NUMERICAL CHARACTERIZATION OF HEMODYNAMICS CONDITIONS NEAR AORTIC VALVE AFTER IMPLANTATION OF LEFT VENTRICULAR ASSIST DEVICE

被引:10
|
作者
Quaini, Annalisa [1 ]
Canic, Suncica [1 ]
Paniagua, David [2 ,3 ]
机构
[1] Univ Houston, Dept Math, Houston, TX 77204 USA
[2] St Lukes Episcopal Hosp, Texas Heart Inst, Dept Cardiol, Houston, TX 77030 USA
[3] Mickael E Debakey VA Med Ctr, Houston, TX 77030 USA
基金
美国国家科学基金会;
关键词
Fluid-structure interaction; structure-structure interaction; monolithic methods; semi-implicit coupling; hemodynamics;
D O I
10.3934/mbe.2011.8.785
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Left Ventricular Assist Devices (LVADs) are implantable mechanical pumps that temporarily aid the function of the left ventricle. The use of LVADs has been associated with thrombus formation next to the aortic valve and close to the anastomosis region, especially in patients in which the native cardiac function is negligible and the aortic valve remains closed. Stagnation points and recirculation zones have been implicated as the main fluid dynamics factors contributing to thrombus formation. The purpose of the present study was to develop and use computer simulations based on a fluid-structure interaction (FSI) solver to study flow conditions corresponding to different strategies in LVAD ascending aortic anastomosis providing a scenario with the lowest likelihood of thrombus formation. A novel FSI algorithm was developed to deal with the presence of multiple structures corresponding to different elastic properties of the native aorta and of the LVAD cannula. A sensitivity analysis of different variables was performed to assess their impact of flow conditions potentially leading to thrombus formation. It was found that the location of the anastomosis closest to the aortic valve (within 4 cm away from the valve) and at the angle of 30 minimizes the likelihood of thrombus formation. Furthermore, it was shown that the rigidity of the dacron anastomosis cannula plays almost no role in generating pathological conditions downstream from the anastomosis. Additionally, the flow analysis presented in this manuscript indicates that compliance of the cardiovascular tissue acts as a natural inhibitor of pathological flow conditions conducive to thrombus formation and should not be neglected in computer simulations.
引用
收藏
页码:785 / 806
页数:22
相关论文
共 50 条
  • [41] Management of a mechanical aortic valve during left ventricular assist device implantation in a previously replaced aortic root
    Tulimat, Tamam
    Osman, Bassam
    Beresian, Jean
    Sfeir, Pierre
    Borgi, Jamil
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2022, 45 (02): : 152 - 154
  • [42] Impact of progressive aortic regurgitation on outcomes after left ventricular assist device implantation
    Gasparovic, Hrvoje
    Jakus, Nina
    Brugts, Jasper J.
    Pouleur, Anne-Catherine
    Timmermans, Philippe
    Rubis, Pawel
    Gaizauskas, Edvinas
    Van Craenenbroeck, Emeline M.
    Barge-Caballero, Eduardo
    Grundmann, Sebastian
    Paolillo, Stefania
    D'Amario, Domenico
    Braun, Oscar O.
    Meyns, Bart
    Droogne, Walter
    Wierzbicki, Karol
    Holcman, Katarzyna
    Planinc, Ivo
    Lovric, Daniel
    Flammer, Andreas J.
    Petricevic, Mate
    Biocina, Bojan
    Lund, Lars H.
    Milicic, Davor
    Ruschitzka, Frank
    Cikes, Maja
    HEART AND VESSELS, 2022, 37 (12) : 1985 - 1994
  • [43] Atrophy of the Heart After Insertion of a Left Ventricular Assist Device and Closure of the Aortic Valve
    Roberts, William C.
    Hall, Shelley A.
    Ko, Jong M.
    McCullough, Peter A.
    Lima, Brian
    AMERICAN JOURNAL OF CARDIOLOGY, 2016, 117 (05): : 878 - 879
  • [44] Percutaneous Transcatheter Closure of the Native Aortic Valve to Treat De Novo Aortic Insufficiency After Implantation of a Left Ventricular Assist Device
    Freed, Benjamin H.
    Paul, Jonathan D.
    Bhave, Nicole M.
    Russo, Mark J.
    Jeevanandam, Valluvan
    Lang, Roberto M.
    Shah, Atman P.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2012, 5 (03) : 358 - 359
  • [45] Left ventricular assist device implantation in patients after left ventricular reconstruction
    Palmen, Meindert
    Braun, Jerry
    Beeres, Saskia L. M. A.
    Klautz, Robert J. M.
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2016, 23 (06) : 979 - 981
  • [46] Aortic valve insufficiency after Impella device insertion that required aortic valve replacement after Heart Mate III left ventricular assist device implantation: a case report.
    Ueda, Kohei
    Yoshitani, Kenji
    Hosotani, Shunsuke
    Hayashi, Hisanori
    Fukushima, Satsuki
    Ohnishi, Yoshihiko
    JOURNAL OF SURGICAL CASE REPORTS, 2021, (10):
  • [47] Valve-in-Ring Transcatheter Aortic Valve Replacement After Left Ventricular Assist Device Therapy
    Chung, Matthew J.
    Ganapathi, Asvin M.
    Vora, Amit N.
    Schroder, Jacob N.
    Kiefer, Todd L.
    Hughes, G. Chad
    ANNALS OF THORACIC SURGERY, 2020, 109 (03): : E163 - E165
  • [48] Addition of Mitral Valve Repair to Left Ventricular Assist Device Implantation
    Taghavi, S.
    Hamad, E.
    Wilson, L.
    Clark, R.
    Jayarajan, S. N.
    Uriel, N.
    Goldstein, D.
    Takayama, H.
    Naka, Y.
    Mangi, A. A.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2012, 31 (04): : S270 - S271
  • [49] Left Ventricular Assist Device Implantation Induced Tricuspid Valve Prolapse
    Bouabdallaoui, Nadia
    Akar, Ramzi Abi
    Ennezat, Pierre Vladimir
    Aissaoui, Nadia
    Jouan, Jerome
    Bricourt, Marie Odile
    Grinda, Jean Michel
    CIRCULATION-HEART FAILURE, 2013, 6 (06) : E71 - E72
  • [50] Tricuspid valve repair with left ventricular assist device implantation: Is it warranted?
    Saeed, Diyar
    Kidambi, Trilokesh
    Shalli, Shanaz
    Lapin, Brittany
    Malaisrie, S. Chris
    Lee, Richard
    Cotts, William G.
    McGee, Edwin C., Jr.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2011, 30 (05): : 530 - 535