Digital sampling setup for calibration of strain gauge bridge calibration unit BN100

被引:0
|
作者
Maslan, Stanislav [1 ]
Sira, Martin [1 ]
Skalicka, Tereza [1 ]
机构
[1] Czech Metrol Inst, Okruzni 31, Brno 63800, Czech Republic
关键词
Impedance; inductive voltage divider; voltage ratio; strain gauge;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes minimalistic, economic, automated, digital sampling calibration setup for calibration of a strain gauge bridge calibration unit BN100 or similar high division ratio dividers. The setup uses impedance divider instead of the IVD as the reference divider. It achieves comparable or better uncertainty for division ratios below 1 mV/V compared to the IVD based methods. For 1 mV/V the expanded uncertainty (k = 2) is 8 nV/V from the input.
引用
收藏
页数:2
相关论文
共 50 条
  • [21] A Reference Merging Unit and Calibration Setup for Sampled Values over Ethernet
    Lehtonen, Tapio
    Hallstrom, Jari
    2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [22] CiNetStrain - Wireless Strain Gauge Network - Calibration and Reliability Measurements
    Hongell, Timo
    Ihalainen, Jukka
    Hakala, Ismo
    Ad Hoc Networks, (ADHOCNETS 2012), 2013, 111 : 275 - 288
  • [23] Digital Calibration of 60 GHz Setup for Use in Power Amplifier Predistortion
    Marsalek, Roman
    Pospisil, Martin
    Gotthans, Tomas
    Urbanec, Tomas
    2019 IEEE 20TH INTERNATIONAL WORKSHOP ON SIGNAL PROCESSING ADVANCES IN WIRELESS COMMUNICATIONS (SPAWC 2019), 2019,
  • [24] Error Analysis and Calibration on Digital Optical Metering Unit
    Zhu, Dongbai
    IECON 2004: 30TH ANNUAL CONFERENCE OF IEEE INDUSTRIAL ELECTRONICS SOCIETY, VOL 2, 2004, : 1555 - 1558
  • [25] Calibration of the DQI (Digital Quartz Inertial Measurements Unit)
    Stave, LP
    PROCEEDINGS OF THE 52ND ANNUAL MEETING - THE INSTITUTE OF NAVIGATION: NAVIGATIONAL TECHNOLOGY FOR THE 3RD MILLENNIUM, 1996, : 607 - 615
  • [26] "Optimal" calibration weights under unit nonresponse in survey sampling
    Andersson, Per Gosta
    SURVEY METHODOLOGY, 2019, 45 (03) : 533 - 542
  • [27] Calibration methods for negative gauge pressure down to-100 kPa
    Bentouati, Djilali
    Durgut, Yasin
    Otal, Pierre
    Plimmer, Mark
    Prazak, Dominik
    Sabuga, Wladimir
    Ehlers, Sven
    Sinir, Ekrem
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2018, 29 (03)
  • [28] CALIBRATION OF STANDARD WATTMETERS USING A CAPACITANCE BRIDGE AND A DIGITAL GENERATOR
    OLDHAM, NM
    PETERSONS, O
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1985, 34 (04) : 521 - 524
  • [29] Theoretical design of calibration beams for strain gauge factor measuring apparatus
    Deng Zubin
    STRAIN, 1998, 34 (03) : 99 - 107
  • [30] Examination of Systematic Bench Errors for Calibration of a Strain-Gauge Balance
    A. A. Bol’shakova
    V. S. Volobuev
    A. R. Gorbushin
    V. V. Petronevich
    Measurement Techniques, 2017, 60 : 763 - 770