Optimal design of Bessel acoustic lens suitable for optical-resolution photoacoustic microscope with large depth of field

被引:0
|
作者
Chen, Zouhua [2 ]
Song, Tongyang [3 ]
Xu, Wenhao [3 ]
Wei, Jianshuang [4 ,5 ]
Song, Lingfang [6 ]
Song, Xianlin [1 ]
机构
[1] Nanchang Univ, Sch Informat Engn, Nanchang 330031, Jiangxi, Peoples R China
[2] Nanchang Univ, Coll Sci, Nanchang 330031, Jiangxi, Peoples R China
[3] Nanchang Univ, Ji Luan Acad, Nanchang 330031, Jiangxi, Peoples R China
[4] Huazhong Univ Sci & Technol, Britton Chance Ctr Biomed Photon, Wuhan Natl Lab Optoelect, Wuhan 430074, Peoples R China
[5] Huazhong Univ Sci & Technol, Dept Biomed Engn, Minist Educ, Moe Key Lab Biomed Photon, Wuhan 430074, Peoples R China
[6] Nanchang Normal Univ, Nanchang 330031, Jiangxi, Peoples R China
关键词
photoacoustic microscopy; axicon; conical concave acoustic lens; Bessel acoustic beam; large depth of field; Zemax;
D O I
10.1117/12.2592413
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Photoacoustic tomography is a new medical imaging technology with the advantages of high resolution, high contrast and high penetration depth. There are three common photoacoustic imaging methods in practical applications: photoacoustic microscopic imaging (PAM), photoacoustic computed tomography (PACT), photoacoustic tomography (PAE). As an important branch of photoacoustic imaging, photoacoustic microscopy combines high contrast of optical imaging with high resolution of ultrasonic imaging. In photoacoustic microscopy, acousto-optic coupling prism is a very important component, which is usually composed of irregular prism and spherical concave acoustic lens at the bottom. Its function is to carry out optical transmission and ultrasonic detection. The ultrasonic depth of field of spherical concave acoustic lens is limited. In order to overcome this defect, researchers propose to use conical concave acoustic lens to produce Bessel sound beam to realize large depth of field ultrasonic detection. But the conical concave acoustic lens affects laser focusing and imaging. In order to solve this problem, we propose an optimization method to eliminate the influence of conical concave acoustic lens on beam transmission. A calibration mirror is added to the acousto-optic coupling prism with conical concave acoustic lens at the bottom, and the deterioration of the cone concave acoustic lens to the beam transmission is eliminated by optimizing the surface shape and thickness of the calibration mirror by Zemax. The optimization effect is evaluated by analyzing the spot. The simulation results show that the optimization method can eliminate the influence of the conical concave acoustic lens on the beam transmission, make the focal point and the focal point keep the coaxial focus, and improve the detection efficiency of the photoacoustic signal. This work is of theoretical significance for the systematic study of large depth of field photoacoustic microscopy.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] High-Sensitivity Optical-Resolution Photoacoustic Microscopy with an Optical-Acoustic Combiner Based on an Off-Axis Parabolic Acoustic Mirror
    Zhang, Xiang
    Liu, Yang
    Tao, Chao
    Yin, Jie
    Hu, Zizhong
    Yuan, Songtao
    Liu, Qinghuai
    Liu, Xiaojun
    PHOTONICS, 2021, 8 (04)
  • [32] Adaptive Scanning Optical Microscope (ASOM): A multidisciplinary optical microscope design for large field of view and high resolution imaging
    Potsaid, B
    Bellouard, Y
    Wen, JT
    OPTICS EXPRESS, 2005, 13 (17): : 6504 - 6518
  • [33] Reflection-mode optical-resolution photoacoustic microscopy with high detection sensitivity by using a perforated acoustic mirror
    Zhang, Xiang
    Ding, Qiuning
    Qian, Xiaoqin
    Tao, Chao
    Liu, Xiaojun
    APPLIED PHYSICS LETTERS, 2018, 113 (18)
  • [34] Large-field-of-view optical-resolution optoacoustic microscopy using a stationary silicon-photonics acoustic detector
    Harary, Tamar
    Nagli, Michael
    Suleymanov, Nathan
    Goykhman, Ilya
    Rosenthal, Amir
    JOURNAL OF BIOMEDICAL OPTICS, 2024, 29
  • [35] Acoustic resolution photoacoustic microscopy with large area optical ultrasound detection
    Nuster, Robert
    Rupp, Patrick
    Paltauf, Guenther
    OPTO-ACOUSTIC METHODS AND APPLICATIONS IN BIOPHOTONICS IV, 2019, 11077
  • [36] Optical-resolution photoacoustic imaging through thick tissue with a thin capillary as a dual optical-in acoustic-out waveguide
    Simandoux, Olivier
    Stasio, Nicolino
    Gateau, Jerome
    Huignard, Jean-Pierre
    Moser, Christophe
    Psaltis, Demetri
    Bossy, Emmanuel
    APPLIED PHYSICS LETTERS, 2015, 106 (09)
  • [37] Wide-field two-dimensional multifocal optical-resolution photoacoustic-computed microscopy
    Xia, Jun
    Li, Guo
    Wang, Lidai
    Nasiriavanaki, Mohammadreza
    Maslov, Konstantin
    Engelbach, John A.
    Garbow, Joel R.
    Wang, Lihong V.
    OPTICS LETTERS, 2013, 38 (24) : 5236 - 5239
  • [38] Full-field illumination approach with multiple speckle for optical-resolution photoacoustic microscopy (Conference Presentation)
    Poisson, Florian
    Bossy, Emmanuel
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2016, 2016, 9708
  • [39] Large-field-of-view optical resolution photoacoustic microscopy
    Qin, Wei
    Jin, Tian
    Guo, Heng
    Xi, Lei
    OPTICS EXPRESS, 2018, 26 (04): : 4271 - 4278
  • [40] Design of an adaptive scanning optical microscope for simultaneous large field of view and high resolution
    Potsaid, B
    Bellouard, Y
    Wen, JT
    2005 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), VOLS 1-4, 2005, : 460 - 465