Effect of Variable Amplitude Loading in Nitinol Fatigue

被引:0
|
作者
Choules, Brian D. [1 ]
Lewis, Alexandra L. [2 ]
Gulker, Brandon G. [2 ]
Metcalf, Justin M. [2 ]
Kelley, Jace D. [2 ]
机构
[1] Embry Riddle Aeronaut Univ, Dept Mech Engn, 3700 Willow Creek Rd, Prescott, AZ 86301 USA
[2] MED Inst, Med Device Engn, 1330 Win Hentschel Blvd,Suite 100, W Lafayette, IN 47906 USA
关键词
nitinol; fatigue; variable; amplitude; cyclic; strain; fatigue-to-fracture; STENT PLACEMENT;
D O I
10.1520/STP161620180035
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
In this investigation, diamond-shaped samples laser-cut from nitinol tubing were used to investigate the effect of variable amplitude fatigue loading. Alternating strain amplitudes of 0.30 %, 0.35 %, and 0.40 % were applied with periodic high strain amplitude of 0.70 %. The high strain was applied one time for every 40 low strain cycles. Using the obtained data, the appropriateness of the Palmgren-Miner's linear and the Marco-Starkey nonlinear cumulative damage models was investigated. The variable amplitude testing results indicated a trend of decreasing fatigue life with the addition of periodic high strain. Data obtained were not sufficient to demonstrate that the Palmgren-Miners model was an accurate predictor. However, the Markco-Starkey model was found to provide additional flexibility in obtaining more precise predictions. With fatigue-to-fracture data and an appropriate variable amplitude model, the life associated with the variable amplitude loading can be estimated, and burdensome and lengthy physical variable amplitude testing can be avoided.