Incidence and Predictors of Permanent Pacemaker Requirement after Transcatheter Aortic Valve Implantation with a Self-Expanding Bioprosthesis

被引:55
|
作者
Ferreira, Nuno Dias [1 ]
Caeiro, Daniel [1 ]
Adao, Luis [1 ]
Oliveira, Marco [1 ]
Goncalves, Helena [1 ]
Ribeiro, Jose [1 ]
Teixeira, Madalena [1 ]
Albuquerque, Anibal [1 ]
Primo, Joao [1 ]
Braga, Pedro [1 ]
Simoes, Lino [1 ]
Ribeiro, Vasco Gama [1 ]
机构
[1] Ctr Hosp Vila Nova de Gaia Espinho, Dept Cardiol, P-4434502 Vila Nova De Gaia, Portugal
来源
关键词
aortic stenosis; transcatheter aortic valve implantation; permanent pacemaker; PROGNOSTIC-SIGNIFICANCE; RISK-FACTORS; FOLLOW-UP; REPLACEMENT; STENOSIS; PROSTHESIS; EXPERIENCE; ANATOMY; BLOCK;
D O I
10.1111/j.1540-8159.2010.02870.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Methods: From August 2007 to October 2009, 32 consecutive patients underwent TAVI with the Medtronic CoreValve (MCV) System (Medtronic Inc., Minneapolis, MN, USA). Three patients paced at baseline and two cases of procedure-related mortality were excluded. We analyzed the 12-lead electrocardiogram at baseline, immediately after procedure and at discharge. Requirements for PPM were documented and potential clinical, electrophysiological, echocardiographic, and procedural predictors of PPM requirement were studied. Results: After TAVI, eight patients (29.6%) required PPM implantation due to high-grade atrioventricular (AV) block. The prevalence of left bundle branch block increased from 13.8% to 57.7% directly after implantation (P = 0.001). Need for PPM was correlated to the depth of prosthesis implantation (r = 0.590; P = 0.001). At a cutoff point of 10.1 mm, the likelihood of pacemaker could be predicted with 87.5% sensitivity and 74% specificity and a receiver operator characteristic curve area of 0.86 +/- 0.07 (P = 0.003). Of the seven patients with preexisting right bundle branch block (RBBB), four (57.1%) required PPM implantation after TAVI. Conclusions: High-grade AV block requiring PPM implantation is a common complication following TAVI and could be predicted by a deeper implantation of the prosthesis. Patients with preexisting RBBB also seem to be at risk for the development of high-grade AV block and subsequent pacemaker implantation. (PACE 2010; 1364-1372).
引用
收藏
页码:1364 / 1372
页数:9
相关论文
共 50 条
  • [1] Predictors for permanent pacemaker requirement after transcatheter aortic valve implantation with the CoreValve bioprosthesis
    Jilaihawi, Hasan
    Chin, Derek
    Vasa-Nicotera, Mariuca
    Jeilan, Mohamed
    Spyt, Tomasz
    Ng, G. Andre
    Bence, Johan
    Logtens, Elaine
    Kovac, Jan
    AMERICAN HEART JOURNAL, 2009, 157 (05) : 860 - 866
  • [2] Incidence rate and predictors of permanent pacemaker implantation after transcatheter aortic valve implantation with self-expanding CoreValve prosthesis
    Valeria Calvi
    Sergio Conti
    Giusi Paola Pruiti
    Davide Capodanno
    Euglena Puzzangara
    Donatella Tempio
    Angelo Di Grazia
    Gian Paolo Ussia
    Corrado Tamburino
    Journal of Interventional Cardiac Electrophysiology, 2012, 34 : 189 - 195
  • [3] Incidence rate and predictors of permanent pacemaker implantation after transcatheter aortic valve implantation with self-expanding CoreValve prosthesis
    Conti, S.
    Pruiti, G. P.
    Capodanno, D.
    Puzzangara, E.
    Di Grazia, A.
    Ussia, G. P.
    Tamburino, C.
    Calvi, V.
    EUROPEAN HEART JOURNAL, 2011, 32 : 940 - 940
  • [4] Incidence rate and predictors of permanent pacemaker implantation after transcatheter aortic valve implantation with self-expanding CoreValve prosthesis
    Calvi, Valeria
    Conti, Sergio
    Pruiti, Giusi Paola
    Capodanno, Davide
    Puzzangara, Euglena
    Tempio, Donatella
    Di Grazia, Angelo
    Ussia, Gian Paolo
    Tamburino, Corrado
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2012, 34 (02) : 189 - 195
  • [5] Permanent Pacemaker Requirement after Transcatheter Aortic Valve Implantation
    Maier, Robert M.
    Hoedl, Ronald M.
    Stoschitzky, Gergana
    Watzinger, Norbert
    Blazek, Stephan
    Paetzold, Dieter
    Obenwalder, Peter J.
    Pieske, Burkert M.
    Luha, Olev
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (13) : B115 - B115
  • [6] Electrocardiographic and further predictors for permanent pacemaker requirement after transcatheter aortic valve implantation
    Erkapic, Damir
    Kim, Won K.
    Weber, Michael
    Moellmann, Helge
    Berkowitsch, Alexander
    Zaltsberg, Sergey
    Pajitnev, Dmitri J.
    Rixe, Johannes
    Neumann, Thomas
    Kuniss, Malte
    Sperzel, Johannes
    Hamm, Christian W.
    Pitschner, Heinz F.
    EUROPACE, 2010, 12 (08): : 1188 - 1190
  • [7] Electrocardiographic and Imaging Predictors for Permanent Pacemaker Requirement after Transcatheter Aortic Valve Implantation
    Koos, Ralf
    Mahnken, Andreas Horst
    Aktug, Oemer
    Dohmen, Guido
    Autschbach, Ruediger
    Marx, Nikolaus
    Hoffmann, Rainer
    JOURNAL OF HEART VALVE DISEASE, 2011, 20 (01): : 83 - 90
  • [8] Incidence, predictors, and implications of permanent pacemaker requirement after transcatheter aortic valve replacement
    Munoz-Garcia, M.
    Munoz-Garcia, E.
    Munoz Garcia, A. J. Antonio Jesus
    Ruiz-Salas, A.
    Medina-Palomo, C.
    Garcia-Pinilla, J. M.
    Jimenez-Navarro, M. F.
    Alonso-Briales, J. H.
    Hernandez-Garcia, J. M.
    Gomez-Doblas, J. J.
    EUROPEAN JOURNAL OF HEART FAILURE, 2022, 24 : 218 - 219
  • [9] Incidence, Predictors, and Implications of Permanent Pacemaker Requirement After Transcatheter Aortic Valve Replacement
    Sammour, Yasser
    Krishnaswamy, Amar
    Kumar, Arnav
    Puri, Rishi
    Tarakji, Khaldoun G.
    Bazarbashi, Najdat
    Harb, Serge
    Griffin, Brian
    Svensson, Lars
    Wazni, Oussama
    Kapadia, Samir R.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (02) : 115 - 134
  • [10] Predictors of Permanent Pacemaker Implantation After Transcatheter Aortic Valve Implantation
    Fateh-Moghadam, Suzanne
    Pachlhofer, Amelie
    Steeg, Martin
    Mueller, Iris
    Doernberger, Volker
    Doernberger, Volker
    Gawaz, Meinrad
    Bocksch, Wolfgang
    CIRCULATION, 2014, 130