Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus

被引:8
|
作者
Wang, Chunxiao [1 ]
Xia, Xueying [2 ]
Tian, Bishan [1 ]
Zhou, Shiyou [1 ]
机构
[1] Sun Yat Sen Univ, Zhongshan Ophthalm Ctr, State Key Lab Ophthalmol, Guangzhou 510060, Guangdong, Peoples R China
[2] Guangdong Women & Children Hosp, Guangzhou 510000, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
HIGH-SPEED; PACHYMETRY;
D O I
10.1155/2015/402925
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective. To compare fourier-domain optical coherence tomography (FD-OCT) and time-domain OCT (TD-OCT) in the determination of thinnest corneal thickness (TCT). Methods. This study included 55 keratoconus patients and 50 healthy volunteers. The RTVue-OCT (FD-OCT) and Visante-OCT (TD-OCT) were used for the measurement of the TCT. Three consecutive scans were performed. The comparison and agreement between the two modalities were analyzed by paired t-test, the Pearson correlation, intraclass correlation coefficient (ICC), and Bland-Altman plots. intraobserver repeatability was analyzed by the intraobserver within-subject standard deviation (S-(w)), coefficient of variation (CV(w)), and ICC. Results. The TCT value of normal corneas was higher by RTVue-OCT (530.4 +/- 19.7 mu m) than Visante-OCT (521.5 +/- 18.3 mu m) (p < 0.001). For keratoconus eyes, the TCT was 425.0 +/- 58.2 mu m and 424.4 +/- 55.7 mu m(difference being 0.6 +/- 10.2 mu m, p = 0.604). Strong correlation (r = 0.938 +/- 0.985) (ICC = 0.915-0.984) was observed between the two OCTs, and each OCT exhibited excellent repeatability in determining the TCT in all subjects (ICC = 0.984-0.994). Conclusions. The values of TCT obtained from RTVue-OCT and Visante-OCT were highly correlated; however, the two values were different. Both OCT instruments exhibited good intraobserver reliability. The existence of systematic differences suggested that the two instruments cannot be used interchangeably.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography for Assessment of Corneal Thickness and Intersession Repeatability
    Prakash, Gaurav
    Agarwal, Amar
    Jacob, Soosan
    Kumar, Dhivya Ashok
    Agarwal, Athiya
    Banerjee, Rumki
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (02) : 282 - 290
  • [2] Time-domain interpolation for Fourier-domain optical coherence tomography
    Zhang, Yudong
    Li, Xiqi
    Wei, Ling
    Wang, Kai
    Ding, Zhihua
    Shi, Guohua
    [J]. OPTICS LETTERS, 2009, 34 (12) : 1849 - 1851
  • [3] Corneal power measurement with Fourier-domain optical coherence tomography
    Tang, Maolong
    Chen, Alex
    Li, Yan
    Huang, David
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (12): : 2115 - 2122
  • [4] Deconvolution techniques in time-domain and Fourier-Domain optical coherence tomography
    Levesque, D.
    Vergnole, S.
    Dufour, M. L.
    Gauthier, B.
    Lamouche, G.
    [J]. REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOL 27A AND 27B, 2008, 975 : 563 - 570
  • [5] MACULAR THICKNESS MEASUREMENTS IN NORMAL EYES WITH TIME-DOMAIN AND FOURIER-DOMAIN OPTICAL COHERENCE TOMOGRAPHY
    Huang, Jingjing
    Liu, Xing
    Wu, Ziqiang
    Xiao, Hui
    Dustin, Laurie
    Sadda, Srinivas
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2009, 29 (07): : 980 - 987
  • [6] Corneal epithelial thickness measured by Fourier-domain optical coherence tomography
    Wang, Ting
    Li, Fengjie
    Wang, Shuting
    Shi, Weiyun
    [J]. SPEKTRUM DER AUGENHEILKUNDE, 2016, 30 (01) : 3 - 9
  • [7] Fourier-Domain Optical Coherence Tomography Imaging in Keratoconus A Corneal Structural Classification
    Sandali, Otman
    El Sanharawi, Mohamed
    Temstet, Cyril
    Hamiche, Taous
    Galan, Alice
    Ghouali, Wajdene
    Goemaere, Isabelle
    Basli, Elena
    Borderie, Vincent
    Laroche, Laurent
    [J]. OPHTHALMOLOGY, 2013, 120 (12) : 2403 - 2412
  • [8] Comparison of ultrasonic pachymetry and Fourier-domain optical coherence tomography for measurement of corneal thickness in dogs with and without corneal disease
    Hoehn, A. L.
    Thomasy, S. M.
    Kass, P. H.
    Horikawa, T.
    Samuel, M.
    Shull, O. R.
    Stewart, K. A.
    Murphy, C. J.
    [J]. VETERINARY JOURNAL, 2018, 242 : 59 - 66
  • [9] Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography
    Jehn-Yu Huang
    Melike Pekmezci
    Stephanie Yaplee
    Shan Lin
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1647 - 1656
  • [10] Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography
    Wylegala, Edward
    Teper, Slawornir
    Nowinska, Anna K.
    Milka, Michal
    Dobrowolski, Dariusz
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (08): : 1410 - 1414