Dosimetric evaluation of the INTRABEAM system for breast intraoperative radiotherapy: A single-institution experience

被引:2
|
作者
Shaikh, Mubin Y. [1 ]
Nalichowski, Adrian [2 ]
Joiner, Michael C. [3 ]
Burmeister, Jay [2 ]
机构
[1] Rochester Reg, Dept Radiat Oncol, 1425 Portland Ave, Rochester, NY 14621 USA
[2] Wayne State Univ, Gershenson Radiat Oncol Ctr, Barbara Ann Karmanos Canc Inst, Dept Oncol,Sch Med, Detroit, MI USA
[3] Wayne State Univ, Gershenson Radiat Oncol Ctr, Dept Oncol, Detroit, MI 48201 USA
关键词
INTRABEAM; IORT; Dose-volume histograms; Breast IORT; Intraoperative dosimetry; RADIOSURGERY; MINIATURE; RADIATION; CANCER; TARGIT; TRIAL;
D O I
10.1016/j.meddos.2019.09.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Breast intraoperative radiotherapy (IORT) with the INTRABEAM system uses a 50 kV x-ray source to deliver a single fraction of radiation therapy to the lumpectomy cavity during breast-conserving surgery. We seek to perform a dosimetric analysis of the lumpectomy cavity for rigid spherical applicators. Water phantom measurements were acquired to validate the vendor-provided x-ray calibration. The planning target volume (PTV) was defined as a 10 mm expansion beyond the spherical applicator, a dose-volume histogram (DVH) was generated and dose-volume parameters [D-min, D-1mm, V-90, V-80, V-50, HI] were reported. Additionally, the therapeutic treatment depth using the 90 and 80% isodose level was computed [R-90, R-80]. When the percent depth dose (PDD) is normalized to the surface of the applicator, smaller applicators have a steeper PDD. For a prescription dose of 20 Gy to the surface of the applicator, the range of dose-volume parameters for the PTV was: 3.15 to 6.84 Gy for D-min, 16.2 to 17.6 Gy for D-1mm, 2.6 to 6.9% for V-90, 5.5 to 15.1% for V-80, and 21.1 to 55.6% for V-50. For applicators 15 to 50 mm in diameter, the reported values were: 6.35 to 2.9 for HI, 0.53 to 0.85 mm for R-90, and 1.18 to 1.85 mm for R-80. Smaller applicators have reduced PTV coverage but elevated HI because the attenuation of the beam proximal to the source is more pronounced. Additionally, the presence of the aluminum filter for small applicators (<= 30 mm) increases PTV coverage but reduces the dose rate on the applicator surface. The delivery of IORT is performed in the OR without the use of image-based planning. To overcome this limitation, we have generated sample DVH's and report dosimetric parameters to offer clinicians a unique dosimetric perspective. (c) 2019 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:E1 / E6
页数:6
相关论文
共 50 条
  • [31] Metaplastic breast cancer: A single-institution experience.
    Silva, Aumilto Augusto
    Linck, Rudinei Diogo Marques
    Lima, Julianne Maria da Silva
    Matutino, Adriana Reis Brandao
    Silva, Saulo Brito
    Vicentini, Maria Fernanda
    Felizola, Marcelo
    Gagliato, Debora De Melo
    Hoff, Paulo Marcelo
    Mano, Max S.
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2016, 34 (15)
  • [32] Revision Surgery for Breast Cancer Single-Institution Experience
    Thorat, Mangesh A.
    Rangole, Ashvin
    Nadkarni, Mandar S.
    Parmar, Vani
    Badwe, Rajendra A.
    [J]. CANCER, 2008, 113 (08) : 2347 - 2352
  • [33] Hypofractionated stereotactic radiotherapy for oligometastases in the brain: a single-institution experience
    Marcello Marchetti
    Ida Milanesi
    Chiara Falcone
    Michela De Santis
    Luisa Fumagalli
    Lorenzo Brait
    Livia Bianchi
    Laura Fariselli
    [J]. Neurological Sciences, 2011, 32 : 393 - 399
  • [34] Routine intraoperative neuromonitoring during thyroidectomy: a UK single-institution experience
    Cvasciuc, Titus
    Lansdown, Mark
    Fraser, Sheila M.
    [J]. MINERVA ENDOCRINOLOGICA, 2019, 44 (02) : 236 - 238
  • [35] Full-dose intraoperative radiotherapy with electrons in breast cancer - First report on late toxicity and cosmetic results from a single-institution experience
    Mussari, Salvatore
    della Sala, Walter Sabino
    Busana, Lucia
    Vanoni, Valentina
    Eccher, Claudio
    Zani, Bruno
    Menegotti, Loris
    Tomio, Luigi
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2006, 182 (10) : 589 - +
  • [36] COMBINATION OF INTRAOPERATIVE RADIOTHERAPY (IORT) BY USING INTRABEAM SYSTEM AND ONCOPLASTIC BREAST SURGERY IN VIETNAM
    Tung Dinh Nguyen
    [J]. ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2014, 10 : 170 - 170
  • [37] Radiosurgery and stereotactic radiotherapy for brain metastases: an early single-institution experience
    Borzillo, V.
    Giugliano, F. M.
    Di Franco, R.
    Falivene, S.
    Iadanza, L.
    Mormile, M.
    Ravo, V.
    Muto, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S4 - S5
  • [38] Palliative radiotherapy for Japanese patients with malignant melanoma: a single-institution experience
    Koiwai, Keiichiro
    Sasaki, Shigeru
    Yoshizawa, Eriko
    Ina, Hironobu
    Fukazawa, Ayumu
    Sakai, Katsuya
    Ozawa, Takesumi
    Matsushita, Hirohide
    Kadoya, Masumi
    [J]. JAPANESE JOURNAL OF RADIOLOGY, 2016, 34 (02) : 125 - 129
  • [39] Hypofractionated Radiotherapy for 35 Patients with Adrenal Metastases: A Single-Institution Experience
    Zhao, Ruizhi
    Ma, Yuchao
    Yang, Siran
    Liu, Qingfeng
    Tang, Yuan
    Wang, Kai
    Zhang, Ye
    Bi, Nan
    Zhang, Hongmei
    Yi, Junlin
    Li, Yexiong
    Luo, Jingwei
    Xiao, Jianping
    [J]. CANCER MANAGEMENT AND RESEARCH, 2020, 12 : 11563 - 11571
  • [40] Palliative radiotherapy for Japanese patients with malignant melanoma: a single-institution experience
    Keiichiro Koiwai
    Shigeru Sasaki
    Eriko Yoshizawa
    Hironobu Ina
    Ayumu Fukazawa
    Katsuya Sakai
    Takesumi Ozawa
    Hirohide Matsushita
    Masumi Kadoya
    [J]. Japanese Journal of Radiology, 2016, 34 : 125 - 129