Lifetime fluorescence apparatus for clinical investigations of tissues

被引:0
|
作者
Marcu, L [1 ]
Fang, QY [1 ]
Papaioannou, T [1 ]
Jo, JA [1 ]
Vaitha, R [1 ]
Shastry, K [1 ]
机构
[1] Cedars Sinai Med Ctr, Dept Surg, MISTI, Los Angeles, CA 90048 USA
关键词
time-resolved fluorescence spectroscopy; intraoperative diagnostics; arterial diseases; fiberoptic probes;
D O I
10.1117/12.501020
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We report on the design, development, and performance of a versatile and mobile fiber-optic based time-resolved fluorescence apparatus for in-vivo spectroscopy of biological tissues. The apparatus is of modular design which allows for facile interfacing with different pulsed light sources and fiber-optic probes. Main features of the apparatus include: 1) fast acquisition of fluorescence decays with high temporal resolution (0.3 ns) for a broad range of emission wavelengths (300 to 850 nm) and lifetimes (0.3 ns to ins), 2) near real time analysis of fluorescence decays via a Laguerre deconvolution technique, and 3) compatibility with the clinical environment (endoscopic procedures or other intraoperative uses). The apparatus is currently employed for in-vivo studies of diseased tissues (cardiovascular and oncological applications).
引用
收藏
页码:40 / 46
页数:7
相关论文
共 50 条
  • [1] Fluorescence Spectroscopy investigations of cutaneous tissues
    Borisova, E.
    Bliznakova, I.
    Momchilov, N.
    Troyanova, P.
    Avramov, L.
    SIX INTERNATIONAL CONFERENCE OF THE BALKAN PHYSICAL UNION, 2007, 899 : 465 - +
  • [2] Investigations on exponential lifetime measurements for fluorescence thermometry
    Fernicola, VC
    Rosso, L
    Galleano, R
    Sun, T
    Zhang, ZY
    Grattan, KTV
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (07): : 2938 - 2943
  • [3] Investigations of surface defects on semiconductor fluorescence lifetime
    Bender, Daniel A.
    Hasselbeck, Michael P.
    Sheik-Bahae, Mansoor
    LASER COOLING OF SOLIDS, 2007, 6461
  • [4] Fluorescence lifetime of normal, benign, and malignant thyroid tissues
    Brandao, Mariana
    Iwakura, Ricardo
    Basilio, Fagne
    Haleplian, Kaique
    Ito, Amando
    Conti de Freitas, Luiz Carlos
    Bachmann, Luciano
    JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (06) : 067003
  • [5] A WORK STATION FOR LABORATORY DATA ACQUISITION - FLUORESCENCE LIFETIME APPARATUS
    FEDERICI, J
    HELMAN, WP
    HUG, GL
    KANE, C
    PATTERSON, LK
    COMPUTERS & CHEMISTRY, 1985, 9 (03): : 171 - 177
  • [6] Frequency-domain fluorescence lifetime imaging for endoscopic clinical cancer photodetection: Apparatus design and preliminary results
    Wagnières G.
    Mizeret J.
    Studzinski A.
    Van Den Bergh H.
    Journal of Fluorescence, 1997, 7 (1) : 75 - 83
  • [7] Non-linear fluorescence lifetime imaging of biological tissues
    Cicchi, Riccardo
    Pavone, Francesco Saverio
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (09) : 2687 - 2697
  • [8] Fluorescence lifetime-based sensing in tissues: a computational study
    Hutchinson, C. L.
    Lakowicz, J. R.
    Sevick-Muraca, E. M.
    Lebedev Physics Institute. Bulletin, 1994, 68 (11PT2):
  • [9] Auto fluorescence lifetime for identifying premalignant and normal oral tissues
    You, C
    Chen, HM
    Chiang, CP
    Chen, CT
    Wang, CY
    APBP 2004: SECOND ASIAN AND PACIFIC RIM SYMPOSIUM ON BIOPHOTONICS, PROCEEDINGS, 2004, : 105 - 106
  • [10] Non-linear fluorescence lifetime imaging of biological tissues
    Riccardo Cicchi
    Francesco Saverio Pavone
    Analytical and Bioanalytical Chemistry, 2011, 400 : 2687 - 2697