Intra-atrial ECG is not a reliable method for positioning left internal jugular vein catheters

被引:49
|
作者
Schummer, W
Herrmann, S
Schummer, C
Funke, F
Steenbeck, J
Fuchs, J
Uhlig, T
Reinhart, K
机构
[1] Univ Jena, Dept Anaesthesiol & Intens Care Med, D-07743 Jena, Germany
[2] Univ Jena, Dept Neonatol & Pediat Intens Care Med, D-07743 Jena, Germany
[3] Univ Jena, Inst Psychol, D-07743 Jena, Germany
[4] Univ Jena, Inst Diagnost & Intervent Radiol, D-07743 Jena, Germany
关键词
heart; catheterization; measurement techniques; echocardiography; transoesophageal; monitoring; electrocardiography; veins; jugular;
D O I
10.1093/bja/aeg208
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. ECG guidance is widely used for positioning central venous catheters (CVCs) in the superior vena cava. We noticed a higher incidence of a more perpendicular angle between the catheter tip and the vessel wall after left-sided ECG-guided catheter positioning. To investigate the value of left-sided ECG guidance, we performed this prospective study. Methods. Of 114 patients, 53 were randomized to right and 61 to left internal jugular vein catheterization using a triple lumen catheter. Three methods to ascertain catheter tip position were sequentially applied in each patient, and the insertion depths (ID) obtained using each of the three methods were recorded: (i) ECG guidance with a Seldinger guide wire (ID-A); (ii) ECG guidance with saline 10% used as an exploring electrode (ID-B); (iii) from position ID-B, the catheter was rotated and advanced until all three lumina could be aspirated easily. The catheter was fixed in that position (ID-C). To determine final catheter tip position, intraoperative transoesophageal echocardiography (TOE) and a postoperative chest X-ray (CXR) were performed. Results. The depth of insertion of a catheter using the three methods varied significantly in left-sided (P<0.001), but not in right-sided catheters. Forty-eight of 57 (84%) left-sided CVCs, correctly positioned according to ECG guidance, had to be advanced further to achieve free aspiration through all three lumina. By this stage, five of the catheter tips had been positioned in the upper right atrium as demonstrated by TOE. There were 13 malpositions (23%) after left-sided insertion. In nine catheter malpositions, undetected by ECG guidance, the angle between the catheter tip and the lateral wall of the superior vena cava exceeded 40degrees on CXR. Conclusions. Intra-atrial ECG does not detect the junction between the superior vena cava and right atrium. It is not a reliable method for confirming position of left-sided CVCs. Post-procedural CXRs are recommended for left-sided, but not right-sided CVCs.
引用
收藏
页码:481 / 486
页数:6
相关论文
共 50 条
  • [31] IS RIGHT OR LEFT BETTER? TUNNELED INTERNAL JUGULAR HAEMODIALYSIS CATHETERS
    Saunders, J.
    Aravindan, A. N.
    Gonzalez, N.
    Howlin, K. J.
    Jefferys, A.
    Makris, A.
    Spicer, S. T.
    Suranyi, M. G.
    Cleland, B. G.
    Wong, J.
    NEPHROLOGY, 2007, 12 : A24 - A24
  • [32] Left intra-atrial thrombus simulating a tumour in a case of hypereosinophilia
    Raissouni, Maha
    Badidi, Mehdi
    Bouzerda, Abdelmjid
    El Ouafi, Noha
    Asfalou, Ilyass
    Benyass, Aatif
    Hamani, Ahmed
    SANG THROMBOSE VAISSEAUX, 2011, 23 (09): : 498 - 500
  • [33] INTERNAL JUGULAR VEIN MALPOSITION OF SUBCLAVIAN CATHETERS - PREVENTION, DETECTION, AND CORRECTION
    WAXMAN, BP
    POLGLASE, AL
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1981, 5 (02) : 186 - 186
  • [34] Rupture of a left Internal Jugular Vein A Rare Case
    Impiombato, F. Ambesi
    Gambacorta, D.
    Zocchi, M.
    De Nisi, M. C.
    Rossi, A.
    NEURORADIOLOGY JOURNAL, 2011, 24 (06): : 942 - 945
  • [35] AN IMPROVED TECHNIQUE FOR THE PLACEMENT OF HEMODIALYSIS CATHETERS IN THE INTERNAL JUGULAR-VEIN
    LADOCSI, LT
    COHN, JD
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 1995, 180 (03) : 343 - 345
  • [36] Malposition of right internal jugular vein cannulation into left brachiocephalic vein
    Sanjeev, G.
    Subha, D.
    Divya, D.
    AIN SHAMS JOURNAL OF ANESTHESIOLOGY, 2023, 15 (01)
  • [37] Malposition of right internal jugular vein cannulation into left brachiocephalic vein
    G. Sanjeev
    D. Subha
    D. Divya
    Ain-Shams Journal of Anesthesiology, 15
  • [38] Left atrial endocardial fibrosis and intra-atrial thrombosis - landmarks of left atrial remodeling in rats with spontaneous atrial tachyarrhythmias
    Scridon, Alina
    Tabib, A.
    Barres, C.
    Julien, C.
    Chevalier, P.
    ROMANIAN JOURNAL OF MORPHOLOGY AND EMBRYOLOGY, 2013, 54 (02): : 405 - 411
  • [39] POSITIONING THE PATIENT WITH INTRA-CRANICAL HYPERTENSION - HOW TURNING AND HEAD ROTATION AFFECT THE INTERNAL JUGULAR VEIN
    LIPE, HP
    MITCHELL, PH
    HEART & LUNG, 1980, 9 (06): : 1031 - 1037
  • [40] A RIGHT ATRIAL CATHETER AFTER PUNCTURE OF THE INTERNAL JUGULAR VEIN
    SCHAPS, D
    DRANSMANN, F
    OELERT, H
    ANAESTHESIST, 1985, 34 (12): : 688 - 689