A High-Throughput LC-MS/MS Method for the Simultaneous Quantification of Twenty-Seven Drug Molecules in Ocular Tissues

被引:4
|
作者
Matter, Brock [1 ]
Bourne, David W. A. [1 ]
Kompella, Uday B. [1 ]
机构
[1] Univ Colorado, Pharmaceut Sci, Anschutz Med Campus,12850 East Montview Blvd, Aurora, CO 80045 USA
关键词
Drug cassette; Corticosteroids; Beta blockers; NSAIDs; Ocular drug delivery; LC-MS/MS method; Reproducibility; BETA-BLOCKERS; CORNEAL; LIPOPHILICITY; PERMEABILITY; PENETRATION; BOVINE; RABBIT; IMPACT;
D O I
10.1208/s12249-022-02333-6
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The purpose of this study was to develop a validated LC-MS/MS analytical method for the simultaneous analysis of a large cassette containing a wide range of drug substances with positive, negative, or neutral charge and further apply the method to assess octanol partition coefficient and eye tissue recovery of the drug cassette. A twenty-seven-drug cassette (N-in-one) including beta blockers, NSAIDs, and corticosteroids that range from extremely hydrophilic (sotalol) to very hydrophobic (triamcinolone hexacetanide) was used to develop an LC-MS/MS assay using QTrap 4500. An LC-MS/MS method based on gradient elution, with an eighteen-minute run time including equilibration time, was developed and validated for the rapid and simultaneous quantitation of drugs with a wide range of lipophilicities. Scheduled multiple reaction monitoring was used to maximize the scan time for each peak, ensuring sufficient scans. Method validation included lower limit of quantitation (LLOQ) and intra- and inter-day reproducibility. The LLOQ ranged from 0.5 (sotalol) to 40 fmols (dexamethasone) on column with a %RSD < 20%. The method was tested by measuring octanol:water and octanol:buffer (PBS, pH 7.4) partition coefficients and by quantitation of the drug cassette extracted from rabbit aqueous humor and cornea. Measured partition coefficients correlated positively with predicted values (r(2) =0.5-0.7). Drug recovery was >= 79% from aqueous humor and between 61 and 67% on average from cornea. A rapid, sensitive LC-MS/MS method suitable for N-in-one drug delivery screening was developed for simultaneous quantification of twenty-seven drugs in aqueous solutions and eye tissues.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] High-Throughput Fragment Screening by Affinity LC-MS
    Minh-Dao Duong-Thi
    Bergstrom, Maria
    Fex, Tomas
    Isaksson, Roland
    Ohlson, Sten
    JOURNAL OF BIOMOLECULAR SCREENING, 2013, 18 (02) : 160 - 171
  • [32] DEVELOPMENT OF HIGH THROUGHPUT LC-MS/MS ASSAYS FOR NUTRITIONAL BIOMARKER QUANTIFICATION
    Mawson, D.
    Marrs, K.
    Teale, P.
    Grace, P.
    ANNALS OF NUTRITION AND METABOLISM, 2013, 63 : 355 - 355
  • [33] Development of extraction method suitable for LC-MS/MS quantification of antibodies in tissues
    Fuchigami, Hirobumi
    Koganemaru, Shigehiro
    Yasunaga, Masahiro
    CANCER SCIENCE, 2025, 116 : 1560 - 1560
  • [34] Fast and high-throughput LC-MS characterization, and peptide mapping of engineered AAV capsids using LC-MS/MS
    Lam, Anh K.
    Zhang, Junping
    Frabutt, Dylan
    Mulcrone, Patrick L.
    Li, Lei
    Zeng, Lifan
    Herzog, Roland W.
    Xiao, Weidong
    MOLECULAR THERAPY-METHODS & CLINICAL DEVELOPMENT, 2022, 27 : 185 - 194
  • [35] Development of the quantification method of antibody concentration in tissues using LC-MS/MS
    Fuchigami, Hirobumi
    Koganemaru, Shigehiro
    Yasunaga, Masahiro
    CANCER SCIENCE, 2023, 114 : 1497 - 1497
  • [36] An LC-MS/MS method for rapid and sensitive high-throughput simultaneous determination of various protein kinase inhibitors in human plasma
    Abdelhameed, Ali S.
    Attwa, Mohamed W.
    Kadi, Adnan A.
    BIOMEDICAL CHROMATOGRAPHY, 2017, 31 (02)
  • [37] LC-MS/MS method for the simultaneous quantification of intestinal CYP and UGT activity
    Busch, Diana
    Fritz, Anja
    Partecke, Lars Ivo
    Heidecke, Claus-Dieter
    Oswald, Stefan
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 155 : 194 - 201
  • [38] A simple LC-MS/MS method for the simultaneous quantification of endocannabinoids in biological samples
    Bobrich, Maria
    Schwarz, Rico
    Ramer, Robert
    Borchert, Philipp
    Hinz, Burkhard
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2020, 1161
  • [39] High-throughput protein identification of complex mixtures by LC-MS/MS.
    Patterson, SD
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1999, 217 : U151 - U151
  • [40] An LC-MS/MS method for the simultaneous quantification of 32 steroids in human plasma
    Simkova, Marketa
    Kolatorova, Lucie
    Drasar, Pavel
    Vitku, Jana
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2022, 1201