Cochlear implant device failure after cardioversion

被引:3
|
作者
Schild, C. [1 ]
Beck, R. [1 ]
Boedeker, C. C. [1 ]
Wesarg, T. [1 ]
Arndt, S. [1 ]
Aschendorff, A. [1 ]
Laszig, R. [1 ]
机构
[1] Univ Freiburg, Hals Nasen Ohren Klin, D-79106 Freiburg, Germany
关键词
Cochlear implant; Cardioversion; Defibrillation; Dysfunction; Reimplantation;
D O I
10.1007/s00106-011-2298-8
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Cochlear implantation has become the standard procedure for the treatment of severe to profound hearing loss, even in patients with underlying diseases. We report the case of a CI patient who underwent cardiac defibrillation, following which he reported a reduction in sound quality, a worsening of tinnitus, as well as headaches. An integrity test showed multiple electrode anomalies which could not be directly attributed to the cardioversion. We performed explantation and reimplantation of the CI. During the course of rehabilitation, the patient showed good results in speech comprehension. Since cardioversion, like any electrical monopolar treatment, may damage CI, it should be avoided in CI patients wherever possible. If it cannot be avoided, we strongly recommend removing the sound processor during treatment sessions. When device-related problems occur, the treatment of choice is reimplantation.
引用
收藏
页码:461 / +
页数:3
相关论文
共 50 条
  • [1] Cochleaimplantatdysfunktion nach KardioversionCochlear implant device failure after cardioversion
    C. Schild
    R. Beck
    C.C. Boedeker
    T. Wesarg
    S. Arndt
    A. Aschendorff
    R. Laszig
    [J]. HNO, 2011, 59
  • [2] COCHLEAR IMPLANT DEVICE FAILURE - DIAGNOSIS AND MANAGEMENT
    KILENY, PR
    MEITELES, LZ
    ZWOLAN, TA
    TELIAN, SA
    [J]. AMERICAN JOURNAL OF OTOLOGY, 1995, 16 (02): : 164 - 171
  • [3] RETRACTED ​ARTICLE: Cochlear implant device activation after 7 days in cochlear implant recipients
    Mingfang Diao
    Jianjun Sun
    Fangjie Tian
    Yili Ding
    Yuqing Wang
    [J]. European Archives of Oto-Rhino-Laryngology, 2019, 276 : 281 - 281
  • [4] The Function of Cochlear Implant After Cardioversion in a Patient With Atrial Flutter: A Case Report
    Kim, Hong Chan
    Sung, Chung Man
    Yang, Hyung Chae
    Cho, Hyong-Ho
    [J]. ENT-EAR NOSE & THROAT JOURNAL, 2023, 102 (05) : NP229 - NP231
  • [5] A multicenter study of device failure in European cochlear implant centers
    Battmer, Rolf-Dieter
    O'Donoghue, Gerard M.
    Lenarz, Thomas
    [J]. EAR AND HEARING, 2007, 28 (02): : 95S - 99S
  • [6] Cochlear Implant Device Failure Secondary to Cholesterol Granuloma-Mediated Cochlear Erosion
    Neilan, Ryan Eric
    Pawlowski, Karen
    Isaacson, Brandon
    Roland, Peter S.
    [J]. OTOLOGY & NEUROTOLOGY, 2012, 33 (05) : 733 - 735
  • [7] The experience of device failure after cochlear implantation
    Jeong Heon Kim
    Yeonjoo Choi
    Woo Seok Kang
    Hong Ju Park
    Joong Ho Ahn
    Jong Woo Chung
    [J]. Journal of Otolaryngology - Head & Neck Surgery, 52
  • [8] The experience of device failure after cochlear implantation
    Kim, Jeong Heon
    Choi, Yeonjoo
    Kang, Woo Seok
    Park, Hong Ju
    Ahn, Joong Ho
    Chung, Jong Woo
    [J]. JOURNAL OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2023, 52 (01)
  • [9] INEXPENSIVE COCHLEAR IMPLANT DEVICE
    KANCHANARAK, C
    SIRIRATWATANAKUL, N
    SAENGIN, A
    BOONYANUKUL, S
    KRAIROJANANAN, T
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1991, 100 (12): : 984 - 988
  • [10] The Effects of Cochlear Implant Electrode Deactivation on Speech Perception and in Predicting Device Failure
    Zeitler, Daniel M.
    Lalwani, Anil K.
    Roland, J. Thomas, Jr.
    Habib, Mirette G.
    Gudis, David
    Waltzman, Susan B.
    [J]. OTOLOGY & NEUROTOLOGY, 2009, 30 (01) : 7 - 13