Design and evaluation of a reflectance diffuse optical tomography system

被引:16
|
作者
Kazanci, Hoseyin Ozgur [1 ]
Mercan, Tanju [1 ]
Canpolat, Murat [1 ]
机构
[1] Akdeniz Univ, Sch Med, Dept Biophys, Biomed Opt Res Unit, TR-07058 Antalya, Turkey
关键词
Reflectance diffuse optical tomography; Optode calibrations; Imaging systems; Medical and biological imaging; BRAIN ACTIVATION; SPATIAL-RESOLUTION; CALIBRATION;
D O I
10.1007/s11082-014-9910-6
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have designed a continuous wave, back-reflection diffuse optical tomography system; and developed a new practical calibration method including both optode efficiency and positional error corrections. System design, data acquisition and calibration protocols are described in detail. Monte Carlo (MC) simulations of photon distribution for tissue phantoms have been used to obtain the weight matrix to be used in the Rytov approximation to the photon diffusion equation. The system has been evaluated by acquiring data from a tissue phantom with a background scattering coefficient () of and absorption coefficient ( of . An inclusion made of 1 % Intralipid and indocyanine green with and was placed at a 2 cm depth from the tissue phantom surface. After calibration, the average value of the measurements over source-detector pairs at the same distance for each neighborhood was calculated. Perturbation data were obtained by subtracting the average data from the measurements with the same source-detector separation. In the reconstruction, weight matrixes obtained from MC simulation for to were used. The Depth Compensation Algorithm was used in the Tikhonov regularization to identify the location of the inclusion correctly in the reconstruction.
引用
收藏
页码:257 / 265
页数:9
相关论文
共 50 条
  • [1] Design and evaluation of a reflectance diffuse optical tomography system
    Hüseyin Özgür Kazanci
    Tanju Mercan
    Murat Canpolat
    Optical and Quantum Electronics, 2015, 47 : 257 - 265
  • [2] Design and evaluation of a continuous-wave diffuse optical tomography system
    Siegel, AM
    Marota, JJA
    Boas, DA
    OPTICS EXPRESS, 1999, 4 (08): : 287 - 298
  • [3] Design and evaluation of a hybrid photoacoustic tomography and diffuse optical tomography system for breast cancer detection
    Xi, Lei
    Li, Xiaoqi
    Yao, Lei
    Grobmyer, Stephen
    Jiang, Huabei
    MEDICAL PHYSICS, 2012, 39 (05) : 2584 - 2594
  • [4] Design of an Advanced Time-Domain Diffuse Optical Tomography System
    Mo, Weirong
    Chen, Nanguang
    IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2010, 16 (03) : 581 - 587
  • [5] Evaluation of a multi-layer diffuse reflectance spectroscopy system using optical phantoms
    Fredriksson, Ingemar
    Saager, Rolf B.
    Durkin, Anthony J.
    Stromberg, Tomas
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES X, 2017, 10056
  • [6] Reflectance diffuse optical tomography: Its application to human brain mapping
    Ueda, Y
    Yamanaka, T
    Yamashita, D
    Suzuki, T
    Ohmae, E
    Oda, M
    Yamashita, Y
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS & EXPRESS LETTERS, 2005, 44 (37-41): : L1203 - L1206
  • [7] Preliminary study on human brain imaging by reflectance diffuse optical tomography
    Ueda, Y
    Yamanaka, T
    Yamashita, D
    Suzuki, T
    Oda, M
    Yamashita, Y
    2005 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, 2005, : 392 - 393
  • [8] Imaging of breast cancer using reflectance diffuse optical tomography (RDOT)
    Yoshizawa, Nobuko
    Asano, Yuko
    Yoshimoto, Kenji
    Wada, Hiroko
    Ohmae, Etsuko
    Ueda, Yukio
    Mimura, Tetsuya
    Sakahara, Harumi
    Koizumi, Kei
    Goshima, Satoshi
    Ogura, Hiroyuki
    OPTICAL REVIEW, 2025,
  • [9] Optical clearing of melanoma in vivo: characterization by diffuse reflectance spectroscopy and optical coherence tomography
    Pires, Layla
    Demidov, Valentin
    Vitkin, I. Alex
    Bagnato, Vanderlei
    Kurachi, Cristina
    Wilson, Brian C.
    JOURNAL OF BIOMEDICAL OPTICS, 2016, 21 (08)
  • [10] Design and implementation of a multifrequency near-infrared diffuse optical tomography system
    Gulsen, G
    Xiong, B
    Birgul, O
    Nalcioglu, O
    JOURNAL OF BIOMEDICAL OPTICS, 2006, 11 (01)