Evaluation of set-up errors and estimation of set-up margin during external beam radiation therapy of prostate cancer using electronic portal imaging device (EPID)

被引:0
|
作者
Khoramian, Daryoush [1 ,2 ]
Sistani, Soroush [1 ,2 ]
Farhood, Bagher [3 ]
机构
[1] Ramezan Zadeh Radiotherapy Ctr, Med Phys Dept, Yazd, Iran
[2] Ayatollah Yasrebi Hosp, Advocate Ctr Clin Res, Dept Med Phys, Kashan, Iran
[3] Kashan Univ Med Sci, Fac Paramed Sci, Dept Med Phys & Radiol, Kashan, Iran
关键词
3DCRT; EPID; prostate cancer; set-up errors; set-up margin; CONFORMAL RADIOTHERAPY; IMPACT; HEAD;
D O I
10.1017/S1460396921000297
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: In radiation therapy, accurate dose distribution in target volume requires accurate treatment setup. The set-up errors are unwanted and inherent in the treatment process. By achieving these errors, a set-up margin (SM) of clinical target volume (CTV) to planning target volume (PTV) can be determined. In the current study, systematic and random set-up errors that occurred during prostate cancer radiotherapy were measured by an electronic portal imaging device (EPID). The obtained values were used to propose the optimum CTV-to-PTV margin in prostate cancer radiotherapy. Materials and methods: A total of 21 patients with prostate cancer treated with external beam radiation therapy (EBRT) participated in this study. A total of 280 portal images were acquired during 12 months. Gross, population systematic (Sigma) and random (sigma) errors were obtained based on the portal images in Anterior-Posterior (AP), Medio-Lateral (ML) and Superior-Inferior (SI) directions. The SM of CTV to PTV were then calculated and compared by using the formulas presented by the International Commission on Radiation Units and Measurements (ICRU) 62, Stroom and Heijmen and Van Herk et al. Results: The findings showed that the population systematic errors during prostate cancer radiotherapy in AP, ML and SI directions were 1.40, 1.95 and 1.94 mm, respectively. The population random errors in AP, ML and SI directions were 2.09, 1.85 and 2.29 mm, respectively. The SM of CTV to PTV calculated in accordance with the formula of ICRU 62 in AP, ML and SI directions were 2.51, 2.68 and 3.00 mm, respectively. And according to Stroom and Heijmen, formula were 4.23, 5.19 and 548 mm, respectively. And Van Herk et al. formula were 4.96, 6.17 and 6.45 mm, respectively. Findings: The SM of CTV to PTV in all directions, based on the formulas of ICRU 62, Stroom and Heijmen and van Herk et al., were equal to 2.73, 4.98 and 5.86 mm, respectively; these values were obtained by averaging the margins in all directions.
引用
收藏
页码:493 / 500
页数:8
相关论文
共 50 条
  • [1] Evaluation of set-up errors and determination of set-up margin in pelvic radiotherapy by electronic portal imaging device (EPID)
    Noghreiyan, Vajiheh Vejdani
    Nasseri, Shahrookh
    Anvari, Kazem
    Naji, Maryam
    Momennezhad, Mehdi
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2020, 19 (02) : 150 - 156
  • [2] Evaluation of patient set-up variation in isocentric breast radiotherapy using an electronic portal imaging device (EPID)
    Petillion, S.
    Swinnen, A.
    Van den Heuvel, F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S205 - S205
  • [3] Prospective study on set-up random errors in 3D-conformal prostate cancer radiotherapy by using an electronic portal imaging device (EPID)
    Zerini, D
    Antognoni, P
    Gandini, S
    Monti, L
    Visconti, M
    Orecchia, R
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S288 - S288
  • [4] Evaluation of set-up errors in head and neck radiotherapy using electronic portal imaging
    Strbac, Bojan
    Jokic, Vesna Spasic
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (05): : 531 - 536
  • [5] Evaluation of set-up errors using electronic portal imaging in head and neck carcinoma
    Bayman, E.
    Uruk Ataman, O.
    Akman, F.
    Kynay, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 82 : S35 - S35
  • [6] Evaluation of set-up errors using electronic portal imaging in head and neck carcinoma
    Ataman, Oe.
    Demirtas Bayman, E.
    Akman, F.
    Kinay, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S306 - S306
  • [7] Set-up verification on a belly-board device using electronic portal imaging
    Roels, Sarah
    Verstraete, Jan
    Haustermans, Karin
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2007, 6 (02) : 73 - 82
  • [8] Interfractional set-up errors evaluation by daily electronic portal imaging of IMRT in head and neck cancer patients
    Pehlivan, Berrin
    Pichenot, Charlotte
    Castaing, Marine
    Auperin, Anne
    Lefkopoulos, Dimitri
    Arriagada, Rodrigo
    Bourhis, Jean
    [J]. ACTA ONCOLOGICA, 2009, 48 (03) : 440 - 445
  • [9] IMPACT OF PORTAL IMAGES FREQUENCY ON PATIENTS' SET-UP ERRORS DURING PROSTATE CANCER RADIOTHERAPY
    Korytkowska, J.
    Skorska, M.
    Ryczkowski, A.
    Bak, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S582 - S582
  • [10] Uncertainty in patient set-up margin analysis in radiation therapy
    Suzuki, Junji
    Tateoka, Kunihiko
    Shima, Katsumi
    Yaegashi, Yuji
    Fujimoto, Kazunori
    Saitoh, Yuichi
    Nakata, Akihiro
    Abe, Tadanori
    Nakazawa, Takuya
    Sakata, Kouichi
    Hareyama, Masato
    [J]. JOURNAL OF RADIATION RESEARCH, 2012, 53 (04) : 615 - 619