Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE

被引:30
|
作者
Schroder, David T. [2 ]
Kelly, Natalie H. [1 ]
Wright, Timothy M. [1 ]
Parks, Michael L. [2 ]
机构
[1] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
[2] Hosp Special Surg, Adult Reconstruct & Joint Replacement Div, New York, NY 10021 USA
关键词
TOTAL HIP; POLYETHYLENE;
D O I
10.1007/s11999-010-1552-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Highly crosslinked UHMWPE is associated with increased wear resistance in hip simulator and clinical studies. Laboratory and case studies, however, have described rim fracture in crosslinked acetabular liners. Controversy exists, therefore, on the relative merits of crosslinked liners over conventional liners in terms of wear performance versus resistance to fatigue cracking. Questions/purposes We asked whether crosslinked liners would show less surface damage than conventional liners but would be more susceptible to fatigue damage. Methods We examined 36 conventional UHMWPE and 39 crosslinked UHMWPE retrieved implants with similar patient demographics and identical design for evidence of wear damage, including articular surface damage, impingement, screw-hole creep, and rim cracks. Results We observed no difference in wear damage scores for the two liners. Conventional liners more frequently impinged but were more often elevated with smaller head sizes. We observed creep in approximately 70% of both types of liners. Incipient rim cracks were found in five crosslinked liners, and one liner had a rim fracture. Only one conventional liner had an incipient rim crack. Conclusions Contrary to our expectation, damage was similar between crosslinked and conventional UHMWPE liners. Moreover, the 15% occurrence (six of 39) of incipient or complete fractures in crosslinked liners as compared with a 3% occurrence (one of 36) in conventional liners may have implications for the long-term performance of crosslinked liners. Longer-term studies will be necessary to establish the fate of rim cracks and thus the overall clinical fatigue performance of crosslinked liners.
引用
收藏
页码:387 / 394
页数:8
相关论文
共 50 条
  • [1] Wear of retrieved UHMWPE hip liners
    Lee, KY
    Kim, SY
    Shin, D
    Lee, S
    Kim, S
    YONSEI MEDICAL JOURNAL, 2004, 45 (02) : 293 - 299
  • [2] Highly Crosslinked Polyethylene Improves Wear But Not Surface Damage in Retrieved Acetabular Liners
    Pang, Hee-Nee
    Naudie, Douglas D. R.
    McCalden, Richard W.
    MacDonald, Steven J.
    Teeter, Matthew G.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (02) : 463 - 468
  • [3] The use of white light interferometry to characterize wear in short term retrieved highly crosslinked UHMWPE liners
    Kurtz, SM
    Turner, JL
    Herr, MP
    Hozack, W
    Edidin, AA
    PROCEEDINGS OF THE IEEE 28TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2002, : 109 - 110
  • [4] Wear of shelf-aged UHMWPE acetabular liners
    Lee, KY
    Lee, KH
    WEAR, 1999, 225 : 728 - 733
  • [5] Wear versus Thickness and Other Features of 5-Mrad Crosslinked UHMWPE Acetabular Liners
    Shen, Fu-Wen
    Lu, Zhen
    McKellop, Harry A.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 395 - 404
  • [6] Surface analysis of early retrieved acetabular polyethylene liners - A comparison of conventional and highly crosslinked polyethylenes
    Muratoglu, OK
    Greenbaum, ES
    Bragdon, CR
    Jasty, M
    Freiberg, AA
    Harris, WH
    JOURNAL OF ARTHROPLASTY, 2004, 19 (01): : 68 - 77
  • [7] Progressive Wear Damage Analysis on Retrieved UHMWPE Tibial Implants
    Camacho, N.
    Stafford, S. W.
    Trueba, L., Jr.
    26TH SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE: SBEC 2010, 2010, 32 : 165 - 170
  • [8] Comparison of Wear and Oxidation in Retrieved Conventional and Highly Cross-Linked UHMWPE Tibial Inserts
    Currier, Barbara H.
    Currier, John H.
    Franklin, Katherine J.
    Mayor, Michael B.
    Reinitz, Steven D.
    Van Citters, Douglas W.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (12): : 2349 - 2353
  • [9] Analysis of a retrieved UHMWPE acetabular cup crosslinked in air with 1000 kGy of gamma radiation
    Sugano, N
    Saito, A
    Yamamoto, T
    Nishii, T
    Yau, SS
    Wang, A
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (04) : 828 - 831
  • [10] The effect of minimum load on the fluid uptake and wear of highly crosslinked UHMWPE total hip acetabular components
    Yao, JQ
    Laurent, MP
    Gilbertson, LN
    Crowninshield, RD
    WEAR, 2001, 250 : 140 - 144