Comparison of regional transcranial oximetry with near infrared spectroscopy (NIRS) and jugular venous bulb oxygen saturation for the monitoring of cerebral oxygenation in infants and children

被引:79
|
作者
Abdul-Khaliq, H [1 ]
Troitzsch, D [1 ]
Berger, F [1 ]
Lange, PE [1 ]
机构
[1] Deutsch Herzzentrum Berlin, Abt Angeborene Herzfehler Kinderkardiol, D-13353 Berlin, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2000年 / 45卷 / 11期
关键词
cerebral oxygenation; near infrared spectroscopy (NIRS); regional cerebral haemoglobin saturation (rSO(2)); jugular venous oxygen saturation (SjvO(2)); infants and children;
D O I
10.1515/bmte.2000.45.11.328
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Using a dual channel near infrared (NIR) in vivo optical spectroscopy (INVOS(R)) system (INVOS 3100A(R), Somanetics Corp. Troy, MI, USA) we investigated the relationship between jugular venous oxygen saturation (SjvO(2)) and regional cerebral oxygen saturation (rSO(2)) in 30 infants and children (mean age 4.5 years) with congenital heart disease undergoing cardiac catheterisation. The NIRS-SomaSensor(R) (emitter and dual receiver probe) was applied at a standardised right fronto-temporal location lover the right frontal cortex) on the infant's head and covered with an adhesive flexible bandage Using NIR Light (730 and 810 nm) and two source-detector spacings (3 and 4 cm from the transmitter), percentage values of rSO(2) were calculated from detected haemoglobin saturations. Simultaneously, jugular venous oxygen saturation (SjvO(2)) monitoring was performed via a catheter placed in the right internal jugular vein with its tip positioned in the jugular bulb, as verified by fluoroscopy. To compare the reliability of NIRS measurement characteristics, jugular venous blood was analysed for SjvO(2) as a reference measure of global cerebral oxygenation, by co-oximetry (OSM3-Hemoximeter(R), Radiometer Copenhagen, Denmark). Other measured variables included pulse oximetry, arterial blood pressure, and venous and arterial oxygen saturations. Over a jugular venous oxygen saturation range of 31-83%, a significant positive linear correlation was found between rSO(2) (NIRS measurement) and SjvO(2) (jugular bulb oximetry) (r = 0.93, p < 0.001). No significant correlation was observed between rSO(2) values and arterial blood saturation or pulse oximetry. The quantitative correlation between rSO(2) (haemoglobin oxygenation in a small hemi-elliptical area of the brain) and reference SjvO(2) measurement (method for monitoring global cerebral oxygenation) suggests that NIRS measurement with subtraction algorithm should identify predominantly intracranial saturation in the pediatric age group, and will tend to reflect global oxygenation under physiological conditions. Transcranial oximetry using dual receiving channel NIRS offers a noninvasive, real-time, reliable and practicable means of monitoring cerebral haemoglobin oxygenation changes infants and children with cyanotic and noncyanotic congenital heart disease.
引用
收藏
页码:328 / 332
页数:5
相关论文
共 50 条
  • [1] Cerebral oxygenation measured by near-infrared spectroscopy: Comparison with jugular bulb oximetry
    Daubeney, PEF
    Pilkington, SN
    Janke, E
    Charlton, GA
    Smith, DC
    Webber, SA
    [J]. ANNALS OF THORACIC SURGERY, 1996, 61 (03): : 930 - 934
  • [2] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Nagdyman, N
    Fleck, T
    Schubert, S
    Ewert, P
    Peters, B
    Lange, PE
    Abdul-Khaliq, H
    [J]. INTENSIVE CARE MEDICINE, 2005, 31 (06) : 846 - 850
  • [3] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Fleck, TPK
    Nagdyman, N
    Schubert, S
    Peters, B
    Reisenkampff, E
    Ewert, P
    Abdul-Khaliq, H
    Lange, PE
    [J]. PEDIATRIC RESEARCH, 2004, 56 (03) : 478 - 478
  • [4] Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
    Nicole Nagdyman
    Thilo Fleck
    Stephan Schubert
    Peter Ewert
    Björn Peters
    Peter E. Lange
    Hashim Abdul-Khaliq
    [J]. Intensive Care Medicine, 2005, 31 : 846 - 850
  • [5] MONITORING OF THE CEREBRAL OXYGENATION WITH NEAR-INFRARED SPECTROSCOPY IN COMPARISON TO JUGULAR BULB OXYGEN-SATURATION DURING CAROTID ENDARTERECTOMY
    SCHINDLER, E
    ZICKMANN, B
    MULLER, M
    BOLDT, J
    KROLL, J
    HEMPELMANN, G
    [J]. VASA-JOURNAL OF VASCULAR DISEASES, 1995, 24 (02): : 168 - 175
  • [6] Jugular venous bulb oxygen saturation compared with cerebral saturation measured by near infrared spectroscopy.
    OwenReece, H
    Smith, M
    Elwell, CE
    Lovell, AT
    Goldstone, J
    [J]. ANESTHESIA AND ANALGESIA, 1997, 84 : S370 - S370
  • [7] 86 Comparison Between Cerebral Tissue Oxygenation Index Measured by Near- Infrared Spectroscopy and Venous Jugular Bulb Saturation in Children
    T P K Fleck
    N Nagdyman
    S Schubert
    B Peters
    E Riesenkampff
    P Ewert
    H Abdul-Khaliq
    P E Lange Deutsches
    [J]. Pediatric Research, 2004, 56 (3) : 478 - 478
  • [8] Cerebral oxygenation measured by near-infrared spectroscopy: Comparison with jugular bulb oximetry - Invited commentary
    Gunter, JB
    [J]. ANNALS OF THORACIC SURGERY, 1996, 61 (03): : 934 - 934
  • [9] Cerebral oxygenation monitoring by near-infrared spectroscopy is not clinically useful in patients with severe closed-head injury: A comparison with jugular venous bulb oximetry
    Lewis, SB
    Myburgh, JA
    Thornton, EL
    Reilly, PL
    [J]. CRITICAL CARE MEDICINE, 1996, 24 (08) : 1334 - 1338
  • [10] Regional cerebral oxygenation by NIRS does not correlate with central or jugular venous oxygen saturation during interventional catheterisation in children
    Knirsch, W.
    Stutz, K.
    Kretschmar, O.
    Tomaske, M.
    Balmer, C.
    Schmitz, A.
    Berger, F.
    Bauersfeld, U.
    Weiss, M.
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2008, 52 (10) : 1370 - 1374