Maltose Interference-Free Test Strips for Blood Glucose Testing at Point-of-Care: A Laboratory Performance Evaluation

被引:7
|
作者
Ng, Wai-Yoong [1 ]
Tiong, Chiau Chiae [1 ]
Jacob, Edward [1 ]
机构
[1] Singapore Gen Hosp, Dept Pathol, Singapore 169608, Singapore
关键词
ICODEXTRIN;
D O I
10.1089/dia.2010.0095
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Maltose interference is a concern with blood glucose testing at point-of-care. We evaluated a maltose interference-free test strip (with a modified recombinant glucose dehydrogenase-pyrroloquinoline quinone system) for the Accu-Chek (R) Performa glucose meter (Roche Diagnostics, Mannheim, Germany). Methods: Blood specimens (n - 120) sent for clinical laboratory glucose testing were used in assessing performance characteristics, including imprecision, linearity, clinical impact analysis, and method comparison, of the test strips. To evaluate sugar interference, two heparinized blood specimens were spiked with maltose, xylose, and galactose (up to 500 mg/dL) followed by testing with modified Performa, Accutrend (R) (Roche Diagnostics), and Advantage II (Roche Diagnostics) test strips and by the laboratory method. Results: Test strips demonstrated total laboratory coefficients of variation of <7%; within-run coefficients of variation were 2.7-5.4% for blood glucose at 2.5-19.7 mmol/L. Clarke Error Grid analysis of the 120 results (0.8-27.6 mmol/L) showed all values to be within critical clinical limits. Comparison with laboratory results gave 0.960 correlation (Spearman's r(2)) with a Deming regression y (Performa) 0.95x (laboratory) - 0.11 mmol/L (SEy vertical bar x0.06 mmol/L). A slight negative bias (-0.5 mmol/L) was demonstrated with the Bland-Altman difference plot. Maltose (up to 13.9 mmol/L) and xylose (33.3 mmol/L) had no effect, but galactose (2.2 mmol/L) showed interference. The sugars also affected test strips for Advantage II but not Accutrend glucose meters. With International Organization for Standardization ISO 15197:2003 criteria, 99% of the 120 results determined by the test strips were within the minimal acceptable performance; only one of 106 (5.9 mmol/L) was >20% from the laboratory result. Conclusions: The modified and improved Performa test strips were not affected by maltose and xylose. They meet ISO 15197:2003 requirements with a slight bias (-0.5 mmol/L) compared to the laboratory method.
引用
收藏
页码:889 / 893
页数:5
相关论文
共 50 条
  • [21] Smartphone-Based Point-of-Care Cholesterol Blood Test Performance Evaluation Compared with a Clinical Diagnostic Laboratory Method
    Yun, Kyongsik
    Choi, Jaekyu
    Song, In-Uk
    Chung, Yong-An
    APPLIED SCIENCES-BASEL, 2019, 9 (16):
  • [22] Evaluation of a point-of-care blood glucose monitor in healthy goats
    Quandt, Jane E.
    Barletta, Michele
    Cornell, Karen K.
    Giguere, Steeve
    Hofmeister, Erik H.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2018, 28 (01) : 45 - 53
  • [23] Evaluating Laboratory Performance on Point-of-Care Glucose Testing with Six Sigma Metric for 151 Institutions in China
    Fei, Yang
    Wang, Wei
    He, Falin
    Zhong, Kun
    Wang, Zhiguo
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (10) : 745 - 754
  • [24] Point-of-care blood gas analysers: a performance evaluation
    St-Louis, P
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 139 - 144
  • [25] Blood glucose testing at the point-of-care - Clinical experience with the HemoCue B
    Koschinsky, T
    Heidkamp, P
    Reinauer, H
    DIABETES, 1998, 47 : A401 - A401
  • [26] The Impact of Inpatient Point-Of-Care Blood Glucose Quality Control Testing
    Corl, Dawn E.
    Yin, Tom S.
    Hoofnagle, Andrew N.
    Whitney, JoAnne D.
    Hirsch, Irl B.
    Wisse, Brent E.
    JOURNAL FOR HEALTHCARE QUALITY, 2012, 34 (04) : 24 - 32
  • [27] Commentary on Impact of Cyanokit Administration on Point-of-Care Glucose and Laboratory Test Accuracy
    Wang, Ping
    CLINICAL CHEMISTRY, 2022, 68 (04) : 510 - 510
  • [28] Point-of-care testing .3. Determining acceptability of blood glucose meters - Evaluating a blood glucose testing system
    Voss, EM
    Bina, DM
    McNeil, LD
    Johnson, ML
    Cembrowkski, GS
    LABORATORY MEDICINE, 1996, 27 (10) : 679 - 682
  • [29] A combined laboratory and field evaluation of the Cholestech LDX and CardioChek PA point-of-care testing lipid and glucose analysers
    Whitehead, Simon J.
    Ford, Clare
    Gama, Rousseau
    ANNALS OF CLINICAL BIOCHEMISTRY, 2014, 51 (01) : 54 - 67
  • [30] Performance of funded point-of-care capillary blood glucose meters at altitude
    Miller, Steven C. M.
    NEW ZEALAND MEDICAL JOURNAL, 2014, 127 (1394) : 97 - 101