A high-performance double-crystal monochromator soft X-ray beamline

被引:46
|
作者
Dann, TE [1 ]
Chung, SC [1 ]
Huang, LJ [1 ]
Juang, JM [1 ]
Chen, CI [1 ]
Tsang, KL [1 ]
机构
[1] Synchrotron Radiat Res Ctr, Hsinchu 300, Taiwan
关键词
beamlines; soft X-rays; double-crystal monochromators;
D O I
10.1107/S0909049597017135
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A high-performance double-crystal-monochromator soft X-ray (DCMSX) beamline has been constructed at the Synchrotron Radiation Research Center (SRRC). This beamline delivers monochromatic photon beams with energies from 1 to 9 keV and a resolving power (E/Delta E) of up to 7000. This beamline provides users with an opportunity to study many important materials, such as high-T-c superconductors, magnetic materials, catalysts, super-alloy compounds etc. Excellent EXAFS and NEXFS spectra have been routinely obtained from this beamline. Several interesting research projects are currently being conducted at this beamline. All the results show that this beamline has been constructed to meet its design goals.
引用
收藏
页码:664 / 666
页数:3
相关论文
共 50 条
  • [41] Mechanical design and performance evaluation for plane grating monochromator in a soft X-ray microscopy beamline at SSRF
    Gong, Xuepeng
    Lu, Qipeng
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2015, 23 (04) : 409 - 421
  • [42] The resolution function of a double-crystal X-ray diffractometer
    E. M. Pashaev
    I. A. Subbotin
    M. A. Chuev
    V. V. Kvardakov
    A. E. Golovanov
    I. A. Likhachev
    Instruments and Experimental Techniques, 2009, 52 : 712 - 720
  • [43] Design of a double-crystal x-ray vacuum spectrometer
    Parratt, LG
    PHYSICAL REVIEW, 1932, 41 (05): : 553 - 560
  • [44] The Resolution Function of a Double-Crystal X-ray Diffractometer
    Pashaev, E. M.
    Subbotin, I. A.
    Chuev, M. A.
    Kvardakov, V. V.
    Golovanov, A. E.
    Likhachev, I. A.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2009, 52 (05) : 712 - 720
  • [45] Improvement of the performance of sagittally focusing double-crystal monochromator
    Fu, Xuan
    Zhou, Renkui
    Zhou, Sizhong
    Guangzi Xuebao/Acta Photonica Sinica, 2000, 29 (04): : 372 - 375
  • [46] EXAMINATION OF HIGH-ANGLE DOUBLE-CRYSTAL X-RAY DIFFRACTOMETRY (HADOX)
    Fujii, Y.
    Soejima, Y.
    Okazaki, A.
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 1996, 52 : C39 - C39
  • [47] A DOUBLE-CRYSTAL (N, -N) X-RAY MONOCHROMATOR WITH SAGITTALLY BENT 2ND CRYSTAL FOR ARBITRARILY DIVERGENT RADIATION
    HRDY, J
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1994, 65 (06): : 2147 - 2148
  • [48] CONSTRUCTION AND PERFORMANCE OF A BENT CRYSTAL X-RAY MONOCHROMATOR
    STEPHENS, PW
    ENG, PJ
    TSE, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (02): : 374 - 378
  • [49] Comparative monochromator studies for a soft x-ray microfocus beamline for BESSY-II
    Weiss, MR
    Follath, R
    Senf, F
    Gudat, W
    JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 1999, 101 : 1003 - 1012
  • [50] Design and feasibility of quick EXAFS scans for a 'broomstick' double-crystal monochromator at PLS beamline
    Lee, JM
    Sung, NE
    Park, JK
    Yoon, JG
    Kim, JH
    Choi, MH
    Lee, KB
    JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 : 524 - 526