Left atrial- and left ventricular-based single lead DDD pacing

被引:2
|
作者
De Cock, CC [1 ]
Van Campen, CMC [1 ]
Vos, DHS [1 ]
Visser, CA [1 ]
机构
[1] Free Univ Amsterdam, Acad Hosp, Dept Cardiol, NL-1081 HV Amsterdam, Netherlands
来源
关键词
D O I
10.1046/j.1460-9592.2001.00486.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Single lead physiological pacing with right ventricular apex stimulation is feasible in patients with normal sinus node function and is associated with lower costs and a reduced complication rate. Left ventricular or biventricular pacing using the tributaries of the coronary sinus was reported to improve functional status and quality-of-life in patients with advanced congestive heart failure and intraventricular conduction delays. In addition, the proximal part of the coronary sinus has been successfully used for left atrial stimulation. The feasibility and pacing characteristics of left atrial- and left ventricular-based temporary pacing using a conventional single pass lead were studied. The tip of the lead was advanced in a distal position of the lateral or posterolateral branches of the coronary sinus, providing a stable position of the middle and proximal electrode pair in the first 2-7 cm of the coronary sinus. A successful stable position was achieved in 13 of 21 consecutive attempts. Using overlapping biphasic (OLBI) stimulation, left atrial and left ventricular pacing threshold were 2.3 +/- 0.6 and 2.5 +/- 0.5 V, respectively. Acceptable sensing values were measured for P waves (4.1 +/- 0.7 mV) and R waves (9.7 +/- 3.9 mV). There was a good correlation between the diameter of the coronary sinus as assessed by quantitative coronary angiography at the level of the middle and proximal rings and atrial pacing threshold (r = 0.92, P < 0.0001). Thus, single lead left atrial and left ventricular-based pacing is feasible although further development is necessary to increase the success rate for stable stimulation at both sires.
引用
收藏
页码:486 / 488
页数:3
相关论文
共 50 条
  • [21] A new lead system far left ventricular pacing
    Gold, MR
    Hummel, JD
    Olsovsky, MR
    Giudici, MC
    Laskey, WL
    Shorofsky, SR
    Tockman, BA
    Ding, J
    Spinelli, JC
    CIRCULATION, 1999, 100 (18) : 515 - 515
  • [22] Inadvertent placement of left ventricular endocardial pacing lead
    Heck, Patrick M.
    Hoole, Stephen P.
    Cooper, John P.
    Begley, David A.
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2012, 13 (10) : 656 - 659
  • [23] Left ventricular anterolateral site is the preferable left ventricular lead position for an upgrade to biventricular pacing
    Ogano, M.
    Iwasaki, Y.
    Okada, T.
    Tanabe, J.
    Shimizu, W.
    Asai, K.
    EUROPEAN HEART JOURNAL, 2023, 44
  • [24] LEFT ATRIAL PACING
    WATSON, CG
    LAU, SH
    COHEN, SI
    STEIN, E
    DAMATO, AN
    NEW ENGLAND JOURNAL OF MEDICINE, 1968, 278 (24): : 1349 - +
  • [25] IMPROVEMENT IN LEFT-VENTRICULAR FUNCTION DURING PHYSIOLOGIC PACING (VENTRICULAR RATE RESPONSIVE AND DDD)
    ALTIERI, PI
    MARTINEZ, JA
    BANCHS, H
    CLINICAL RESEARCH, 1988, 36 (03): : A258 - A258
  • [26] Does permanent left ventricular-based pacing decrease mitral regurgitation in patients with severe congestive heart failure?
    Etienne, Y
    Valls-Bertault, V
    Touiza, A
    Mansourati, J
    Gilard, M
    Blanc, JJ
    EUROPEAN HEART JOURNAL, 2000, 21 : 410 - 410
  • [27] EFFECT OF PAIRED ATRIAL PACING ON LEFT ATRIAL AND VENTRICULAR PERFORMANCE IN DOG
    CHAMBERLAIN, DA
    LEINBACH, RC
    SANDERS, CA
    CARDIOVASCULAR RESEARCH, 1970, 4 (01) : 116 - +
  • [28] BIPOLAR CORONARY SINUS LEAD FOR LEFT ATRIAL AND LEFT VENTRICULAR RECORDING
    CASTELLANOS, A
    CASTILLO, CA
    MYERBURG, RJ
    AMERICAN HEART JOURNAL, 1971, 81 (06) : 832 - +
  • [29] ECHOCARDIOGRAPHIC ASSESSMENT OF LEFT VENTRICULAR FUNCTION DETERMINED BY ATRIAL OR VENTRICULAR PACING
    YAMAGUCH.T
    UMEDA, T
    OMOTO, R
    FURUTA, S
    MATSUDA, M
    TAKAHASH.H
    MACHII, K
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1973, 37 (06): : 603 - 603
  • [30] A dedicated left atrial pacing and sensing lead improves the outcome of atrial resynchronization pacing for atrial fibrillation
    Sankaranarayanan, Rajiv
    Holloway, Russell
    James, Michael
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 28A - 28A