In Vivo Effect of Titanium Implants with Porous Zinc-Containing Coatings Prepared by Plasma Electrolytic Oxidation Method on Osseointegration in Rabbits

被引:24
|
作者
He, Jing [1 ]
Feng, Wei [2 ]
Zhao, Bao-Hong [1 ]
Zhang, Wei [3 ,4 ]
Lin, Zeng [5 ]
机构
[1] China Med Univ, Sch Stomatol, Ctr Implant Dent, Nanjing North St 117, Shenyang 110002, Liaoning, Peoples R China
[2] Jining Med Univ, Affiliated Hosp, Dept Oral Implants, Jining, Peoples R China
[3] Chinese Acad Sci, Inst Met Res, Shenyang Natl Lab Mat Sci, Corros & Protect Div, Shenyang, Liaoning, Peoples R China
[4] Jilin Normal Univ, Minist Educ, Key Lab Preparat & Applicat Environm Friendly Mat, Changchun, Jilin, Peoples R China
[5] Northeastern Univ, Sch Mech Engn & Automat, Shenyang, Liaoning, Peoples R China
关键词
bone-to-implant contact; dental implant; mineral apposition ratio; osseointegration; plasma electrolyte oxidation; zinc; CALCIUM-PHOSPHATE CEMENT; PROMOTING BONE-FORMATION; BLOOD MONONUCLEAR-CELLS; MICRO-ARC OXIDATION; TRICALCIUM PHOSPHATE; CORROSION BEHAVIOR; PURE TITANIUM; SURFACE; HYDROXYAPATITE; BRUSHITE;
D O I
10.11607/jomi.5764
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The aim of this study was to confirm the in vivo effect of surface-modified titanium (Ti) implants with zinc (Zn) prepared by a plasma electrolytic oxidation (PEO) method on the process of osseointegration. Materials and Methods: Forty-five New Zealand white rabbits were used. Implants were divided into three groups due to different surface treatments (the PEO-Zn group, the PEO-calcium phosphorus [Ca/P] group, and the sandblasted [control] group) and were randomly inserted into rabbits' mandibles. The implant surface morphology and chemical state of the relevant elements were observed. The maximum push-out force, the bone-to-implant contact (BIC), and the mineral apposition ratio (MAR) were detected at 4, 8, and 12 weeks after implantation. Results: A two-layer structure, with a dense inner layer and a porous outer layer, was found in the PEO samples. In the PEO-Zn group, CaHPO4 center dot 2H(2)O (dicalcium phosphate dihydrate) was the main calcium compound, and ZnO was identified as the predominant form of Zn. More bone deposition was found on the top of the PEO-Zn implant, which was also confirmed by scanning electron microscope (SEM) afterward, and the shear strength reached a maximum value of 1.57 +/- 0.26 MPa at 12 weeks compared with the sandblasted group and the PEO-Ca/P group. The BIC values and the MARs of the PEO-Zn group were significantly higher than the control group, especially at the early weeks (P < .05), and first reached maximum values of 73.91% +/- 9.01% and 4.89 +/- 0.60 mu m/d at 12 weeks, respectively. Conclusion: Implants with Zn coating by the PEO method could accelerate bone formation and bone remodeling, shorten the osseointegration period, and enhance bone-implant bonding force.
引用
收藏
页码:298 / 310
页数:13
相关论文
共 50 条
  • [41] Investigation of hydroxyapatite (HAP) containing coating on grade 2 titanium alloy prepared by plasma electrolytic oxidation (PEO) at low voltage
    Kostelac, Lorena
    Pezzato, Luca
    Settimi, Alessio Giorgio
    Franceschi, Mattia
    Gennari, Claudio
    Brunelli, Katya
    Rampazzo, Chiara
    Dabala, Manuele
    SURFACES AND INTERFACES, 2022, 30
  • [42] In vitro and in vivo mechanical evaluations of plasma-sprayed hydroxyapatite coatings on titanium implants: The effect of coating characteristics
    Yang, CY
    Lin, RM
    Wang, BC
    Lee, TM
    Chang, E
    Hang, YS
    Chen, PQ
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1997, 37 (03): : 335 - 345
  • [43] ENERGY-DISPERSIVE X-RAY SPECTROSCOPY MAPPING OF POROUS COATINGS OBTAINED ON TITANIUM BY PLASMA ELECTROLYTIC OXIDATION IN A SOLUTION CONTAINING CONCENTRATED PHOSPHORIC ACID WITH COPPER NITRATE
    Rokosz, K.
    Hryniewicz, T.
    Dudek, L.
    Schuetz, A.
    Heeg, J.
    Wienecke, M.
    ADVANCES IN MATERIALS SCIENCE, 2016, 16 (03): : 15 - 25
  • [44] Influence of magnetostatic interactions on magnetization process of iron-containing coatings, produced using the plasma electrolytic oxidation method
    Kharitonskii, P. V.
    Frolov, A. M.
    Boev, S. A.
    Rudnev, V. S.
    Tkachenko, I. A.
    Morozova, V. P.
    Lukiyanchuk, I. V.
    Adigamova, M. V.
    Ustinov, A. Yu.
    TRENDS IN MAGNETISM: NANOMAGNETISM, 2014, 215 : 200 - 203
  • [45] On the Stability of ZrO2 + TiO2 Coatings on Titanium Formed by Plasma Electrolytic Oxidation in a Chlorine-Containing Medium
    I. V. Malyshev
    V. S. Rudnev
    Protection of Metals and Physical Chemistry of Surfaces, 2020, 56 : 369 - 373
  • [46] Electrochemical corrosion properties of CeO2-containing coatings on AZ31 magnesium alloys prepared by plasma electrolytic oxidation
    Lim, Tae Seop
    Ryu, Hyun Sam
    Hong, Seong-Hyeon
    CORROSION SCIENCE, 2012, 62 : 104 - 111
  • [47] An X-ray photoelectron spectroscopy study of Ni, Cu-containing coatings formed by plasma electrolytic oxidation on aluminum and titanium
    E. A. Koblova
    A. Yu. Ustinov
    V. S. Rudnev
    I. V. Lukiyanchuk
    I. V. Chernykh
    Journal of Structural Chemistry, 2017, 58 : 1129 - 1136
  • [48] On the Stability of ZrO2+ TiO2Coatings on Titanium Formed by Plasma Electrolytic Oxidation in a Chlorine-Containing Medium
    Malyshev, I. V.
    Rudnev, V. S.
    PROTECTION OF METALS AND PHYSICAL CHEMISTRY OF SURFACES, 2020, 56 (02) : 369 - 373
  • [49] An X-ray photoelectron spectroscopy study of Ni, Cu-containing coatings formed by plasma electrolytic oxidation on aluminum and titanium
    Koblova, E. A.
    Ustinov, A. Yu.
    Rudnev, V. S.
    Lukiyanchuk, I. V.
    Chernykh, I. V.
    JOURNAL OF STRUCTURAL CHEMISTRY, 2017, 58 (06) : 1129 - 1136
  • [50] In-vitro corrosion and bioactivity behavior of tailored calcium phosphate-containing zinc oxide coating prepared by plasma electrolytic oxidation
    Bordbar-Khiabani, Aidin
    Ebrahimi, Sema
    Yarmand, Benyamin
    Corrosion Science, 2021, 173