Response of lithium formate EPR dosimeters at photon energies relevant to the dosimetry of brachytherapy

被引:26
|
作者
Adolfsson, Emelie [1 ]
Carlsson, Gudrun Alm [1 ]
Grindborg, Jan-Erik [2 ]
Gustafsson, Hakan [1 ]
Lund, Eva [1 ]
Tedgren, Asa Carlsson [1 ,2 ]
机构
[1] Linkoping Univ, Dept Med & Hlth Sci IMH, SE-58185 Linkoping, Sweden
[2] Swedish Radiat Safety Author, SE-17116 Stockholm, Sweden
关键词
lithium formate EPR; brachytherapy; dosimetry; detector response; kV x rays; HDR BRACHYTHERAPY; DOSE MEASUREMENTS; IR-192; SOURCE; X-RAYS; DETECTOR; SPECTRA; TLD; VERIFICATION; PROTOCOL; SYSTEM;
D O I
10.1118/1.3475938
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate experimentally the energy dependence of the detector response of lithium formate EPR dosimeters for photon energies below 1 MeV relative to that at Co-60 energies. High energy photon beams are used in calibrating dosimeters for use in brachytherapy since the absorbed dose to water can be determined with high accuracy in such beams using calibrated ion chambers and standard dosimetry protocols. In addition to any differences in mass-energy absorption properties between water and detector, variations in radiation yield (detector response) with radiation quality, caused by differences in the density of ionization in the energy imparted (LET), may exist. Knowledge of an eventual deviation in detector response with photon energy is important for attaining high accuracy in measured brachytherapy dose distributions. Methods: Lithium formate EPR dosimeters were irradiated to known levels of air kerma in 25-250 kV x-ray beams and in Cs-137 and Co-60 beams at the Swedish Secondary Standards Dosimetry Laboratory. Conversions from air kerma free in air into values of mean absorbed dose to the detectors were made using EGSnrc MC simulations and x-ray energy spectra measured or calculated for the actual beams. The signals from the detectors were measured using EPR spectrometry. Detector response (the EPR signal per mean absorbed dose to the detector) relative to that for Co-60 was determined for each beam quality. Results: Significant decreases in the relative response ranging from 5% to 6% were seen for x-ray beams at tube voltages <= 180 kV. No significant reduction in the relative response was seen for Cs-137 and 250 kV x rays. Conclusions: When calibrated in Co-60 or MV photon beams, corrections for the photon energy dependence of detector response are needed to achieve the highest accuracy when using lithium formate EPR dosimeters for measuring absorbed doses around brachytherapy sources emitting photons in the energy range of 20-150 keV such as 169Yb and electronic sources. c 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3475938]
引用
收藏
页码:4946 / 4959
页数:14
相关论文
共 28 条
  • [1] Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (&lt;1 MeV)
    Tedgren, Asa Carlsson
    Hedman, Angelica
    Grindborg, Jan-Erik
    Carlsson, Gudrun Alm
    MEDICAL PHYSICS, 2011, 38 (10) : 5539 - 5550
  • [2] Dosimetry of stereotactic radiosurgery using lithium formate EPR dosimeters
    Waldeland, Einar
    Horling, Magnus
    Hole, Eli Olaug
    Sagstuen, Einar
    Malinen, Eirik
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (08): : 2307 - 2316
  • [3] Dosimetry of ion beams using lithium formate EPR dosimeters
    Waldeland, E
    Hole, E
    Sagstuen, E
    Stenerlöw, B
    Grusell, E
    Malinen, E
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S122 - S122
  • [4] LITHIUM FORMATE EPR DOSIMETRY FOR DOSIMETRY VERIFICATIONS IN RADIOTHERAPY
    Olsson, S.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S27 - S27
  • [5] A practical EPR dosimetry system for routine use in radiotherapy: uncertainty analysis of lithium formate dosimeters at the therapeutic dose level
    Hoefel, Sebastian
    Stehle, Michael
    Zwicker, Felix
    Fix, Michael K.
    Drescher, Malte
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (04):
  • [6] The energy dependence of lithium formate EPR dosimeters for clinical electron beams
    Malinen, Eirik
    Waldeland, Einar
    Hole, Eli Olaug
    Sagstuen, Einar
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14): : 4361 - 4369
  • [7] LET effects following neutron irradiation of lithium formate EPR dosimeters
    Malinen, E
    Waldeland, E
    Hole, EO
    Sagstuen, E
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2006, 63 (04) : 861 - 869
  • [8] Evaluation of a lithium formate EPR dosimetry system for dose measurements around 192Ir brachytherapy sources
    Antonovic, Laura
    Gustafsson, Hakan
    Carlsson, Gudrun Alm
    Tedgren, Asa Carlsson
    MEDICAL PHYSICS, 2009, 36 (06) : 2236 - 2247
  • [9] Measurement of absorbed dose to water around an electronic brachytherapy source. Comparison of two dosimetry systems: lithium formate EPR dosimeters and radiochromic EBT2 film
    Adolfsson, Emelie
    White, Shane
    Landry, Guillaume
    Lund, Eva
    Gustafsson, Hakan
    Verhaegen, Frank
    Reniers, Brigitte
    Tedgren, Asa Carlsson
    Carlsson, Gudrun Alm
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (09): : 3869 - 3882
  • [10] Characterization of lithium formate EPR dosimeters for high dose applications - Comparison with alanine
    Waldeland, Einar
    Helt-Hansen, Jakob
    Malinen, Eirik
    RADIATION MEASUREMENTS, 2011, 46 (02) : 213 - 218