Comparison of different near-infrared spectroscopic cerebral oxygenation indices with central venous and jugular venous oxygenation saturation in children

被引:86
|
作者
Nagdyman, Nicole [1 ]
Ewert, Peter [1 ]
Peters, Bjoern [1 ]
Miera, Oliver [1 ]
Fleck, Thilo [1 ]
Berger, Felix [2 ]
机构
[1] Deutsch Herzzentrum Berlin, Dept Congenital Heart Dis Pediat Cardiol, D-13353 Berlin, Germany
[2] Charite Univ Med Berlin, Dept Pediat Cardiol, Berlin, Germany
关键词
near-infrared spectroscopy; tissue oxygenation index; regional cerebral oxygenation; jugular venous oxygen saturation; central venous oxygen saturation; congenital heart defects;
D O I
10.1111/j.1460-9592.2007.02365.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: We compared two different near-infrared spectrophotometers: cerebral tissue oxygenation index (TOI) measured by NIRO 200 and regional cerebral oxygenation index (rSO(2)) measured by INVOS 5100 with venous oxygen saturation in the jugular bulb (SjO(2)) and central SvO(2) from the superior caval vein (SVC) during elective cardiac catheterization in children. Methods: A prospective observational clinical study in 31 children with congenital heart defects in a catheterization laboratory was undertaken. TOI was compared with SjO(2) in the left jugular bulb and with SvO(2). rSO(2) was compared with SjO(2) from the right jugular bulb and SvO(2). Linear regression analysis and Pearson's correlation coefficient were calculated and Bland-Altman analyses were performed. Results: Cerebral TOI and SjO(2) were significantly correlated (r = 0.56, P < 0.0001), as well as TOI and SvO(2) with r = 0.74 (P < 0.0001). Bland-Altman plots showed a mean bias of -4.3% with limits of agreement of 15.7% and -24.3% for TOI and SjO(2) and a mean bias of -4.9% with limits of agreement of 10.3% and -20.1% for TOI and SvO(2). Cerebral rSO(2) and SjO(2) showed a significant correlation (r = 0.83, P < 0.0001) and rSO(2) and SvO(2) showed excellent correlation with r = 0.93 (P < 0.0001). Bland-Altman plots showed a mean bias of -5.2% with limits of agreement of between 8.4% and -18.8% for rSO(2) and SjO(2) and a mean bias of 5.6% with limits of agreement of 13.4% and -2.2% for rSO(2) and SvO(2). Conclusions: Both near-infrared spectroscopy devices demonstrate a significant correlation with SjO(2) and SvO(2) values; nevertheless both devices demonstrate a substantial bias of the measurements to both SjO(2) and SvO(2).
引用
收藏
页码:160 / 166
页数:7
相关论文
共 50 条
  • [1] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Nagdyman, N
    Fleck, T
    Schubert, S
    Ewert, P
    Peters, B
    Lange, PE
    Abdul-Khaliq, H
    [J]. INTENSIVE CARE MEDICINE, 2005, 31 (06) : 846 - 850
  • [2] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Fleck, TPK
    Nagdyman, N
    Schubert, S
    Peters, B
    Reisenkampff, E
    Ewert, P
    Abdul-Khaliq, H
    Lange, PE
    [J]. PEDIATRIC RESEARCH, 2004, 56 (03) : 478 - 478
  • [3] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Nicole Nagdyman
    Thilo Fleck
    Stephan Schubert
    Peter Ewert
    Björn Peters
    Peter E. Lange
    Hashim Abdul-Khaliq
    [J]. Intensive Care Medicine, 2005, 31 : 846 - 850
  • [4] Near-infrared spectroscopic cerebral oxygenation reading in neonates and infants is associated with central venous oxygen saturation
    Weiss, M
    Dullenkopf, A
    Kolarova, A
    Schulz, G
    Frey, B
    Baenziger, O
    [J]. PEDIATRIC ANESTHESIA, 2005, 15 (02) : 102 - 109
  • [5] JUGULAR VENOUS OXYGEN SATURATION AS AN INDEX OF CEREBRAL OXYGENATION
    LARSON, CP
    EHRENFELD, WK
    WADE, JG
    WYLIE, EJ
    [J]. SURGERY, 1967, 62 (01) : 31 - +
  • [6] 86 Comparison Between Cerebral Tissue Oxygenation Index Measured by Near- Infrared Spectroscopy and Venous Jugular Bulb Saturation in Children
    T P K Fleck
    N Nagdyman
    S Schubert
    B Peters
    E Riesenkampff
    P Ewert
    H Abdul-Khaliq
    P E Lange Deutsches
    [J]. Pediatric Research, 2004, 56 (3) : 478 - 478
  • [7] MEASUREMENT OF CEREBRAL VENOUS OXYHEMOGLOBIN SATURATION IN CHILDREN BY NEAR-INFRARED SPECTROSCOPY AND PARTIAL JUGULAR VENOUS OCCLUSION
    YOXALL, CW
    WEINDLING, AM
    DAWANI, NH
    PEART, I
    [J]. PEDIATRIC RESEARCH, 1995, 38 (03) : 319 - 323
  • [8] Validation of cerebral venous oxygenation measured using near-infrared spectroscopy and partial jugular venous occlusion in the newborn lamb
    Wong, Flora Y.
    Barfield, Charles P.
    Campbell, Louisa
    Brodecky, Vojta A.
    Walker, Adrian M.
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2008, 28 (01): : 74 - 80
  • [9] Comparison of regional transcranial oximetry with near infrared spectroscopy (NIRS) and jugular venous bulb oxygen saturation for the monitoring of cerebral oxygenation in infants and children
    Abdul-Khaliq, H
    Troitzsch, D
    Berger, F
    Lange, PE
    [J]. BIOMEDIZINISCHE TECHNIK, 2000, 45 (11): : 328 - 332
  • [10] Transcutaneously measured near-infrared spectroscopic liver tissue oxygenation does not correlate with hepatic venous oxygenation in children
    Weiss, M
    Schulz, G
    Fasnacht, M
    Balmer, C
    Fischer, JE
    Gerber, AC
    Bucher, HU
    Baenziger, O
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2002, 49 (08): : 824 - 829