical system design
    Ge Mingda
    Wei Qun
    2015 INTERNATIONAL CONFERENCE ON OPTOELECTRONICS AND MICROELECTRONICS (ICOM), 2015, : 485 - 487
  • [34] A RANDOMIZED STUDY TO EVALUATE THE ACUTE TOXICITY IN PATIENTS OF CERVICAL CANCER TREATED WITH THREE-DIMENSIONAL CONFORMAL RADIOTHERAPY VERSUS IMAGE GUIDED RADIOTHERAPY
    Rai, B.
    Singh, M.
    Saha, S.
    Oinam, A.
    Ghoshal, S.
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2018, 28 : 74 - 74
  • [35] Design and implementation of a computer vision-guided greenhouse crop diagnostics system
    Story, David
    Kacira, Murat
    MACHINE VISION AND APPLICATIONS, 2015, 26 (04) : 495 - 506
  • [36] Design and implementation of a PC-based image-guided surgical system
    Stefansic, JD
    Bass, WA
    Hartmann, SL
    Beasley, RA
    Sinha, TK
    Cash, DM
    Herline, AJ
    Galloway, RL
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2002, 69 (03) : 211 - 224
  • [37] Reference Model Guided System Design and Implementation for Interoperable Environmental Research Infrastructures
    Zhao, Zhiming
    Martin, Paul
    Grosso, Paola
    Los, Wouter
    de laat, Cees
    Jeffrey, Keith
    Hardisty, Alex
    Vermeulen, Alex
    Castelli, Donatella
    Legre, Yannick
    Kutsch, Werner
    2015 IEEE 11TH INTERNATIONAL CONFERENCE ON E-SCIENCE, 2015, : 551 - 556
  • [38] Design and implementation of time reversal focusing inspection system for guided wave in pipeline
    He, Cunfu
    Zhou, Jinjie
    Wu, Bin
    Zheng, Yang
    Fu, Hao
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2012, 33 (02): : 342 - 348
  • [39] Design and implementation of a computer vision-guided greenhouse crop diagnostics system
    David Story
    Murat Kacira
    Machine Vision and Applications, 2015, 26 : 495 - 506
  • [40] DEVELOPMENT OF A MICRO-COMPUTED TOMOGRAPHY-BASED IMAGE-GUIDED CONFORMAL RADIOTHERAPY SYSTEM FOR SMALL ANIMALS
    Zhou, Hu
    Rodriguez, Manuel
    van den Haak, Fred
    Nelson, Geoffrey
    Jogani, Rahil
    Xu, Jiali
    Zhu, Xinzhi
    Xian, Yongjiang
    Tran, Phuoc T.
    Felsher, Dean W.
    Keall, Paul J.
    Graves, Edward E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (01): : 297 - 305