A novel paper-based assay for the simultaneous determination of Rh typing and forward and reverse ABO blood groups

被引:48
|
作者
Noiphung, Julaluk [1 ]
Talalak, Kwanrutai [1 ]
Hongwarittorrn, Inn [1 ]
Pupinyo, Naricha [2 ]
Thirabowonkitphithan, Pannawich [2 ]
Laiwattanapaisal, Wanida [3 ]
机构
[1] Chulalongkorn Univ, Fac Allied Hlth Sci, Grad Program Clin Biochem & Mol Med, Bangkok 10330, Thailand
[2] Chulalongkorn Univ, Fac Allied Hlth Sci, Undergrad Program Med Technol, Bangkok 10330, Thailand
[3] Chulalongkorn Univ, Fac Allied Hlth Sci, Dept Clin Chem, Bangkok 10330, Thailand
来源
关键词
ABO blood group; Forward and reverse typing; Haemagglutination; Paper-based analytical devices (PADs); Rh typing; Wax dipping; LOW-COST; DEVICE; WAX; MICROFLUIDICS; SYSTEM;
D O I
10.1016/j.bios.2014.09.011
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We propose a new, paper-based analytical device (PAD) for blood typing that allows for the simultaneous determination of ABO and Rh blood groups on the same device. The device was successfully fabricated by using a combination of wax printing and wax dipping methods. A 1:2 blood dilution was used for forward grouping, whereas whole blood could be used for reverse grouping. A 30% cell suspension of A-cells or B-cells was used for haemagglutination on the reverse grouping side. The total assay time was 10 min. The ratio between the distance of red blood cell movement and plasma separation is the criterion for agglutination and indicates the presence of the corresponding antigen or antibody. The proposed PAD has excellent reproducibility in that the same blood groups, namely A, AB, and 0, were reported by using different PADs that were fabricated on the same day (n=10). The accuracy for detecting blood group A (n=12), B (n=13), AB (n=9), 0 (n=14), and Rh (n=48) typing were 92%, 85%, 89%, 93%, and 96%, respectively, in comparison with the conventional slide test method. The haematocrit of the sample affects the accuracy of the results, and appropriate dilution is suggested before typing. In conclusion, this study proposes a novel method that is straightforward, time-saving, and inexpensive for the simultaneous determination of ABO and Rh blood groups, which is promising for use in developing countries. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:485 / 489
页数:5
相关论文
共 50 条
  • [1] Simultaneous forward and reverse ABO blood group typing using a paper-based device and barcode-like interpretation
    Songjaroen, Temsiri
    Laiwattanapaisal, Wanida
    [J]. ANALYTICA CHIMICA ACTA, 2016, 921 : 67 - 76
  • [2] Validation of Paper-Based Assay for Rapid Blood Typing
    Al-Tamimi, Mohammad
    Shen, Wei
    Zeineddine, Rania
    Huy Tran
    Garnier, Gil
    [J]. ANALYTICAL CHEMISTRY, 2012, 84 (03) : 1661 - 1668
  • [3] Forward and reverse typing discrepancy and crossmatch incompatibility of ABO blood groups: cause analysis and treatment
    Qiu, Hongmei
    Wang, Xuechun
    Shao, Yan
    [J]. HEMATOLOGY, 2023, 28 (01)
  • [4] Paper-based device for rapid typing of secondary human blood groups
    Li, Miaosi
    Then, Whui Lyn
    Li, Lizi
    Shen, Wei
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (03) : 669 - 677
  • [5] Paper-based device for rapid typing of secondary human blood groups
    Miaosi Li
    Whui Lyn Then
    Lizi Li
    Wei Shen
    [J]. Analytical and Bioanalytical Chemistry, 2014, 406 : 669 - 677
  • [6] Enhanced fully cellulose based forward and reverse blood typing assay
    Casals-Terre, Jasmina
    Farre-Llados, Josep
    Lopez, Joan A.
    Vidal, Teresa
    Blanca Roncero, Maria
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2020, 108 (02) : 439 - 450
  • [7] A portable microfluidic paper-based analytical device for blood detection and typing assay
    Ansari, Niha
    Trambadiya, Nehal
    Lodha, Anand
    Menon, S. K.
    [J]. AUSTRALIAN JOURNAL OF FORENSIC SCIENCES, 2021, 53 (04) : 407 - 418
  • [8] Paper-based microfluidics for typing of primary and secondary human blood groups in "text"
    Li, Miaosi
    Then, Whui Lyn
    Tian, Junfei
    Shen, Wei
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2015, 249
  • [9] Paper-based assay for red blood cell antigen typing by the indirect antiglobulin test
    Natasha Yeow
    Heather McLiesh
    Liyun Guan
    Wei Shen
    Gil Garnier
    [J]. Analytical and Bioanalytical Chemistry, 2016, 408 : 5231 - 5238
  • [10] Paper-based assay for red blood cell antigen typing by the indirect antiglobulin test
    Yeow, Natasha
    McLiesh, Heather
    Guan, Liyun
    Shen, Wei
    Garnier, Gil
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (19) : 5231 - 5238