Reproducibility-Repeatability of Choroidal Thickness Calculation Using Optical Coherence Tomography

被引:32
|
作者
Benavente-Perez, Alexandra [1 ]
Hosking, Sarah L. [2 ,3 ,4 ]
Logan, Nicola S. [2 ]
Bansal, Dheeraj [5 ]
机构
[1] SUNY State Coll Optometry, Dept Biol Sci, New York, NY USA
[2] Aston Univ, Sch Life & Hlth Sci, Ophthalm Res Grp, Birmingham B4 7ET, W Midlands, England
[3] City Univ London, Dept Optometry & Visual Sci, London, England
[4] Univ Melbourne, Dept Ophthalmol, Melbourne, Vic, Australia
[5] InnZ Med AB, Ljungskile, Sweden
关键词
choroidal thickness; optical coherence tomography; in vivo manual measurement; axial length; FIBER LAYER THICKNESS; AXIAL LENGTH; OCULAR BIOMETRY; MYOPIA; CHICKS; CHILDREN; OCT; AGE;
D O I
10.1097/OPX.0b013e3181f3eced
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To evaluate the repeatability and reproducibility of subfoveal choroidal thickness (CT) calculations performed manually using optical coherence tomography (OCT). Methods. The CT was imaged in vivo at each of two visits on 11 healthy volunteers (mean age, 35.72 +/- 13.19 years) using the spectral domain OCT. CT was manually measured after applying ImageJ processing filters on 15 radial subfoveal scans. Each radial scan was spaced 12 degrees from each other and contained 2500 A-scans. The coefficient of variability, coefficient of repeatability (CoR), coefficient of reproducibility, and intraclass correlation coefficient determined the reproducibility and repeatability of the calculation. Axial length (AL) and mean spherical equivalent refractive error were measured with the IOLMaster and an open view autorefractor to study their potential relationship with CT. Results. The within-visit and between-visit coefficient of variability, CoR, coefficient of reproducibility, and intraclass correlation coefficient were 0.80, 2.97% 2.44%, and 99%, respectively. The subfoveal CT correlated significantly with AL (R = -0.60, p = 0.05). Conclusions. The subfoveal CT could be measured manually in vivo using OCT and the readings obtained from the healthy subjects evaluated were repeatable and reproducible. It is proposed that OCT could be a useful instrument to perform in vivo assessment and monitoring of CT changes in retinal disease. The preliminary results suggest a negative correlation between subfoveal CT and AL in such a way that it decreases with increasing AL but not with refractive error. (Optom Vis Sci 2010;87:867-872)
引用
收藏
页码:867 / 872
页数:6
相关论文
共 50 条
  • [1] Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography
    Muscat, S
    McKay, N
    Parks, S
    Kemp, E
    Keating, D
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 (06) : 1791 - 1795
  • [2] Repeatability and Reproducibility of Macular Thickness Measurements Using Fourier Domain Optical Coherence Tomography
    Bruce, Alison
    Pacey, Ian E.
    Dharni, Poonam
    Scally, Andy J.
    Barrett, Brendan T.
    [J]. OPEN OPHTHALMOLOGY JOURNAL, 2009, 3 : 10 - 14
  • [3] Influence of choroidal thickness on subfoveal choroidal thickness measurement repeatability using enhanced depth imaging optical coherence tomography
    Cho, A. R.
    Choi, Y. J.
    Kim, Y. T.
    [J]. EYE, 2014, 28 (10) : 1152 - 1159
  • [4] Influence of choroidal thickness on subfoveal choroidal thickness measurement repeatability using enhanced depth imaging optical coherence tomography
    A R Cho
    Y J Choi
    Y T Kim
    [J]. Eye, 2014, 28 : 1151 - 1160
  • [5] Repeatability and reproducibility of retinal and choroidal thickness measurements in Diabetic Macular Edema using Swept-source Optical Coherence Tomography
    Sala-Puigdollers, Anna
    Figueras-Roca, Marc
    Hereu, Mireia
    Hernandez, Teresa
    Morato, Montse
    Adan, Alfredo
    Zarranz-Ventura, Javier
    [J]. PLOS ONE, 2018, 13 (07):
  • [6] Repeatability and reproducibility of retinal and choroidal thickness measurements in Diabetic Macular Edema using Swept-source Optical Coherence Tomography
    Sala-Puigdollers, Anna
    Zarranz-Ventura, Javier
    Figueras-Roca, Marc
    Hereu, Mireia
    Hernandez, Teresa
    Morato, Montserrat
    Mesquida, Marina
    Llorens, Victor
    Teresa Sainz de la Maza, Maria
    Adan Civera, Alfredo
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [7] Repeatability of choroidal thickness measurements in healthy subjects using RTVue XR optical coherence tomography
    Mansoori, Tarannum
    Charan, Aknoor Sree Ram
    Suresh, Narala
    Pesala, Veerendranath
    Nagalla, Balakrishna
    [J]. JOURNAL OF CURRENT OPHTHALMOLOGY, 2022, 34 (04): : 436 - 441
  • [8] Repeatability and Reproducibility of Manual Choroidal Volume Measurements Using Enhanced Depth Imaging Optical Coherence Tomography
    Chhablani, Jay
    Barteselli, Giulio
    Wang, Haiyan
    El-Emam, Sharif
    Kozak, Igor
    Doede, Aubrey L.
    Bartsch, Dirk-Uwe
    Cheng, Lingyun
    Freeman, William R.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (04) : 2274 - 2280
  • [9] Repeatability and reproducibility of fast macular thickness mapping with stratus optical coherence tomography
    Polito, A
    Del Borrello, M
    Isola, M
    Zemella, N
    Bandello, F
    [J]. ARCHIVES OF OPHTHALMOLOGY, 2005, 123 (10) : 1330 - 1337
  • [10] Choroidal thickness measurements in different ethnicities using swept source optical coherence tomography: repeatability and assessment
    Alkhaldi, Saleh A.
    Aldakhil, Sulaiman
    Gray, Lyle S.
    [J]. CLINICAL AND EXPERIMENTAL OPTOMETRY, 2022, 105 (05) : 527 - 533