Technical note: Dosimetric study for the new BEBIG 60Co HDR source used in brachytherapy in water and different media using Monte Carlo N-Particle eXtended code

被引:3
|
作者
Elboukhari, Said [1 ]
Yamni, Khalid [1 ]
Ouabi, Hmad [2 ]
Bouassa, Taoufiq [3 ]
Mlouk, Lahcen Ait [4 ]
机构
[1] Molay Ismail Univ Meknes, Fac Sci, Dept Chem, EMaMePS, Meknes, Morocco
[2] Ctr Oncol ALAZHAR, Rabat, Morocco
[3] Mohammed V Univ Rabat, Fac Sci, Rabat, Morocco
[4] Molay Ismail Univ Meknes, Fac Sci, Meknes, Morocco
关键词
Dosimetry parameters; HDR brachytherapy; Dose rate; Co-60; HEBD working group; AAPM;
D O I
10.1016/j.apradiso.2020.109087
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
This work aims to study the dosimetrc parameters of the new Co-60-source model CoO.A86 used in high dose rate brachytherapy and manufactured by BEBIG (Eckert & Ziegler BEBIG GmbH, Germany). The radial dose function, 2D along&away dose rates and the dose rate constant were investigated in a water phantom. Accordingly, we use the recommendations from the AAPM and ESTRO on dose calculation for high-energy (average energy higher than 50 keV) photon-emitting brachytherapy sources cited in the HEBD working group report. Furthermore, we compared the obtained results with the quoted values in the previous studies. The value of air-kerma strength calculated in this work was 3.030 +/- 0.002 IL13q(-1). Moreover, the radial dose function and 2D along&away dose rates are evaluated in different tissues and compared with the results obtained in the water and we found a notable difference. To reach the goal we have used MCNPX code for simulation.
引用
收藏
页数:10
相关论文
共 17 条
  • [1] EGSnrc Monte Carlo-aided dosimetric studies of the new BEBIG 60Co HDR brachytherapy source
    Anwarul, Islam Mohammad
    Akramuzzaman, Mir Md
    Abu Zakaria, Golam
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2013, 5 (03) : 148 - 156
  • [2] Technical Note: EGSnrc-based dosimetric study of the BEBIG 60Co HDR brachytherapy sources
    Selvam, T. Palani
    Bhola, Subhalaxmi
    MEDICAL PHYSICS, 2010, 37 (03) : 1365 - 1370
  • [3] Monte Carlo dosimetric study of the new BEBIG Co-60HDR source
    Ballester, F
    Granero, D
    Pérez-Calatayud, J
    Casal, E
    Agramunt, S
    Cases, R
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S138 - S138
  • [4] Monte Carlo dosimetric study of the new BEBIG Co-60HDR source
    Perez-Calatayud, J
    Granero, D
    Ballester, F
    Casal, E
    Agramunt, S
    Cases, R
    MEDICAL PHYSICS, 2005, 32 (06) : 1958 - 1958
  • [5] Dosimetric investigation for the new BEBIG cobalt 60 source used in brachytherapy using Monte Carlo N-Particles eXtended code (MCNPX)
    Elboukhari, S.
    Yamni, K.
    Ouabi, H.
    Bouassa, T.
    Mlouk, L. A.
    INTERNATIONAL CONFERENCE ON ADVANCED MATERIALS, MICROSCOPY AND ENERGY (ICAMME) 2019, 2020, 783
  • [6] Monte Carlo dosimetric study of the BEBIG Co-60HDR source
    Ballester, F
    Granero, D
    Pérez-Calatayud, J
    Casal, E
    Agramunt, S
    Cases, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21): : N309 - N316
  • [7] BEBIG 60Co HDR brachytherapy source dosimetric parameters validation using GATE Geant4-based simulation code
    Joya, Musa
    Geraily, Ghazale
    Ghorbani, Mahdi
    Sheikhzadeh, Peyman
    Nedaie, Hassan Ali
    Arani, Mahmud Naraqi
    HELIYON, 2022, 8 (03)
  • [8] Dosimetric comparison between the microSelectron HDR Ir-192 v2 source and the BEBIG Co-60 source for HDR brachytherapy using the EGSnrc Monte Carlo transport code
    Islam, M. Anwarul
    Akramuzzaman, M. M.
    Zakaria, G. A.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (04) : 219 - 225
  • [9] Technical note:: Dosimetric study of a. new Co-60 source used in brachytherapy
    Granero, D.
    Perez-Calatayud, J.
    Ballester, F.
    MEDICAL PHYSICS, 2007, 34 (09) : 3485 - 3488
  • [10] Monte Carlo Dosimetric Study of the GZP 60Co Brachytherapy Source with Stainless Steel Applicator
    Wu, J.
    Wang, P.
    Wang, X.
    Kang, S.
    Li, J.
    Tang, B.
    Orlandini, L. C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E564 - E564