Rotational artifacts in on-board cone beam computed tomography

被引:3
|
作者
Ali, E. S. M. [1 ]
Webb, R. [2 ]
Nyiri, B. J. [1 ,3 ]
机构
[1] Ottawa Hosp Canc Ctr, Dept Med Phys, Ottawa, ON, Canada
[2] Elekta, Montreal, PQ, Canada
[3] Univ Ottawa, Dept Radiol, Ottawa, ON, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 04期
关键词
cone beam computed tomography; XVI; image artifacts; image lag; rotational errors; STEREOTACTIC BODY RADIOTHERAPY; SETUP ERRORS; QUALITY-ASSURANCE; CONSEQUENCES; SYSTEM;
D O I
10.1088/0031-9155/60/4/1461
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Rotational artifacts in image guidance systems lead to registration errors that affect non-isocentric treatments and dose to off-axis organs-at-risk. This study investigates a rotational artifact in the images acquired with the on-board cone beam computed tomography system XVI (Elekta, Stockholm, Sweden). The goals of the study are to identify the cause of the artifact, to characterize its dependence on other quantities, and to investigate possible solutions. A 30 cm diameter cylindrical phantom is used to acquire clockwise and counterclockwise scans at five speeds (120 to 360 deg min(-1)) on six Elekta linear accelerators from three generations (MLCi, MLCi2 and Agility). Additional scans are acquired with different pulse widths and focal spot sizes for the same mAs. Image quality is evaluated using a common phantom with an in-house three dimensional contrast transfer function attachment. A robust, operator-independent analysis is developed which quantifies rotational artifacts with 0.02 degrees accuracy and imaging system delays with 3 ms accuracy. Results show that the artifact is caused by mislabelling of the projections with a lagging angle due to various imaging system delays. For the most clinically used scan speed (360 deg min(-1)), the artifact is similar to 0.5 degrees, which corresponds to similar to 0.25 degrees error per scan direction with the standard Elekta procedure for angle calibration. This leads to a 0.5 mm registration error at 11 cm off-center. The artifact increases linearly with scan speed, indicating that the system delay is independent of scan speed. For the most commonly used pulse width of 40 ms, this delay is 34 +/- 1 ms, part of which is half the pulse width. Results are consistent among the three linac generations. A software solution that corrects the angles of individual projections is shown to eliminate the rotational error for all scan speeds and directions. Until such a solution is available from the manufacturer, three clinical solutions are presented, which reduce the rotational error without compromising image quality.
引用
收藏
页码:1461 / 1476
页数:16
相关论文
共 50 条
  • [1] Rotational errors with on-board cone beam computed tomography
    Ali, E. S. M.
    Webb, R.
    Nyiri, B.
    [J]. MEDICAL PHYSICS, 2014, 41 (08) : 8 - 8
  • [2] On-Board Cone Beam Computed Tomography as Radiation Therapy Simulation for Emergent Cases
    Johnson, G.
    Raymond, J. F.
    Moore, R.
    Nottingham, H.
    Quiles, M.
    Kaurin, D. G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E494 - E495
  • [3] Quality assurance of the on-board imager and cone-beam computed tomography systems
    Yoo, S.
    Kim, G.
    Hammoud, R.
    Elder, E.
    Pawlicki, T.
    Guan, H.
    Fox, T.
    Luxton, G.
    Yin, F.
    Munro, P.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2131 - 2132
  • [4] Four-dimensional cone-beam computed tomography using an on-board imager
    Li, Tianfang
    Xing, Lei
    Munro, Peter
    McGuinness, Christopher
    Chao, Ming
    Yang, Yong
    Loo, Bill
    Koong, Albert
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3825 - 3833
  • [5] Motion correction for improved target localization with on-board cone-beam computed tomography
    Li, T
    Schreibmann, E
    Yang, Y
    Xing, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (02): : 253 - 267
  • [6] Assessment of setup accuracy for prostate radiotherapy using on-board imaging and cone beam computed tomography
    Kasibhalta, M
    Yoo, S
    Yin, F
    Godfrey, D
    Oldham, M
    Das, SK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S535 - S535
  • [7] Quantification of metal artifacts on cone beam computed tomography images
    Pauwels, Ruben
    Stamatakis, Harry
    Bosmans, Hilde
    Bogaerts, Ria
    Jacobs, Reinhilde
    Horner, Keith
    Tsiklakis, Kostas
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 : 94 - 99
  • [8] Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography
    Hirschinger, Veronika
    Hanke, Sebastian
    Hirschfelder, Ursula
    Hofmann, Elisabeth
    [J]. JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2015, 76 (02): : 152 - 163
  • [9] Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts
    Freitas, Deborah Queiroz
    Fontenele, Rocharles Cavalcante
    Leandro Nascimento, Eduarda Helena
    Vasconcelos, Taruska Ventorini
    Noujeim, Marcel
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2018, 47 (08)
  • [10] Evaluation of artifacts of cochlear implant electrodes in cone beam computed tomography
    Bevis, Nicholas
    Effertz, Thomas
    Beutner, Dirk
    Gueldner, Christian
    [J]. EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2021, 278 (05) : 1381 - 1386