Implementation of 3 T Lactate-Edited 3D 1H MR Spectroscopic Imaging with Flyback Echo-Planar Readout for Gliomas Patients

被引:32
|
作者
Park, Ilwoo [1 ,2 ]
Chen, Albert P. [3 ]
Zierhut, Matthew L. [1 ]
Ozturk-Isik, Esin [1 ]
Vigneron, Daniel B. [1 ,2 ]
Nelson, Sarah J. [1 ,2 ]
机构
[1] Univ Calif San Francisco, Surbeck Lab Adv Imaging, Dept Radiol & Biomed Imaging, San Francisco, CA 94158 USA
[2] UCSF UCB Joint Grad Grp Bioengn, San Francisco, CA USA
[3] GE Healthcare, Toronto, ON M5V 3Y3, Canada
关键词
3D magnetic resonance spectroscopic imaging; In vivo lactate detection; Brain tumor; Glioblastoma multiforme; BRAIN-TUMORS; IN-VIVO; CEREBRAL GLIOMAS; EMT6; TUMORS; LIPIDS; VOLUME; TESLA; EXCITATION; SEQUENCES; CHOLINE;
D O I
10.1007/s10439-010-0128-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to implement a new lactate-edited 3D H-1 magnetic resonance spectroscopic imaging (MRSI) sequence at 3 T and demonstrate the feasibility of using this sequence for measuring lactate in patients with gliomas. A 3D PRESS MRSI sequence incorporating shortened, high bandwidth 180 degrees pulses, new dual BASING lactate-editing pulses, high bandwidth very selective suppression (VSS) pulses and a flyback echo-planar readout was implemented at 3 T. Over-prescription factor of PRESS voxels was optimized using phantom to minimize chemical shift artifacts. The lactate-edited flyback sequence was compared with lactate-edited MRSI using conventional elliptical k-space sampling in a phantom and volunteers, and then applied to patients with gliomas. The results demonstrated the feasibility of detecting lactate within a short scan time of 9.5 min in both phantoms and patients. Over-prescription of voxels gave less chemical shift artifacts allowing detection of lactate on the majority of the selected volume. The normalized SNR of brain metabolites using the flyback encoding were comparable to the SNR of brain metabolites using conventional phase encoding MRSI. The specialized lactate-edited 3D MRSI sequence was able to detect lactate in brain tumor patients at 3 T. The implementation of this technique means that brain lactate can be evaluated in a routine clinical setting to study its potential as a marker for prognosis and response to therapy.
引用
收藏
页码:193 / 204
页数:12
相关论文
共 50 条
  • [21] Feasibility of using echo planar spectroscopic imaging (EPSI) for 3 dimensional 1H metabolic imaging in the human brain
    Kimura, T
    Taki, H
    Hori, M
    Naruse, S
    Hirata, S
    Komura, K
    Takahashi, T
    ULTRAFAST MAGNETIC RESONANCE IMAGING IN MEDICINE, 1999, 1192 : 255 - 258
  • [22] Readout-segmented Echo-planar Imaging for Diffusion-weighted Imaging in the Pelvis at 3T-A Feasibility Study
    Thian, Yee Liang
    Xie, Wanying
    Porter, David A.
    Ang, Bertrand Weileng
    ACADEMIC RADIOLOGY, 2014, 21 (04) : 531 - 537
  • [23] Assessment of malignancy in gliomas by 3T 1H MR spectroscopy
    Jeun, SS
    Kim, MC
    Kim, BS
    Lee, JM
    Chung, ST
    Oh, CH
    Lee, SY
    Choe, BY
    CLINICAL IMAGING, 2005, 29 (01) : 10 - 15
  • [24] High-resolution functional MRI at 3 T: 3D/2D echo-planar imaging with optimized physiological noise correction
    Lutti, Antoine
    Thomas, David L.
    Hutton, Chloe
    Weiskopf, Nikolaus
    MAGNETIC RESONANCE IN MEDICINE, 2013, 69 (06) : 1657 - 1664
  • [25] Spherical echo-planar time-resolved imaging (sEPTI) for rapid 3D quantitative T2* and susceptibility imaging
    Wang, Nan
    Liao, Congyu
    Cao, Xiaozhi
    Nishimura, Mark
    Brackenier, Yannick W. E.
    Yurt, Mahmut
    Gao, Mengze
    Abraham, Daniel
    Alkan, Cagan
    Iyer, Siddharth Srinivasan
    Zhou, Zihan
    Jeong, Hwihun
    Kerr, Adam
    Haldar, Justin P.
    Setsompop, Kawin
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (01) : 121 - 137
  • [26] Post mortem high resolution diffusion MRI for large specimen imaging at 11.7 T with 3D segmented echo-planar imaging
    Sebille, Sophie Bernadette
    Rolland, Anne-Sophie
    Welter, Marie-Laure
    Bardinet, Eric
    Santin, Mathieu David
    JOURNAL OF NEUROSCIENCE METHODS, 2019, 311 : 222 - 234
  • [27] Simple and broadly applicable automatic quality control for 3D 1H MR spectroscopic imaging data of the prostate
    Tayari, Nassim
    Obels, Jiri
    Kobus, Thiele
    Scheenen, Tom W. J.
    Heerschap, Arend
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (05) : 2887 - 2895
  • [28] Fast 3D 1H spectroscopic imaging at 3 tesla using spectroscopic missing-pulse SSFP with 3D spatial preselection
    Schuster, Christian
    Dreher, Wolfgang
    Geppert, Christian
    Leibfritz, Dieter
    MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (01) : 82 - 89
  • [29] BREATH-HOLD 3D STAR MR-ANGIOGRAPHY OF THE RENAL-ARTERIES USING SEGMENTED ECHO-PLANAR IMAGING
    WIELOPOLSKI, PA
    ADAMIS, M
    PRASAD, P
    GAA, J
    EDELMAN, R
    MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (03) : 432 - 438
  • [30] 3D ''blipped'' echo-planar imaging of cortical activation with volumetric interpolation of k-space
    Mock, BJ
    Sorenson, JA
    RADIOLOGY, 1996, 201 : 704 - 704