Dose verification for respiratory-gated volumetric modulated arc therapy

被引:36
|
作者
Qian, Jianguo [1 ]
Xing, Lei [1 ]
Liu, Wu [1 ]
Luxton, Gary [1 ]
机构
[1] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 15期
关键词
ION-CHAMBER ARRAY; CONE-BEAM CT; QUALITY-ASSURANCE; PHOTON BEAMS; DELIVERY; RADIOTHERAPY; IMRT; RECONSTRUCTION;
D O I
10.1088/0031-9155/56/15/013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel commercial medical linac system (TrueBeam (TM), Varian Medical Systems, Palo Alto, CA) allows respiratory-gated volumetric modulated arc therapy (VMAT), a new modality for treating moving tumors with high precision and improved accuracy by allowing for regular motion associated with a patient's breathing during VMAT delivery. The purpose of this work is to adapt a previously-developed dose reconstruction technique to evaluate the fidelity of VMAT treatment during gated delivery under clinic-relevant periodic motion related to patient breathing. A Varian TrueBeam system was used in this study. VMAT plans were created for three patients with lung or pancreas tumors. Conventional 6 and 15 MV beams with flattening filter and high-dose-rate 10 MV beams with no flattening filter were used in these plans. Each patient plan was delivered to a phantom first without gating and then with gating for three simulated respiratory periods (3, 4.5 and 6 s). Using the adapted log-file-based dose reconstruction procedure supplemented with ion chamber array (Seven29 (TM), PTW, Freiburg, Germany) measurements, the delivered dose was used to evaluate the fidelity of gated VMAT delivery. Comparison of Seven29 measurements with and without gating showed good agreement with gamma-index passing rates above 99% for 1%/1 mm dose accuracy/distance-to-agreement criteria. With original plans as reference, gamma-index passing rates were 100% for the reconstituted plans (1%/1 mm criteria) and 93.5-100% for gated Seven29 measurements (3%/3 mm criteria). In the presence of leaf error deliberately introduced into the gated delivery of a pancreas patient plan, both dose reconstruction and Seven29 measurement consistently indicated substantial dosimetric differences from the original plan. In summary, a dose reconstruction procedure was demonstrated for evaluating the accuracy of respiratory-gated VMAT delivery. This technique showed that under clinical operation, the TrueBeam system faithfully realized treatment plans with gated delivery. This methodology affords a useful tool for machine- and patient-specific quality assurance of the newly available respiratory-gated VMAT.
引用
收藏
页码:4827 / 4838
页数:12
相关论文
共 50 条
  • [1] Pre-clinical evaluation of respiratory-gated delivery of volumetric modulated arc therapy with RapidArc
    Nicolini, Giorgia
    Vanetti, Eugenio
    Clivio, Alessandro
    Fogliata, Antonella
    Cozzi, Luca
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12): : N347 - N357
  • [2] Stereotactic body radiation therapy using a respiratory-gated volumetric-modulated arc therapy technique for small hepatocellular carcinoma
    Jeong, Yuri
    Jung, Jinhong
    Cho, Byungchul
    Kwak, Jungwon
    Jeong, Chiyoung
    Kim, Jong Hoon
    Park, Jin-hong
    Kim, So Yeon
    Shim, Ju Hyun
    Kim, Kang Mo
    Lim, Young-Suk
    Lee, Han Chu
    Yoon, Sang Min
    [J]. BMC CANCER, 2018, 18
  • [3] Stereotactic body radiation therapy using a respiratory-gated volumetric-modulated arc therapy technique for small hepatocellular carcinoma
    Yuri Jeong
    Jinhong Jung
    Byungchul Cho
    Jungwon Kwak
    Chiyoung Jeong
    Jong Hoon Kim
    Jin-hong Park
    So Yeon Kim
    Ju Hyun Shim
    Kang Mo Kim
    Young-Suk Lim
    Han Chu Lee
    Sang Min Yoon
    [J]. BMC Cancer, 18
  • [4] Respiratory-Gated Volumetric Modulated Arc Therapy for Lung SBRT with An Elekta Digital Accelerator: A Pre-Clinical Evaluation
    Cui, Guoqiang
    Housley, David
    Chen, Fan
    Mehta, Vivek
    Shepard, David
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3783 - 3783
  • [5] Stereotactic Body Radiation Therapy (SBRT) Using Respiratory-Gated Volumetric Modulated Arc Therapy (VMAT) with Flattening Filter Free (FFF)
    Qian, J.
    Luxton, G.
    Xing, L.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [6] Dual Gated Volumetric Modulated Arc Therapy
    Wu, J.
    Fahimian, B.
    Wu, H.
    Xing, L.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3909 - 3909
  • [7] Dosimetric planning study of respiratory-gated volumetric modulated arc therapy for early-stage lung cancer with stereotactic body radiation therapy
    Xhaferllari, Ilma
    Chen, Jeff Z.
    MacFarlane, Michael
    Yu, Edward
    Gaede, Stewart
    [J]. PRACTICAL RADIATION ONCOLOGY, 2015, 5 (03) : 156 - 161
  • [8] Stereotactic Body Radiation Therapy (SBRT) Using Respiratory-Gated Volumetric Modulated Arc Therapy (VMAT) With Flattening Filter Free (FFF) Beams
    Qian, J.
    Luxton, G.
    Mok, E.
    Wang, L.
    Maxim, P.
    Kielar, K.
    Atwood, T.
    Xing, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S851 - S852
  • [9] On-Line Dosimetric Verification of Respiratory Gated Volumetric Modulated Arc Therapy Using the Electronic Portal Imaging Device
    Schaly, B.
    Xhaferllari, I.
    Gaede, S.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3469 - 3470
  • [10] Dual-gated volumetric modulated arc therapy
    Fahimian, Benjamin
    Wu, Junqing
    Wu, Huanmei
    Geneser, Sarah
    Xing, Lei
    [J]. RADIATION ONCOLOGY, 2014, 9