Comparison of different registration methods in cone-beam computed tomography for breast boost radiation therapy

被引:1
|
作者
Lastrucci, A. [1 ]
Fedeli, L. [2 ]
Marciello, L. [3 ]
Serventi, E. [1 ]
Segnini, S. [1 ]
Meucci, F. [2 ]
Bernardi, L. [2 ]
Ricci, R. [1 ]
Marzano, S. [3 ]
机构
[1] Santo Stefano Hosp, Dept Tech Hlth Profess, Radiat Oncol Unit, Azienda USL Toscana Ctr, I-59100 Prato, Italy
[2] Santo Stefano Hosp, Med Phys Unit, Azienda USL Toscana Ctr, I-59100 Prato Pistoia, Italy
[3] Santo Stefano Hosp, Dept Oncol, Radiat Oncol Unit, Azienda USL Toscana Ctr, I-59100 Prato, Italy
关键词
Boost; breast; CBCT; Clipbox; IGRT; Mask; CONSERVING THERAPY; CANCER; RADIOTHERAPY; ELECTRON;
D O I
10.1017/S1460396922000176
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: The aim of this study is to compare patient geometrical uncertainties in the treatment of breast boost three-dimensional conformal radiation therapy (3D-CRT) considering both manual alignment and automatic different registration methods in cone-beam computed tomography (CBCT). Methods: A total of 85 patients were chosen for this study. A total of 254 registrations of CBCT vs planning computed tomography (CT) were retrospectively performed using automatic registration algorithms from Elekta XVI system (Clipbox and Mask) to detect patient setup uncertainties. All CBCTs were also matched manually by three health professionals. Mean shift values obtained with manual registration performed by health professionals were used as reference. Absolute value of difference between automatic algorithm shifts and reference values shifts was collected for each enrolled patient considering the three different spatial directions (x, y ,z), and the magnitude was calculated (delta m for Mask and delta c for Clipbox). Results: Data analysis showed a significant difference in delta m and delta c. t-Test statistics showed a high difference between Mask and Clipbox, in particular mean delta m = (1.3 +/- 0.1) mm and delta c = (3.3 +/- 1.2) mm (p-value <0.0001). Mask algorithm was performed in a very similar way with respect to the reference alignment, and the differences between these two procedures were of the order of 1 mm. Clipbox algorithm showed larger differences with manual registration. Conclusions: These results suggest that the Mask algorithm may be the optimal choice for patient setup verification in clinical practice for breast boost treatment in 3D-CRT.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Stereotactic registration using cone-beam computed tomography
    Carlson, Jonathan Dennis
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2019, 182 : 107 - 111
  • [2] Dose Reduction in Kilovotage Cone-Beam Computed Tomography for Radiation Therapy
    Wang, J.
    Li, T.
    Liang, J.
    Xing, L.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [3] Dose reduction for kilovotage cone-beam computed tomography in radiation therapy
    Wang, Jing
    Li, Tianfang
    Liang, Zhengrong
    Xing, Lei
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11): : 2897 - 2909
  • [4] Quantitative evaluation of a cone-beam computed tomography-planning computed tomography deformable image registration method for adaptive radiation therapy
    Lawson, Joshua D.
    Schreibmann, Eduard
    Jani, Ashesh B.
    Fox, Tim
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (04): : 96 - 113
  • [5] Automated Orientation and Registration of Cone-Beam Computed Tomography Scans
    Anchling, Luc
    Hutin, Nathan
    Huang, Yanjie
    Barone, Selene
    Roberts, Sophie
    Miranda, Felicia
    Gurgel, Marcela
    Al Turkestani, Najla
    Tinawi, Sara
    Bianchi, Jonas
    Yatabe, Marilia
    Ruellas, Antonio
    Prieto, Juan Carlos
    Cevidanes, Lucia
    CLINICAL IMAGE-BASED PROCEDURES, FAIRNESS OF AI IN MEDICAL IMAGING, AND ETHICAL AND PHILOSOPHICAL ISSUES IN MEDICAL IMAGING, CLIP 2023, FAIMI 2023, EPIMI 2023, 2023, 14242 : 43 - 58
  • [6] Cone-beam micro computed tomography dedicated to the breast
    Sarno, Antonio
    Mettivier, Giovanni
    Di Lillo, Francesca
    Cesarelli, Mario
    Bifulco, Paolo
    Russo, Paolo
    MEDICAL ENGINEERING & PHYSICS, 2016, 38 (12) : 1449 - 1457
  • [7] Evaluation of Megavoltage Cone-Beam Computed Tomography Doses in Thoracic Radiation Therapy
    Yan, S.
    Zhu, J.
    Yan, S.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [8] Comparison of Methods for Acceptance and Constancy Testing in Dental Cone-beam Computed Tomography
    Steiding, C.
    Kolditz, D.
    Kalender, W.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2015, 187 (04): : 283 - 290
  • [9] Different setup errors assessed by weekly cone-beam computed tomography on different registration in nasopharyngeal carcinoma treated with intensity-modulated radiation therapy
    Su, Jiqing
    Chen, Wen
    Yang, Huiyun
    Hong, Jidong
    Zhang, Zijian
    Yang, Guangzheng
    Li, Li
    Wei, Rui
    ONCOTARGETS AND THERAPY, 2015, 8 : 2545 - 2553
  • [10] Study to Compare the Effect of Different Registration Methods on Patient Setup Uncertainties in Cone-beam Computed Tomography during Volumetric Modulated Arc Therapy for Breast Cancer Patients
    Mohandass, P.
    Khanna, D.
    Kumar, T. Manoj
    Thiyagaraj, T.
    Saravanan, C.
    Bhalla, Narendra Kumar
    Puri, Abhishek
    JOURNAL OF MEDICAL PHYSICS, 2018, 43 (04) : 207 - 213