Specific Absorption Rate Distribution on a Human Head Model from Inductive Power Coils

被引:0
|
作者
Ng, David C. [1 ,2 ]
Wang, Xiaoting [2 ]
Felic, Gordana K. [1 ,2 ]
Bai, Shun [2 ]
Boyd, Clive S. [1 ]
Halpern, Mark [1 ,2 ]
Skafidas, Efstratios [1 ,2 ]
机构
[1] NICTA, Victorian Res Lab, Melbourne, Vic, Australia
[2] Univ Melbourne, Dept Elect & Elect Engn, Melbourne, Vic, Australia
基金
澳大利亚研究理事会;
关键词
specific absorption rate (SAR); inductive coil; retinal implant; DESIGN;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We are interested in the amount of power that is absorbed by the eye and surrounding head tissue due to operation of inductive coils which are placed in proximity to the eye. These coils provide power to a retinal implant device wirelessly by means of electromagnetic induction. Using numerical simulation, we computed the specific absorption rate (SAR) due to the coils which are expected to deliver an average of 50 mW of power to the implanted device. By varying the distance of the coil from the eye and its operation frequency, we are able to find the dependence of SAR due to these parameters. From the result, we found that the maximum SAR value to be below 1 mW/kg. This is several orders of magnitude lower than the safety limits set by the Federal Communications Commission.
引用
收藏
页码:79 / 83
页数:5
相关论文
共 50 条
  • [21] Specific Absorption Rate in the Human Head due to Different Far Field Exposure Sources
    Lazarescu, Catalin
    Dafinescu, Vlad
    David, Valeriu
    PROCEEDINGS OF THE 2012 INTERNATIONAL CONFERENCE AND EXPOSITION ON ELECTRICAL AND POWER ENGINEERING (EPE 2012), 2012, : 683 - 687
  • [22] Investigation of Hand Worn Jewellery Effects on the Specific Absorption Rate in the Human Head and Hand
    Islam, Mohammad Tariqul
    Abidin, Hafizah Zainool
    Faruque, Mohammad Rashed Iqbal
    Misran, Norbahiah
    MECHANICAL AND AEROSPACE ENGINEERING, PTS 1-7, 2012, 110-116 : 4559 - 4563
  • [23] Numerical estimation of local specific absorption rate and temperature elevation in a human head model implanted with visual prosthesis system
    Hikage T.
    Nojima T.
    Yonezawa E.
    Sakai A.
    Osawa K.
    Transactions of Japanese Society for Medical and Biological Engineering, 2017, 55 (01): : 32 - 38
  • [24] Analysis of specific absorption rate and heat transfer in human head due to mobile phones
    Seetharaman, R.
    Tharun, M.
    Gayathri, S.
    Mole, S. S. Sreeja
    Anandan, K.
    MATERIALS TODAY-PROCEEDINGS, 2022, 51 : 2365 - 2374
  • [25] Assessment of specific energy absorption rate (SAR) in the head from a TETRA handset
    Dimbylow, P
    Khalid, M
    Mann, S
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (23): : 3911 - 3926
  • [26] Specific absorption rate distribution in human body caused by PIFA for bluetooth applications
    School of Telecommunication and Network Technology, Beijing University of Posts and Telecommunications, Beijing 100876, China
    不详
    不详
    Dianbo Kexue Xuebao, 2008, 2 (360-364):
  • [27] Specific absorption rate (SAR) revisited for the ellipsoidal model of a human body
    Lattarulo, Francesco
    ENGINEERING ANALYSIS WITH BOUNDARY ELEMENTS, 2014, 49 : 31 - 36
  • [28] Specific absorption rate and path loss in specific body location in heterogeneous human model
    Kurup, Divya
    Joseph, Wout
    Vermeeren, Gunter
    Martens, Luc
    IET MICROWAVES ANTENNAS & PROPAGATION, 2013, 7 (01) : 35 - 43
  • [29] Coils Parallelization: A Strategy for Improving Power Transfer Efficiency and Specific Absorption Rate in Wireless Power Transfer System for Implantable Medical Devices
    Nezaratizadeh, Ali
    Mahmud, Sultan
    Khalifa, Adam
    PROCEEDINGS OF 2024 IEEE WIRELESS POWER TECHNOLOGY CONFERENCE AND EXPO, WPTCE, 2024, : 217 - 222
  • [30] Investigation of reactive power distribution between two coils of inductive power transfer system by Poynting vector analysis
    Liu, Yuan
    Zhang, Kehan
    Hu, Aiguo Patrick
    INTERNATIONAL JOURNAL OF ELECTRICAL POWER & ENERGY SYSTEMS, 2022, 136