SUGGESTED DIAGNOSTIC REFERENCE LEVELS FOR PAEDIATRIC X-RAY EXAMINATIONS IN INDIA

被引:6
|
作者
Sonawane, A. U. [1 ]
Kumar, J. V. K. Sunil [1 ]
Singh, Meghraj [1 ]
Pradhan, A. S. [2 ]
机构
[1] Atom Energy Regulatory Board, Radiol Safety Div, Bombay 400094, Maharashtra, India
[2] Ex Bhabha Atom Res Ctr, Bombay, Maharashtra, India
关键词
RADIATION;
D O I
10.1093/rpd/ncq458
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Exposure of children to ionising radiation is considered to carry higher risk than that of adults; therefore a need to suggest diagnostic reference levels (DRLs) for the common paediatric diagnostic X-ray procedures was recognised for the X-ray machines meeting the requirements of the recently implemented Safety Code for Medical Diagnostic X-ray Equipment and Installations in India. Measurements were carried out for entrance surface air kerma (free in air) in conventional paediatric X-ray diagnostic examinations among four age groups: <1, 1-4, 5-9 and 10-15 y. A total of 2240 air kerma measurements at different fixed focus to skin distances were studied for 7 paediatric diagnostic examinations with 11 different projections on 62 X-ray machines installed in 22 selected hospitals in the country. The third quartile values of air kerma per paediatric examination for the age group of 5-9 y were considered as values of paediatric DRLs. The suggested values of DRLs are 0.2 mGy for chest AP/PA, 0.3 mGy for chest LAT, 0.7 mGy for lumber spine AP, 1.3 mGy for lumber spine LAT, 0.3 mGy for thoracic spine AP, 0.6 mGy for thoracic spine LAT, 0.5 mGy for abdomen AP, 0.7 mGy for pelvis AP, 0.6 mGy for skull PA, 0.5 mGy for skull LAT and 0.8 mGy for hip joints AP.
引用
收藏
页码:423 / 428
页数:6
相关论文
共 50 条
  • [1] Diagnostic reference levels for thorax X-ray examinations of paediatric patients
    Kiljunen, T.
    Jaervinen, H.
    Savolainen, S.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (954): : 452 - 459
  • [2] Diagnostic reference levels for X-ray examinations
    Veit, R.
    Guggenberger, R.
    Nosske, D.
    Brix, G.
    [J]. RADIOLOGE, 2010, 50 (10): : 907 - +
  • [3] Local diagnostic reference levels in standard X-ray examinations
    Compagnone, G
    Pagan, L
    Bergamini, C
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 113 (01) : 54 - 63
  • [4] Consideration of diagnostic reference levels for pediatric chest X-ray examinations
    Asada, Yasuki
    Ichikawa, Takuma
    [J]. RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (04) : 382 - 387
  • [5] Consideration of diagnostic reference levels for pediatric chest X-ray examinations
    Yasuki Asada
    Takuma Ichikawa
    [J]. Radiological Physics and Technology, 2019, 12 : 382 - 387
  • [6] Local Diagnostic Reference Levels for X-Ray Examinations in an Australian Tertiary Hospital
    Ng, Curtise K. C.
    Sun, Zhonghua
    Parry, Helen
    Burrage, John
    [J]. JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2014, 4 (02) : 297 - 302
  • [7] Local Diagnostic Reference Levels (LDRLs) for routine X-ray examinations in Morocco
    Talbi, M.
    Tahiri, Z.
    Eddaoui, K.
    El Mansouri, M.
    Nhila, O.
    Benmessaoud, M.
    Ait Ouaggou, I.
    Chakir, E.
    Sebihi, R.
    Khalis, M.
    [J]. RADIOPROTECTION, 2022, 57 (02) : 149 - 155
  • [8] Diagnostic reference levels (DRLs) for routine X-ray examinations in Lorestan province, Iran
    Gholami, M.
    Maziar, A.
    Khosravi, H. R.
    Ebrahimzadeh, F.
    Mayahi, S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2015, 13 (01): : 85 - 90
  • [9] ESTIMATION OF SKIN ENTRANCE DOSES (SEDs) FOR COMMON MEDICAL X-RAY DIAGNOSTIC EXAMINATIONS IN INDIA AND PROPOSED DIAGNOSTIC REFERENCE LEVELS (DRLs)
    Sonawane, A. U.
    Shirva, V. K.
    Pradhan, A. S.
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 138 (02) : 129 - 136
  • [10] Local diagnostic reference levels for some common diagnostic X-ray examinations in Tehran county of Iran
    Toosi, Mohammad Taghi Bahreyni
    Asadinezhad, Mohsen
    [J]. RADIATION PROTECTION DOSIMETRY, 2007, 124 (02) : 137 - 144