Performance characterization of a prototype dual-layer cone-beam computed tomography system

被引:0
|
作者
Stahl, Fredrik [1 ,2 ]
Schafer, Dirk [3 ]
Omar, Artur [4 ,5 ]
van de Haar, Peter [6 ]
van Nijnatten, Fred [6 ]
Withagen, Paul [6 ]
Thran, Axel [3 ]
Hummel, Erik [6 ]
Menser, Bernd [7 ]
Holmberg, Ake [1 ]
Soderman, Michael [1 ,2 ]
Delgado, Anna Falk [1 ,2 ]
Poludniowski, Gavin [4 ,8 ]
机构
[1] Karolinska Univ Hosp, Dept Neuroradiol, Eugeniavaegen 3, S-17176 Stockholm, Sweden
[2] Karolinska Inst, Dept Clin Neurosci, Stockholm, Sweden
[3] Philips Res Hamburg, Hamburg, Germany
[4] Karolinska Univ Hosp, Med Radiat Phys & Nucl Med, Stockholm, Sweden
[5] Karolinska Inst, Dept Oncol Pathol, Stockholm, Sweden
[6] Phillips Healthcare, Image Guided Therapy, Best, Netherlands
[7] Philips Res, Eindhoven, Netherlands
[8] Karolinska Inst, Dept Clin Sci Intervent & Technol, Huddinge, Sweden
基金
欧盟地平线“2020”;
关键词
cone-beam computed tomography; dual-energy CBCT; dual-layer; flat detector; material decomposition; IMAGE-QUALITY; CT; IMPROVE;
D O I
10.1002/mp.15240
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Conventional cone-beam computed tomography CT (CBCT) provides limited discrimination between low-contrast tissues. Furthermore, it is limited to full-spectrum energy integration. A dual-energy CBCT system could be used to separate photon energy spectra with the potential to increase the visibility of clinically relevant features and acquire additional information relevant in a multitude of clinical imaging applications. In this work, the performance of a novel dual-layer dual-energy CBCT (DL-DE-CBCT) C-arm system is characterized for the first time. Methods A prototype dual-layer detector was fitted into a commercial interventional C-arm CBCT system to enable DL-DE-CBCT acquisitions. DL-DE reconstructions were derived from material-decomposed Compton scatter and photoelectric base functions. The modulation transfer function (MTF) of the prototype DL-DE-CBCT was compared to that of a commercial CBCT. Noise and uniformity characteristics were evaluated using a cylindrical water phantom. Effective atomic numbers and electron densities were estimated in clinically relevant tissue substitutes. Iodine quantification was performed (for 0.5-15 mg/ml concentrations) and virtual noncontrast (VNC) images were evaluated. Finally, contrast-to-noise ratios (CNR) and CT number accuracies were estimated. Results The prototype and commercial CBCT showed similar spatial resolution, with a mean 10% MTF of 5.98 cycles/cm and 6.28 cycles/cm, respectively, using a commercial standard reconstruction. The lowest noise was seen in the 80 keV virtual monoenergetic images (VMI) (7.40 HU) and the most uniform images were seen at VMI 60 keV (4.74 HU) or VMI 80 keV (1.98 HU), depending on the uniformity measure used. For all the tissue substitutes measured, the mean accuracy in effective atomic number was 98.2% (SD 1.2%) and the mean accuracy in electron density was 100.3% (SD 0.9%). Iodine quantification images showed a mean difference of -0.1 (SD 0.5) mg/ml compared to the true iodine concentration for all blood and iodine-containing objects. For VNC images, all blood substitutes containing iodine averaged a CT number of 43.2 HU, whereas a blood-only substitute measured 44.8 HU. All water-containing iodine substitutes measured a mean CT number of 2.6 in the VNC images. A noise-suppressed dataset showed a CNR peak at VMI 40 keV and low at VMI 120 keV. In the same dataset without noise suppression applied, a peak in CNR was obtained at VMI 70 keV and a low at VMI 120 keV. The estimated CT numbers of various clinically relevant objects were generally very close to the calculated CT number. Conclusions The performance of a prototype dual-layer dual-energy C-arm CBCT system was characterized. Spatial resolution and noise were comparable with a commercially available C-arm CBCT system, while offering dual-energy capability. Iodine quantifications, effective atomic numbers, and electron densities were in good agreement with expected values, indicating that the system can be used to reliably evaluate the material composition of clinically relevant tissues. The VNC and monoenergetic images indicate a consistent ability to separate clinically relevant tissues. The results presented indicate that the system could find utility in diagnostic, interventional, and radiotherapy planning settings.
引用
收藏
页码:6740 / 6754
页数:15
相关论文
共 50 条
  • [21] Cone-Beam Computed Tomography as an Adjunct to Performance of Percutaneous Cementoplasty of the Acetabulum
    Munk, Peter L.
    Heran, Manraj K.
    Liu, David M.
    Ouellette, Hugue A.
    Clarkson, Paul W.
    [J]. CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2012, 63 (03): : S7 - S10
  • [22] Cone-Beam Computed Tomography (CBCT) to Computed Tomography (CT) Image Translation with Perceptual Loss to Eliminate Cone-Beam Artifacts
    Yoo, S.
    Choi, B.
    Park, I.
    Kim, J.
    Kim, H.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [23] Asymmetric-Filter Cone-Beam Dual-Energy Computed Tomography
    Yamazaki, Yoichi
    Toda, Naohiro
    [J]. 2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3726 - 3728
  • [24] Effect of noise in dual-energy helical cone-beam computed tomography
    Sidky, EY
    Zou, Y
    Pan, XC
    [J]. MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 396 - 402
  • [25] Noise Characterization of a Clinical Flat-Panel Cone-Beam Computed Tomography (CBCT) System
    Vedantham, S.
    Gounis, M.
    van der Bom, M.
    Karellas, A.
    Wakhloo, A.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [26] A moving blocker system for cone-beam computed tomography scatter correction
    Ouyang, Luo
    Song, Kwang
    Wang, Jing
    [J]. MEDICAL PHYSICS, 2013, 40 (07)
  • [27] A Moving Blocker System for Cone-Beam Computed Tomography Scatter Correction
    Ouyang, L.
    Song, K.
    Solberg, T.
    Wang, J.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [28] A moving blocker system for cone-beam computed tomography scatter correction
    Ouyang, Luo
    Song, Kwang
    Solberg, Timothy
    Wang, Jing
    [J]. MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [29] Automatic Classification System for Periapical Lesions in Cone-Beam Computed Tomography
    Andrade Calazans, Maria Alice
    Ferreira, Felipe Alberto B. S.
    Melo Guedes Alcoforado, Maria de Lourdes
    dos Santos, Andrezza
    Pontual, Andrea dos Anjos
    Madeiro, Francisco
    [J]. SENSORS, 2022, 22 (17)
  • [30] Computed gray levels in multislice and cone-beam computed tomography
    Azeredo, Fabiane
    de Menezes, Luciane Macedo
    Enciso, Reyes
    Weissheimer, Andre
    de Oliveira, Rogerio Belle
    [J]. AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2013, 144 (01) : 147 - 155