Dose verification and calibration of the Cyberknife® by EPR/alanine dosimetry

被引:10
|
作者
Garcia, Tristan [1 ]
Lacornerie, Thomas [2 ]
Popoff, Romain [1 ]
Lourenco, Valerie [1 ]
Bordy, Jean-Marc [1 ]
机构
[1] CEA, LIST, Lab Natl Henri Becquerel, F-91191 Gif Sur Yvette, France
[2] Ctr Oscar Lambret, Serv Phys Med, F-59020 Lille, France
关键词
Cyberknife; EPR; Alanine; Dosimetry; Radiotherapy;
D O I
10.1016/j.radmeas.2011.03.031
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Standard calibration protocols in radiotherapy (AAPM TG-51 and IAEA TRS 398) recommend a 10 x 10 cm(2) irradiation field at a distance of 100 cm. However, the specific geometry of the Cyberknife machine enables only circular radiation beams with a maximum diameter of 6 cm field size at a distance of 80 cm from the source which prevents from a direct use of these protocols. Thereby, specific calibration and verification protocols are required. Within this framework, the Laboratoire National Henri Becquerel and the Centre Oscar-Lambret decided to study two points of interest: i) the feasibility of the calibration of the Cyberknife beam; ii) the verification of the delivered dose during a treatment with a very small diameter beam using Electron Paramagnetic Resonance (EPR)/alanine dosimetry. Alanine dosimeters are suitable for this work thanks to their small size, their integrating capability, and their low energy dependence as a function of both dose rate and beam energy in the radiotherapy dose range. The results of the dose rate calibration using EPR/alanine are very good, their deviations compared to the dose rate measured in the Cyberknife reference conditions are small (about 0.5%). The doses delivered during a treatment, measured within a 2% standard uncertainty, showed less than 4% deviation from the doses calculated by the treatment planning system. So, a rather good agreement is found between the measured and theoretically delivered doses. Nevertheless, the smaller the collimator diameter is, the larger the deviation is. This problem could be due to the evaluation of the output factor correction which is difficult to achieve for such small fields or to the Treatment Planning System algorithm itself and in particular the profile modelling. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:952 / 957
页数:6
相关论文
共 50 条
  • [1] Absolute calibration of the Gamma Knife® Perfexion™ and delivered dose verification using EPR/alanine dosimetry
    Hornbeck, Amaury
    Garcia, Tristan
    Cuttat, Marguerite
    Jenny, Catherine
    [J]. MEDICAL PHYSICS, 2014, 41 (06)
  • [2] Alanine dosimetry in dose verification for IMRT
    Schaeken, B.
    De Ost, B.
    Van Gestel, D.
    Meijnders, P.
    Jacobs, I.
    Goor, C.
    Vanderkam, S.
    Coelmont, A.
    Van den Weyngaert, D.
    Schaeken, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S365 - S366
  • [3] A unique calibration curve for alanine EPR dosimetry systems
    Wieser, A
    Girzikowsky, R
    [J]. APPLIED RADIATION AND ISOTOPES, 1996, 47 (11-12) : 1269 - 1275
  • [4] Calibration of EPR spectrometer and measurement of dose from radiation field around linac by alanine dosimetry
    Liang, Renyou
    Xu, Yunhua
    Li, Yuxiong
    Li, Juexin
    Gao, Juncheng
    [J]. He Jishu/Nuclear Techniques, 1993, 16 (04): : 254 - 256
  • [5] Calibration of helical tomotherapy machine using EPR/alanine dosimetry
    Perichon, Nicolas
    Garcia, Tristan
    Francois, Pascal
    Lourenco, Valerie
    Lesven, Caroline
    Bordy, Jean-Marc
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1168 - 1177
  • [6] Epr/alanine dosimetry: Just another dose measuring system?
    Schaeken, B
    Goor, C
    Scalliet, P
    [J]. EUROPEAN JOURNAL OF CANCER, 1995, 31A : 142 - 142
  • [7] Electron arc therapy treatment planning verification with alanine/EPR dosimetry
    De Angelis, C
    Mattacchioni, A
    Onori, S
    Aragno, D
    de Paula, U
    Panichelli, V
    [J]. APPLIED RADIATION AND ISOTOPES, 2000, 52 (05) : 1203 - 1207
  • [8] Alanine EPR dosimetry in brachytherapy
    De Angelis, C
    Onori, S
    Petetti, E
    Piermattei, A
    Azario, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (05): : 1181 - 1191
  • [9] Validation of dose painting of lung tumours using alanine/EPR dosimetry
    Knudtsen, Ingerid Skjei
    Svestad, Jorund Graadal
    Sande, Erlend Peter Skaug
    Rekstad, Bernt Louni
    Rodal, Jan
    van Elmpt, Wouter
    Ollers, Michel
    Hole, Eli Olaug
    Malinen, Eirik
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (06): : 2243 - 2254
  • [10] Alanine/EPR dosimetry applied to the verification of a total body irradiation protocol and treatment planning dose calculation using a humanoid phantom
    Schaeken, B.
    Lelie, S.
    Meijnders, P.
    Van den Weyngaert, D.
    Janssens, H.
    Verellen, D.
    [J]. MEDICAL PHYSICS, 2010, 37 (12) : 6292 - 6299