Measurement of capillary refill time with a handheld prototype device: a comparative validation study in healthy volunteers

被引:5
|
作者
Ruste, M. [1 ,2 ,4 ]
Cazenave, L. [1 ]
Tardif, M. [3 ]
Saint-Jean, C. [3 ]
Fellahi, J. -L. [1 ,2 ,4 ]
Lagreze, M. Jacquet [1 ,2 ,4 ]
机构
[1] Hosp Civils Lyon, Louis Pradel Hosp, Dept Anesthesiol & Intens Care, Lyon, France
[2] Univ Claude Bernard Lyon 1, Fac Med Lyon Est, Lyon, France
[3] Rochelle Univ, Lab MIA, F-17031 La Rochelle, France
[4] INSERM, Lab CarMeN, IRIS Team, U1060, Lyon, France
关键词
Capillary refill time; Device; Vascular occlusion test; Agreement; Healthy volunteers; Validation study; DiCART (TM); VASCULAR RESPONSIVENESS; PERIPHERAL PERFUSION; AGREEMENT;
D O I
10.1007/s10877-021-00757-2
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Validity and reproducibility of clinical capillary refill time (CRT) measurement depend on many factors in daily routine practice. We conducted a prospective validation study of an automatized handheld prototype device providing standardized CRT assessment (DiCART (TM)) in 20 healthy volunteers. Three different methods of CRT measurement were compared before and during dynamic circulatory changes induced by venous and arterial occlusion tests at both upper and lower limb levels: CRTCLIN corresponding to basic clinical assessment and considered as the reference method; CRTVIDEO corresponding to off-line videos reviewed by investigators recorded by DiCART (TM); and CRTDiCART corresponding to on-line videos analysed by a built-in proprietary mathematical algorithm included in DiCART (TM). Five subjects were excluded because of a DiCART (TM) dysfunction. ROCAUC to detect arterial occlusion test changes at the upper limb level were 1.00 (95%CI 1.00; 1.00), 0.96 (95%CI 0.88; 1.00), and 0.92 (95%CI 0.79; 1.00) for CRTCLIN, CRTVIDEO, and CRTDiCART, respectively. Precision of CRTCLIN and CRTVIDEO were significantly better than CRTDiCART (0.18 and 0.20 vs. 0.28; P < 0.05). Percentages of error were 76% and 87% for CRTVIDEO and CRTDiCART, respectively. DiCART (TM) had an excellent discrimination to detect major changes in CRT induced by arterial ischemia. However, the perfectible precision, the poor agreement with clinical assessment and numerous device dysfunctions give leads to the development of a further version of the prototype before promoting its use in clinical practice.
引用
收藏
页码:1271 / 1278
页数:8
相关论文
共 50 条
  • [21] Evaluation of Automated Finger Compression for Capillary Refill Time Measurement in Pediatrics
    Nickel, Amanda J.
    Jiang, Shen
    Napolitano, Natalie
    Donoghue, Aaron
    Nadkarni, Vinay M.
    Nishisaki, Akira
    [J]. PEDIATRIC EMERGENCY CARE, 2024, 40 (08) : 586 - 590
  • [22] Performance of Early Capillary Refill Time Measurement on Outcomes in Cardiogenic Shock An Observational, Prospective Multicentric Study
    Merdji, Hamid
    Curtiaud, Anais
    Aheto, Antoine
    Studer, Antoine
    Harjola, Veli-Pekka
    Monnier, Alexandra
    Duarte, Kevin
    Girerd, Nicolas
    Kibler, Marion
    Ait-Oufella, Hafid
    Helms, Julie
    Mebazaa, Alexandre
    Levy, Bruno
    Kimmoun, Antoine
    Meziani, Ferhat
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2022, 206 (10) : 1230 - 1238
  • [23] Optical Fibre Sensor for Simultaneous Measurement of Capillary Refill Time and Contact Pressure
    Liu, Chong
    Correia, Ricardo
    Ballaji, Hattan
    Korposh, Serhiy
    Hayes-Gill, Barrie
    Morgan, Stephen
    [J]. SENSORS, 2020, 20 (05)
  • [24] Impact of Skin Color on Accuracy of Capillary Refill Time Measurement by Pulse Oximeter
    Nickel, Amanda
    Jiang, Shen
    Napolitano, Natalie
    Saeki, Kota
    Hirahara, Hideaki
    Nadkarni, Vinay
    Nishisaki, Akira
    [J]. CIRCULATION, 2018, 138
  • [25] Measurement of capillary refill time (CRT) in term neonates is not a reliable indicator of cardiovascular status
    LeFlore, JL
    Engle, WD
    [J]. PEDIATRIC RESEARCH, 2003, 53 (04) : 419A - 419A
  • [26] Feedback function for capillary refilling time measurement device
    Shinozaki, Masayoshi
    Nakada, Taka-aki
    Kawaguchi, Rui
    Yoshimura, Yuichiro
    Nakaguchi, Toshiya
    Haneishi, Hideaki
    Oda, Shigeto
    [J]. CRITICAL CARE, 2019, 23 (01):
  • [27] Feedback function for capillary refilling time measurement device
    Masayoshi Shinozaki
    Taka-aki Nakada
    Rui Kawaguchi
    Yuichiro Yoshimura
    Toshiya Nakaguchi
    Hideaki Haneishi
    Shigeto Oda
    [J]. Critical Care, 23
  • [28] Retinol concentrations in capillary dried blood spots from healthy volunteers: method validation
    Craft, NE
    Bulux, J
    Valdez, C
    Li, YK
    Solomons, NW
    [J]. AMERICAN JOURNAL OF CLINICAL NUTRITION, 2000, 72 (02): : 450 - 454
  • [29] Pilot Study: Safety and Performance Validation of an Ingestible Medical Device for Collecting Small Intestinal Liquid in Healthy Volunteers
    Tronel, Alexandre
    Silvent, Anne-Sophie
    Buelow, Elena
    Giai, Joris
    Leroy, Corentin
    Proust, Marion
    Martin, Donald
    Le Gouellec, Audrey
    Soranzo, Thomas
    Mathieu, Nicolas
    [J]. METHODS AND PROTOCOLS, 2024, 7 (01)
  • [30] Capillary refill time: a study of interobserver reliability among nurses and nurse assistants
    Brabrand, Mikkel
    Hosbond, Susanne
    Folkestad, Lars
    [J]. EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2011, 18 (01) : 46 - 49