Autocalibration of dew point hygrometer based on integrated semiconductor detector

被引:0
|
作者
Paczesny, D. [1 ]
Tarapata, G. [1 ]
Weremczuk, J. [1 ]
Jachowicz, R. S. [1 ]
机构
[1] Warsaw Univ Technol, Inst Elect Syst, PL-00665 Warsaw, Poland
关键词
sensor autocalibration; dew point hygrometer; humidity measurement;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
The autocalibration procedure of the fast dew point hygrometer based on integrated semiconductor dew detector is presented in the paper. Proposed method is based on detector threshold level setting at 98% of difference between maximal impedance (dry detector) and minimal impedance (fully wet detector). The threshold value was set experimentally by analyzing of detector impedance characteristic in the region close to dew point. Tests of the device proved efficiency of proposed method.
引用
收藏
页码:293 / 296
页数:4
相关论文
共 50 条
  • [41] A new method of super-cooled water recognition in dew point hygrometer by morphology-based image processing
    Weremczuk, J
    Wawrzyniak, Z
    Jachowicz, RS
    SENSORS AND ACTUATORS A-PHYSICAL, 2004, 111 (2-3) : 240 - 244
  • [42] SEMICONDUCTOR SENSOR MIRROR FOR DEW-POINT MEASUREMENT
    DZHAMAGIDZE, SZ
    KOBAKHIDZE, KA
    TODUA, AM
    CHATOV, VA
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1989, 32 (02) : 503 - 504
  • [43] ESTIMATION OF INITIAL AND RESPONSE-TIMES OF LASER DEW-POINT HYGROMETER BY MEASUREMENT SIMULATION
    MATSUMOTO, S
    TOYOOKA, S
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1995, 34 (10): : 5847 - 5852
  • [44] Evaluation of water barrier function of human skin using super fast dew point hygrometer
    Tarapata, Grzegorz
    Weremczuk, Jerzy
    Jachowicz, Ryszard
    MEMEA 2007: SECOND - IEEE INTERNATIONAL WORKSHOP ON MEDICAL MEASUREMENT AND APPLICATIONS, 2007, : 14 - +
  • [45] Development of measurement simulation of the laser dew-point hygrometer using an optical fiber cable
    Matsumoto, S
    Advanced Sensor Systems and Applications II Pt 1and 2, 2004, 5634 : 111 - 122
  • [46] FIELD PERFORMANCE OF DEW-POINT HYGROMETER IN STUDIES OF SOIL-ROOT WATER RELATIONS
    NNYAMAH, JU
    BLACK, TA
    CANADIAN JOURNAL OF SOIL SCIENCE, 1977, 57 (04) : 437 - 444
  • [49] Autocalibration technique based on SIA and integrated multisensor chip
    Ipatov, Andrey
    Abramova, Natalia
    Bratov, Andrey
    TALANTA, 2008, 77 (02) : 581 - 586
  • [50] An ultrahigh-accuracy Miniature Dew Point Sensor based on an Integrated Photonics Platform
    Tao, Jifang
    Luo, Yu
    Wang, Li
    Cai, Hong
    Sun, Tao
    Song, Junfeng
    Liu, Hui
    Gu, Yuandong
    SCIENTIFIC REPORTS, 2016, 6