Development of a New Modified Screen-Printed and Carbon Paste Electrodes for Selective Determination of Cetyltrimethylammonium Bromide in Different Water Samples

被引:0
|
作者
Ali, Tamer Awad [1 ]
Soliman, M. H. [2 ]
Mohamed, Gehad G. [3 ]
Farag, A. B. [2 ]
Samah, Mahmoud K. [3 ]
机构
[1] EPRI, Cairo 11727, Egypt
[2] Helwan Univ, Fac Sci, Dept Chem, Cairo, Egypt
[3] Cairo Univ, Fac Sci, Dept Chem, Giza 12613, Egypt
来源
关键词
Cetyltrimethylammonium bromide; Modified screen-printed; Modified carbon paste; Zeolite ionophore; Selectivity coefficient; Different water samples; ENVIRONMENTAL POLLUTED SAMPLES; POTENTIOMETRIC DETERMINATION; CATIONIC SURFACTANT; GOLD-NANOPARTICLES; ZEOLITE FRAMEWORKS; ION-EXCHANGER; MEMBRANE; TITRATION; CONSTRUCTION; PERFORMANCE;
D O I
暂无
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Cetyltrimethylammonium bromide (CTAB)-selective electrodes of the types carbon paste (CPE) and screen-printed sensors (SPE) based on incorporation of zeolite ionophore have been constructed. The influences of paste composition, temperature, pH of the test solution, and foreign ions on the electrodes performance were investigated. The electrodes showed Nernstian behavior with linear concentration range of 4.61 x 10(-7) -1.0 x 10(-2) and 1.26 x 10(-7) -1.0 x 10(-2) mol L-1, slope of 57.56 +/- 1.25 and 58.92 +/- 0.27 mV decade(-1) and lower limit of detection of 4.61 x 10(-7) and 1.26 x 10(-7) mol L-1 for modified CPE and SPE sensors, respectively. The electrodes display good selectivity for CTAB with respect to a number of common foreign inorganic and organic species. The response is not affected by pH variation between 2.0 - 8.5 and 2.0 - 9.0 for modified CPE and SPE, respectively. The sensors were successfully used for determination of CTAB both in pure solution and in different spiked real water samples. The frequently used CTAB of analytical and technical grade as well as different water samples has been successfully titrated and the results obtained agreed with those obtained with standard two phase titration method. The sensitivity of the proposed method is comparable with the official method and ability of field measurements.
引用
收藏
页码:3192 / 3206
页数:15
相关论文
共 50 条
  • [1] Modified Screen-Printed and Carbon Paste as Ion-Selective Electrodes for the Determination of Ramipril Drug in Pharmaceutical and Biological Samples
    Mohamed, Gehad G.
    Frag, Eman Yossri
    El Mohamed, Marwa
    Elhassan, Mohamed O.
    [J]. ANALYTICAL CHEMISTRY LETTERS, 2019, 9 (03) : 311 - 328
  • [2] New Screen-Printed Ion-Selective Electrodes for Potentiometric Titration of Cetyltrimethylammonium Bromide in Different Civilic Media
    Mohamed, Gehad G.
    Ali, Tamer Awad
    El-Shahat, M. F.
    Al-Sabagh, A. M.
    Migahed, M. A.
    [J]. ELECTROANALYSIS, 2010, 22 (21) : 2587 - 2599
  • [3] Modified Screen-Printed Ion Selective Electrodes for Potentiometric Determination of Sodium Dodecylsulfate in Different Samples
    Ali, Tamer Awad
    Mohamed, Gehad G.
    [J]. JOURNAL OF AOAC INTERNATIONAL, 2015, 98 (01) : 116 - 123
  • [5] A Novel Screen-Printed and Carbon Paste Electrodes for Potentiometric Determination of Uranyl(II) Ion in Spiked Water Samples
    Tamer Awad Ali
    Gehad G. Mohamed
    Refat F. Aglan
    Mai A. Mourad
    [J]. Russian Journal of Electrochemistry, 2018, 54 : 201 - 215
  • [6] A Novel Screen-Printed and Carbon Paste Electrodes for Potentiometric Determination of Uranyl(II) Ion in Spiked Water Samples
    Ali, Tamer Awad
    Mohamed, Gehad G.
    Aglan, Refat F.
    Mourad, Mai A.
    [J]. RUSSIAN JOURNAL OF ELECTROCHEMISTRY, 2018, 54 (02) : 201 - 215
  • [7] New Chemically Modified Screen-Printed Electrode for Co(II) Determination in Different Water Samples
    Ali, Tamer Awad
    Aglan, Refat F.
    Mohamed, Gehad G.
    Mourad, Mai A.
    [J]. INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2014, 9 (04): : 1812 - 1826
  • [8] A new screen-printed electrode for selective determination of bismuth in different authentic samples
    Refaat F. Aglan
    Mostafa M. Hamed
    Hosam M. Saleh
    [J]. Journal of the Iranian Chemical Society, 2023, 20 : 1481 - 1490
  • [9] A new screen-printed electrode for selective determination of bismuth in different authentic samples
    Aglan, Refaat F.
    Hamed, Mostafa M.
    Saleh, Hosam M.
    [J]. JOURNAL OF THE IRANIAN CHEMICAL SOCIETY, 2023, 20 (06) : 1481 - 1490
  • [10] Novel screen-printed electrode for the determination of dodecyltrimethylammonium bromide in water samples
    Mohamed, Gehad G.
    Ali, Tamer Awad
    El-Shahat, M. F.
    Migahed, M. A.
    Al-Sabagh, A. M.
    [J]. DRUG TESTING AND ANALYSIS, 2012, 4 (12) : 1009 - 1013