Multimodal phantoms for clinical PET/MRI

被引:12
|
作者
Lennie, Eve [1 ]
Tsoumpas, Charalampos [2 ,3 ]
Sourbron, Steven [1 ]
机构
[1] Univ Sheffield, Dept Infect Immun & Cardiovasc Dis, Sheffield, S Yorkshire, England
[2] Univ Leeds, Biomed Imaging Sci Dept, Leeds, W Yorkshire, England
[3] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, Groningen, Netherlands
基金
英国工程与自然科学研究理事会; 英国医学研究理事会;
关键词
Positron emission tomography; Magnetic resonance imaging; Phantoms; Hybrid imaging; PET ATTENUATION CORRECTION; CEREBRAL-BLOOD-FLOW; WHOLE-BODY PET/MR; MOTION CORRECTION; PERFORMANCE EVALUATION; IMAGING PHANTOMS; TAGGED-MRI; TIME; QUANTIFICATION; COMPENSATION;
D O I
10.1186/s40658-021-00408-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Phantoms are commonly used throughout medical imaging and medical physics for a multitude of applications, the designs of which vary between modalities and clinical or research requirements. Within positron emission tomography (PET) and nuclear medicine, phantoms have a well-established role in the validation of imaging protocols so as to reduce the administration of radioisotope to volunteers. Similarly, phantoms are used within magnetic resonance imaging (MRI) to perform quality assurance on clinical scanners, and gel-based phantoms have a longstanding use within the MRI research community as tissue equivalent phantoms. In recent years, combined PET/MRI scanners for simultaneous acquisition have entered both research and clinical use. This review explores the designs and applications of phantom work within the field of simultaneous acquisition PET/MRI as published over the period of a decade. Common themes in the design, manufacture and materials used within phantoms are identified and the solutions they provided to research in PET/MRI are summarised. Finally, the challenges remaining in creating multimodal phantoms for use with simultaneous acquisition PET/MRI are discussed. No phantoms currently exist commercially that have been designed and optimised for simultaneous PET/MRI acquisition. Subsequently, commercially available PET and nuclear medicine phantoms are often utilised, with CT-based attenuation maps substituted for MR-based attenuation maps due to the lack of MR visibility in phantom housing. Tissue equivalent and anthropomorphic phantoms are often developed by research groups in-house and provide customisable alternatives to overcome barriers such as MR-based attenuation correction, or to address specific areas of study such as motion correction. Further work to characterise materials and manufacture methods used in phantom design would facilitate the ability to reproduce phantoms across sites.
引用
收藏
页数:24
相关论文
共 50 条
  • [1] Multimodal phantoms for clinical PET/MRI
    Eve Lennie
    Charalampos Tsoumpas
    Steven Sourbron
    [J]. EJNMMI Physics, 8
  • [2] Bone material analogues for PET/MRI phantoms
    Chandramohan, Dharshan
    Cao, Peng
    Han, Misung
    An, Hongyu
    Sunderland, John J.
    Kinahan, Paul E.
    Laforest, Richard
    Hope, Thomas A.
    Larson, Peder E. Z.
    [J]. MEDICAL PHYSICS, 2020, 47 (05) : 2161 - 2170
  • [3] Multimodal imaging - PET and MRI - in dementia
    Le Jeune, F.
    Ferre, J-C
    [J]. MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2015, 39 (03): : 268 - 278
  • [4] Quantitative imaging of bone material analogues for PET/MRI phantoms
    Chandramohan, Dharshan
    Cao, Peng
    Han, Misung
    An, Hongyu
    Fowler, Kathryn
    Stupic, Karl
    Keenan, Kathryn
    Sunderland, John
    Kinahan, Paul
    Laforest, Richard
    Hope, Thomas
    Larson, Peder
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [5] Precision medicine with multimodal imaging PET/MRI
    Pichler, Bernd
    [J]. JOURNAL OF PHARMACOLOGICAL SCIENCES, 2017, 133 (03) : S17 - S17
  • [6] Clinical PET/MRI
    Mansi, Luigi
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (02) : 240 - 241
  • [7] Nanoparticle-based multimodal PET/MRI probes
    Garcia, Joel
    Tang, Tang
    Louie, Angelique Y.
    [J]. NANOMEDICINE, 2015, 10 (08) : 1343 - 1359
  • [8] MULTIMODAL PET/MRI RADIOMICS AND CLINICAL PARAMETERS FOR OVERALL SURVIVAL PREDICTION IN PATIENTS WITH IDH WILDTYPE GLIOBLASTOMA
    Gutsche, Robin
    Bauer, Elena
    Kocher, Martin
    Werner, Jan-Michael
    Fink, Gereon
    Shah, Nadim
    Langen, Karl-Josef
    Galldiks, Norbert
    Lohmann, Philipp
    [J]. NEURO-ONCOLOGY, 2022, 24 : 184 - 184
  • [9] Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer
    Rausch, Ivo
    Valladares, Alejandra
    Sundar, Lalith Kumar Shiyam
    Beyer, Thomas
    Hacker, Marcus
    Meyerspeer, Martin
    Unger, Ewald
    [J]. EJNMMI PHYSICS, 2021, 8 (01)
  • [10] Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer
    Ivo Rausch
    Alejandra Valladares
    Lalith Kumar Shiyam Sundar
    Thomas Beyer
    Marcus Hacker
    Martin Meyerspeer
    Ewald Unger
    [J]. EJNMMI Physics, 8