k-space weighted image average (KWIA) for ASL-based dynamic MR angiography and perfusion imaging

被引:0
|
作者
Zhao, Chenyang [1 ]
Shao, Xingfeng [1 ]
Yan, Lirong [1 ]
Wang, Danny J. J. [1 ]
机构
[1] Univ Southern Calif, Keck Sch Med, USC Mark & Mary Stevens Neuroimaging & Informat I, Lab FMRI Technol LOFT, Los Angeles, CA 90033 USA
关键词
k-space weighted image average; Denoising; Arterial spin labeling; Perfusion MRI; Dynamic magnetic resonance angiography; SPIN; FLOW;
D O I
10.1016/j.mri.2021.11.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel denoising algorithm termed k-space weighted image average (KWIA) was proposed to improve the signal-to-noise ratio (SNR) of dynamic MRI, such as arterial spin labeling (ASL)-based dynamic magnetic resonance angiography (dMRA) and perfusion imaging. KWIA divides the k-space of each time frame into multiple rings, the central ring of the k-space remains intact to preserve the image contrast and temporal resolution, while outer rings are progressively averaged with neighboring time frames to increase SNR. Simulations and in-vivo dMRA and multi-delay ASL studies were performed to evaluate the performance of KWIA under various MRI acquisition conditions. SNR ratios and temporal signal errors between KWIA-processed and the original data were measured. Visualization of dynamic blood flow signals as well as quantitative parametric maps were evaluated for KWIA-processed images as compared to the original images. KWIA achieved a SNR ratio of 1.73 for dMRA and 2.0 for multi-delay ASL respectively, which were in accordance with the theoretical predictions. Improved visualization of dynamic blood flow signals was demonstrated using KWIA in distal small vessels in dMRA and small brain structures in multi-delay ASL. Approximately 5% temporal errors were observed in both KWIA-processed dMRA and ASL signals. Fine anatomical features were revealed in the quantitative parametric maps of dMRA, and the residuals of model fitting were reduced for multi-delay ASL. Compared to other conventional denoising methods, KWIA is a flexible denoising algorithm that improves the SNR of ASL-based dMRA and perfusion MRI by up to 2-fold without compromising spatial and temporal resolution or quantification accuracy.
引用
收藏
页码:94 / 106
页数:13
相关论文
共 49 条
  • [21] High-resolution dynamic oxygen-17 MR imaging of mouse brain with golden-ratio-based radial sampling and k-space-weighted image reconstruction
    Liu, Yuchi
    Zhang, Yifan
    Wu, Chunying
    Zhu, Junqing
    Wang, Charlie
    Tomko, Nicholas
    Linetsky, Mikhail D.
    Salomon, Robert G.
    Ramos-Estebanez, Ciro
    Wang, Yanming
    Yu, Xin
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (01) : 256 - 263
  • [22] Noncontrast Enhanced Four-Dimensional Dynamic MRA with Golden Angle Radial Acquisition and K-space Weighted Image Contrast (KWIC) Reconstruction
    Song, Hee Kwon
    Yan, Lirong
    Smith, Robert X.
    Xue, Yiqun
    Rapacchi, Stanislas
    Srinivasan, Subashini
    Ennis, Daniel B.
    Hu, Peng
    Pouratian, Nader
    Wang, Danny J. J.
    MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (06) : 1541 - 1551
  • [23] Micturating cystourethrography performed using dynamic, gadolinium-enhanced real-time MR imaging with continuous radial k-space sampling
    Nolte-Ernsting, CC
    Glowinski, A
    Katterbach, FJ
    Buecker, A
    Adam, GB
    Rasche, V
    RADIOLOGY, 1998, 209P : 336 - 336
  • [24] Influence of the k-space trajectory on the dynamic T1-weighted signal in quantitative first-pass cardiac perfusion MRI at 3T
    Kim, Daniel
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (01) : 202 - 208
  • [25] Diagnostic performance of maximum slope: A kinetic parameter obtained from ultrafast dynamic contrast-enhanced magnetic resonance imaging of the breast using k-space weighted image contrast (KWIC)
    Ohashi, Akane
    Kataoka, Masako
    Kanao, Shotaro
    Iima, Mami
    Murata, Katsutoshi
    Weiland, Elisabeth
    Onishi, Natsuko
    Kawai, Makiko
    Toi, Masakazu
    Togashi, Kaori
    EUROPEAN JOURNAL OF RADIOLOGY, 2019, 118 : 285 - 292
  • [26] Usefulness of T2-Weighted Magnetic Resonance Image based on Radial K-space Filling Technique for Liver Cancer Radiation Therapy
    Park, J.
    Yoo, G. S.
    Kim, H. Y.
    Kim, C.
    Park, Y.
    Park, H. C.
    Cho, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E307 - E308
  • [27] MR image quality testing of k-space- and image-based parallel imaging techniques using the ACR phantom
    Chen, C. C.
    Wai, Y. Y.
    Liu, H. L.
    MEDICAL PHYSICS, 2006, 33 (06) : 2009 - 2010
  • [28] Attention incorporated network for sharing low-rank, image and k-space information during MR image reconstruction to achieve single breath-hold cardiac Cine imaging
    Xu, Siying
    Hammernik, Kerstin
    Lingg, Andreas
    Kuebler, Jens
    Krumm, Patrick
    Rueckert, Daniel
    Gatidis, Sergios
    Kuestner, Thomas
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2025, 120
  • [29] Translation of 1D Inverse Fourier Transform of K-space to an Image Based on Deep Learning for Accelerating Magnetic Resonance Imaging
    Eo, Taejoon
    Shin, Hyungseob
    Kim, Taeseong
    Jun, Yohan
    Hwang, Dosik
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION - MICCAI 2018, PT I, 2018, 11070 : 241 - 249
  • [30] Technical innovation in dynamic contrast-enhanced magnetic resonance imaging of musculoskeletal tumors: an MR angiographic sequence using a sparse k-space sampling strategy
    Laura M. Fayad
    Charles Mugera
    Theodoros Soldatos
    Aaron Flammang
    Filippo del Grande
    Skeletal Radiology, 2013, 42 : 993 - 1000