Towards real-time 2D/3D registration for organ motion monitoring in Image-Guided Radiation Therapy

被引:0
|
作者
Gendrin, C. [1 ]
Spoerk, J. [1 ]
Bloch, C. [1 ]
Pawiro, S. A. [1 ]
Weber, C. [1 ]
Figl, M. [1 ]
Markelj, P. [2 ]
Pernus, F. [2 ]
Georg, D. [3 ]
Bergmann, H. [1 ]
Birkfellner, W. [1 ]
机构
[1] Med Univ Vienna, Ctr Biomed Engn & Phys, Vienna, Austria
[2] Univ Ljubljana, Fac Elect Engn, Lab Imaging Technol, Ljubljana, Slovenia
[3] Med Univ Vienna, Div Med Phys, Univ Clin Radiotherapy, Vienna, Austria
关键词
Image guided radiation therapy; 2D/3D registration; general purpose computations on graphics hardware;
D O I
10.1117/12.844201
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Nowadays, radiation therapy systems incorporate kV imaging units which allow for the real-time acquisition of intra-fractional X-ray images of the patient with high details and contrast. An application of this technology is tumor motion monitoring during irradiation. For tumor tracking, implanted markers or position sensors are used which requires an intervention. 2D/3D intensity based registration is an alternative, non-invasive method but the procedure must be accelerate to the update rate of the device, which lies in the range of 5 Hz. In this paper we investigate fast CT to a single kV X-ray 2D/3D image registration using a new porcine reference phantom with seven implanted fiducial markers. Several parameters influencing the speed and accuracy of the registrations are investigated. First, four intensity based merit functions, namely Cross-Correlation, Rank Correlation, Mutual Information and Correlation Ratio, are compared. Secondly, wobbled splatting and ray casting rendering techniques are implemented on the GPU and the influence of each algorithm on the performance of 2D/3D registration is evaluated. Rendering times for a single DRR of 20 ms were achieved. Different thresholds of the CT volume were also examined for rendering to find the setting that achieves the best possible correspondence with the X-ray images. Fast registrations below 4 s became possible with an inplane accuracy down to 0.8 mm.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Real-time registration of 3D to 2D ultrasound images for image-guided prostate biopsy
    Gillies, Derek J.
    Gardi, Lori
    De Silva, Tharindu
    Zhao, Shuang-ren
    Fenster, Aaron
    [J]. MEDICAL PHYSICS, 2017, 44 (09) : 4708 - 4723
  • [2] MONITORING LUNG TUMOR MOTION BY REAL-TIME 2D/3D REGISTRATION
    Gendrin, C.
    Weber, C.
    Bloch, C.
    Stock, M.
    Georg, D.
    Figl, M.
    Bergmann, H.
    Birkfellner, W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S453 - S453
  • [3] Real-Time 3D Image-Guided HIFU Therapy
    Ziadloo, Ali
    Vaezy, Shahram
    [J]. 2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 4459 - 4462
  • [4] Implementation of real-time 2D/3D image registration in radiation oncology
    Gendrin, C.
    Weber, C.
    Figl, M.
    Georg, D.
    Bergmann, H.
    Birkfellner, W.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 6, 2009, 25 : 51 - +
  • [5] FLIRT: a software suite for real-time 2D/3D image registration for image guided radiotherapy
    Furtado, H.
    Gendrin, C.
    Spoerk, J.
    Figl, M.
    Georg, D.
    Birkfellner, W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S485 - S485
  • [6] A review of 3D/2D registration methods for image-guided interventions
    Markelj, P.
    Tomazvic, D.
    Likar, B.
    Pernus, F.
    [J]. MEDICAL IMAGE ANALYSIS, 2012, 16 (03) : 642 - 661
  • [7] Real-time 2D/3D registration using kV-MV image pairs for tumor motion tracking in image guided radiotherapy
    Furtado, Hugo
    Steiner, Elisabeth
    Stock, Markus
    Georg, Dietmar
    Birkfellner, Wolfgang
    [J]. ACTA ONCOLOGICA, 2013, 52 (07) : 1464 - 1471
  • [8] Monitoring tumor motion by real time 2D/3D registration during radiotherapy
    Gendrin, Christelle
    Furtado, Hugo
    Weber, Christoph
    Bloch, Christoph
    Figl, Michael
    Pawiro, Supriyanto Ardjo
    Bergmann, Helmar
    Stock, Markus
    Fichtinger, Gabor
    Georg, Dietmar
    Birkfellner, Wolfgang
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 102 (02) : 274 - 280
  • [9] Real-time 3D image guided patient positioning in radiation therapy
    Liu, Dezhi
    Yin, Gongjie
    Li, Shidong
    [J]. THREE-DIMENSIONAL IMAGE CAPTURE AND APPLICATIONS VII, 2006, 6056
  • [10] Real-Time Tumor Motion Tracking in 3D Using Planning 4D CT Images during Image-Guided Radiation Therapy
    Choi, Jang-Hwan
    Lee, Sooyeul
    [J]. ALGORITHMS, 2018, 11 (10):