Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE

被引:30
|
作者
Schroder, David T. [2 ]
Kelly, Natalie H. [1 ]
Wright, Timothy M. [1 ]
Parks, Michael L. [2 ]
机构
[1] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
[2] Hosp Special Surg, Adult Reconstruct & Joint Replacement Div, New York, NY 10021 USA
关键词
TOTAL HIP; POLYETHYLENE;
D O I
10.1007/s11999-010-1552-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Highly crosslinked UHMWPE is associated with increased wear resistance in hip simulator and clinical studies. Laboratory and case studies, however, have described rim fracture in crosslinked acetabular liners. Controversy exists, therefore, on the relative merits of crosslinked liners over conventional liners in terms of wear performance versus resistance to fatigue cracking. Questions/purposes We asked whether crosslinked liners would show less surface damage than conventional liners but would be more susceptible to fatigue damage. Methods We examined 36 conventional UHMWPE and 39 crosslinked UHMWPE retrieved implants with similar patient demographics and identical design for evidence of wear damage, including articular surface damage, impingement, screw-hole creep, and rim cracks. Results We observed no difference in wear damage scores for the two liners. Conventional liners more frequently impinged but were more often elevated with smaller head sizes. We observed creep in approximately 70% of both types of liners. Incipient rim cracks were found in five crosslinked liners, and one liner had a rim fracture. Only one conventional liner had an incipient rim crack. Conclusions Contrary to our expectation, damage was similar between crosslinked and conventional UHMWPE liners. Moreover, the 15% occurrence (six of 39) of incipient or complete fractures in crosslinked liners as compared with a 3% occurrence (one of 36) in conventional liners may have implications for the long-term performance of crosslinked liners. Longer-term studies will be necessary to establish the fate of rim cracks and thus the overall clinical fatigue performance of crosslinked liners.
引用
收藏
页码:387 / 394
页数:8
相关论文
共 50 条
  • [21] The wear of highly crosslinked UHMWPE in the presence of abrasive particles: hip and knee simulator studies
    Laurent, MP
    Blanchard, CR
    Yao, JQ
    Johnson, TS
    Gilbertson, LN
    Swarts, DF
    Crowninshield, RD
    CROSSLINKED AND THERMALLLY TREATED ULTRA-HIGH MOLECULAR WEIGHT POLYETHYLENE FOR JOINT REPLACEMENTS, 2004, 1445 : 86 - 103
  • [22] Notched stress-strain behavior of a conventional and a sequentially annealed highly crosslinked UHMWPE
    Sobieraj, Michael C.
    Kurtz, Steven M.
    Wang, A.
    Manley, Michael M.
    Rimnac, Clare M.
    BIOMATERIALS, 2008, 29 (35) : 4575 - 4583
  • [23] Acetabular UHMWPE Survival and Wear Changes With Different Manufacturing Techniques
    Meding, John B.
    Keating, E. Michael
    Davis, Kenneth E.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 405 - 411
  • [24] More wear-resistant and ductile UHMWPE composite prepared by the addition of radiation crosslinked UHMWPE powder
    Wang, Honglong
    Xu, Lu
    Zhang, Mingxing
    Li, Rong
    Xing, Zhe
    Hu, Jiangtao
    Wang, Mouhua
    Wu, Guozhong
    JOURNAL OF APPLIED POLYMER SCIENCE, 2017, 134 (13)
  • [25] Vitamin E diffused, highly crosslinked UHMWPE: a review
    Oral, Ebru
    Muratoglu, Orhun K.
    INTERNATIONAL ORTHOPAEDICS, 2011, 35 (02) : 215 - 223
  • [26] Phase transformation in explanted highly crystalline UHMWPE acetabular cups and debris after in vivo wear
    Reggiani, M
    Tinti, A
    Taddei, P
    Visentin, M
    Stea, S
    De Clerico, M
    Fagnano, C
    JOURNAL OF MOLECULAR STRUCTURE, 2006, 785 (1-3) : 98 - 105
  • [27] Vitamin E diffused, highly crosslinked UHMWPE: a review
    Ebru Oral
    Orhun K. Muratoglu
    International Orthopaedics, 2011, 35 : 215 - 223
  • [28] Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners
    Bradford, L
    Baker, DA
    Graham, J
    Chawan, A
    Ries, MD
    Pruitt, LA
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06): : 1271 - 1282
  • [29] Comparison of Polyethylene Wear between Highly Cross linked and Annealed UHMWPE and Conventional UHMWPE against Ceramic Heads in Total Hip Arthroplasty
    Sato, Taishi
    Nakashima, Yasuharu
    Akiyama, Mio
    Yamamoto, Takuaki
    Mawatari, Taro
    Itokawa, Takashi
    Ohishi, Masanobu
    Motomura, Goro
    Hirata, Masanobu
    Iwamoto, Yukihide
    BIOCERAMICS 24, 2013, 529-530 : 279 - 284
  • [30] High crystallinity makes excellent wear resistance in crosslinked UHMWPE
    Wang, Zirui
    Li, Quan
    Chen, Xianchun
    Zhang, Qin
    Wang, Ke
    POLYMER, 2023, 280