Noise Reduction Using a Theoretically-Exact Algorithm for Helical Cone-Beam Tomography

被引:0
|
作者
Venkataraman, Rajesh [1 ]
Noo, Frederic [2 ]
Kudo, Hiroyuki [3 ]
机构
[1] Univ Utah, Dept Elect & Comp Engn, Salt Lake City, UT 84112 USA
[2] Univ Utah, Dept Radiol, Salt Lake City, UT USA
[3] Univ Tsukuba, Dept Comp Sci, Tsukuba, Ibaraki, Japan
基金
美国国家卫生研究院;
关键词
Computed Tomography; multi-slice CT; cone-beam reconstruction; helical geometry; redundancy; detector configuration; noise reduction; theoretically-exact reconstruction;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper we show that the signal-to-noise ratio of helical cone-beam reconstructions based on Katsevish's formula can be significantly improved for 16-and 32-row scanners, using the observation that Katsevich's formula heavily relies on the concept of pi-lines. A pi-line is any segment of line that connects two source positions separated by less than one helix turn; pi-lines are such that each point within the helix cylinder belongs to one and only one pi-line. Katsevich's formula reconstructs the sought density function at a given point using exclusively the data between the two source positions that define the pi-line through this point. We observe that the pi-line through a given point can change significantly when the point is displaced along a direction of the rotation axis of the scanner, and we use this observation to improve image noise as follows: for CT imaging with a prescribed slice thickness Delta z, we apply Katsevich's formula to achieve reconstruction on slices separated by a distance Delta z/M, where M is a positive integer, then obtain any wanted slice as the average of the M closest reconstructed slices. For Katsevich's formula, proceeding in this way with M large is fundamentally different from reconstructing with M = 1 using pre-smoothing of the data to match the slice thickness target.
引用
收藏
页码:1674 / 1678
页数:5
相关论文
共 50 条
  • [21] Exact Efficient Handling of Interrupted Illumination in Helical Cone-Beam Computed Tomography with Arbitrary Pitch
    Harald Schndube
    Karl Stierstorfer
    Frédéric Noo
    Tsinghua Science and Technology, 2010, 15 (01) : 36 - 43
  • [22] Analysis of image noise with exact cone-beam reconstruction
    Lee, SW
    Wang, G
    RADIOLOGY, 2002, 225 : 253 - 253
  • [23] Exact Efficient Handling of Interrupted Illumination in Helical Cone-Beam Computed Tomography with Arbitrary Pitch
    Schöndube H.
    Stierstorfer K.
    Noo F.
    Tsinghua Science and Technology, 2010, 15 (01) : 36 - 43
  • [24] Noise performance of fast hierarchical 3D backprojection for helical cone-beam tomography
    Brokish, Jeffrey
    Bresler, Yoram
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1860 - 1862
  • [25] Optoelectronic implementation of helical cone-beam computed tomography algorithms
    Madec, Morgan
    Fasquel, Jean-Baptiste
    Uhring, Wilfried
    Joffre, Pascal
    Herve, Yannick
    OPTICAL ENGINEERING, 2008, 47 (05)
  • [26] High-helical-pitch, cone-beam computed tomography
    Silver, MD
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04): : 847 - 855
  • [27] Exact reconstruction of volumetric images in reverse helical cone-beam CT
    Cho, Seungryong
    Xia, Dan
    Pelizzari, Charles A.
    Pan, Xiaochuan
    MEDICAL PHYSICS, 2008, 35 (07) : 3030 - 3040
  • [28] Exact image reconstruction in a helical cone-beam scan with a variable pitch
    Zou, Y
    Pan, XC
    Xia, D
    Wang, G
    Jiang, TZ
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4200 - 4203
  • [29] Exact image reconstruction for selected ROI in helical cone-beam CT
    Pan, XC
    Zou, Y
    Xia, D
    Sidky, E
    Jiang, TZ
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3773 - 3776
  • [30] Physical phantom studies of helical cone-beam CT with exact reconstruction
    Tan, Jun
    Li, H. Harold
    Klein, Eric
    Li, Hua
    Parikh, Parag
    Yang, Deshan