Overestimation of HDL-cholesterol using a homogeneous "Direct" assay

被引:2
|
作者
Zhao, W
Chaffin, C
Desmond, RA
Hodges, B
Daly, TM
Robinson, CA
Hardy, RW
机构
[1] Univ Alabama, Dept Pathol, Div Lab Med, Birmingham, AL 35294 USA
[2] Univ Alabama, Dept Med, Med Stat Div, Birmingham, AL 35294 USA
关键词
D O I
10.1002/jcla.20000
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Direct HDL-cholesterol (HDLc) assays have several advantages over other assays that are more laborious and time-consuming. A recent College of American Pathologists (CAP) report indicates that at least 385 Beckman LX-20 analyzers use the selective detergent (homogeneous or direct) method for analyzing HDLc. There is no published evaluation of direct HDLc assays on the Beckman platform. Here we report our evaluation of a direct HDLc assay marketed by Beckman for the Beckman LX-20 analyzers. In general, the assay performed well; however, the total error was not within National Cholesterol Education Program (NCEP) guidelines. This was largely because of a significant positive bias that appeared to be due, at least in part, to triglycerides. This bias was worse at HDLc concentrations <40 mg/dL, which may make it difficult to identify a low HDLc concentration and consequently an increased risk for cardiovascular disease (CVD). (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:42 / 44
页数:3
相关论文
共 50 条
  • [21] False negative results in the Roche assay for HDL-cholesterol
    Bakker, AJ
    ANNALS OF CLINICAL BIOCHEMISTRY, 2003, 40 : 572 - 575
  • [22] A new liquid homogeneous assay for the determination of HDL-cholesterol -: A comparison to precipitation with phosphotungstic acid/MgCl2 and a lyophilized homogeneous assay
    Nauck, M
    Neumann, I
    März, W
    Wieland, H
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1999, 37 (05) : 537 - 543
  • [23] OPTIMIZATION OF AN HDL-CHOLESTEROL ASSAY ON THE ABBOTT TDX SYSTEM
    MELOY, MN
    STEINER, P
    STEIN, EA
    HEFELE, J
    SHAFFAR, M
    CLINICAL CHEMISTRY, 1985, 31 (06) : 947 - 948
  • [24] Evaluation of the Vitros® magnetic HDL-cholesterol assay.
    Masse, J
    Bergeron, J
    Bourget, C
    CLINICAL CHEMISTRY, 2000, 46 (06) : A97 - A97
  • [25] HDL-CHOLESTEROL ASSAY WITH THE ABBOTT VISION SYSTEM EVALUATED
    NG, RH
    ALTAFFER, M
    CLINICAL CHEMISTRY, 1987, 33 (10) : 1943 - 1943
  • [26] Multicenter evaluation of a homogeneous assay for HDL cholesterol
    Marz, W
    Nauck, M
    Jarausch, J
    Cobbaert, C
    Sagers, A
    Bernard, D
    Delanghe, J
    Honauer, G
    Lehmann, P
    Ostrich, E
    vonEckardstein, A
    Walch, S
    Wieland, H
    Assmann, G
    ATHEROSCLEROSIS, 1997, 130 : 53 - 53
  • [27] EVALUATION OF THE REFLOTRON HDL ASSAY - A 90-SECOND METHOD FOR DIRECT MEASUREMENT OF HDL-CHOLESTEROL WITH NO MANUAL PRECIPITATION STEP
    NG, RH
    SPARKS, KM
    STATLAND, BE
    CLINICAL CHEMISTRY, 1990, 36 (06) : 962 - 962
  • [28] BRAN AND HDL-CHOLESTEROL
    DIXON, M
    BMJ-BRITISH MEDICAL JOURNAL, 1978, 1 (6112): : 578 - 578
  • [29] HDL-CHOLESTEROL AND DIABETES
    WINGERD, J
    PETITTI, DB
    LANCET, 1978, 2 (8091): : 676 - 676
  • [30] HDL-CHOLESTEROL IN DEMENTIA
    KELLETT, JM
    LANCET, 1985, 2 (8445): : 43 - 43