Kalman Filtering for Posture-Adaptive in-Bed Breathing Rate Monitoring Using Bed-Sheet Pressure Sensors

被引:10
|
作者
Matar, Georges [1 ,2 ]
Kaddoum, Georges [1 ]
Carrier, Julie [2 ,3 ]
Lina, Jean-Marc [1 ,2 ]
机构
[1] Ecole Technol Super, Dept Elect Engn, Montreal, PQ H3C 1K3, Canada
[2] Ctr Adv Res Sleep Med CEAMS, Montreal, PQ H4J 1C5, Canada
[3] Univ Montreal, Dept Psychol, Montreal, PQ H3T 1J4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Monitoring; Belts; Pressure sensors; Biomedical monitoring; Kalman filters; Adaptation models; Breathing rate; unobtrusive monitoring; pressure sensor mattress; respiration; breathing movements; OBSTRUCTIVE SLEEP-APNEA; APPROXIMATE ENTROPY; EPIDEMIOLOGY; SYSTEM; CONSEQUENCES; COMPLEXITY; RISK;
D O I
10.1109/JSEN.2020.3034207
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Breathing rate (BR) is one of the vital signs used in physiological monitoring. Conventional BR monitoring requires attaching wired canula/thermistor on the buco-nasal area to measure air-flow, inducing discomfort to the subject. Abdominal/thoracic belts are also used to detect breathing movements whereas esophageal pressure is the gold standard to measure breathing effort. In this paper, we aim to validate the consistency of using only bed-sheet pressure sensors to monitor the BR in healthy adults. We propose a method and demonstrate that it could be used interchangeably with respiratory belts which were approved for medical use by the American Association of Sleep Medicine (AASM). We build a ten-sinusoidal model-based extended Kalman Filter to adaptively estimate the breathing movements' signal from the body pressure distribution data. The model is posture-specific, I.e., parameters are optimized based on the detected posture. An artificial neural network (ANN) model was used to detect four bed postures to perform the kalman filter parameters' optimization step. The BRs of 12 healthy adults are recorded using the pressure mattress and a reference respiratory belt. To validate the method as a surrogate measure, a Bland-Altman (BA) analysis was performed on both pressure and belt data, and the linear relationship is evaluated using Pearson Correlation Coefficient (PCC). Interestingly, a high inter-rater agreement, an average maximum difference of 1.93 Breaths Per Minute (BrPM), a confidence interval of 95%, along with a strong linear relationship of 95.8% on average between the two methods were interestingly obtained. The presented results show the suitability of the proposed solution in medical applications requiring respiration monitoring.
引用
收藏
页码:14339 / 14351
页数:13
相关论文
共 17 条
  • [11] Garment vs. Bed-Sheet Sensors: To Deal with Pressure Dispersion Cushion Use in Pressure Ulcer Prevention
    Onose, Ryosuke
    Mase, Kenji
    Enokibori, Yu
    PROCEEDINGS OF THE 2017 ACM INTERNATIONAL JOINT CONFERENCE ON PERVASIVE AND UBIQUITOUS COMPUTING AND PROCEEDINGS OF THE 2017 ACM INTERNATIONAL SYMPOSIUM ON WEARABLE COMPUTERS (UBICOMP/ISWC '17 ADJUNCT), 2017, : 169 - 172
  • [12] Fuzzy monitoring of in-bed postural changes for the prevention of pressure ulcers using inertial sensors attached to clothing
    Bernal Monroy, Edna
    Polo Rodriguez, Aurora
    Espinilla Estevez, Macarena
    Medina Quero, Javier
    JOURNAL OF BIOMEDICAL INFORMATICS, 2020, 107
  • [13] Unconstrained Monitoring of ECG and Respiratory Variation in Infants with Underwear during Sleep Using a Bed-Sheet Electrode Unit
    Ueno, Akinori
    Yama, Yoshihiro
    2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 2329 - +
  • [14] Simultaneous Measurements of Capacitive Electrocardiogram and Ballistocardiogram Using In-Bed Fabric Sheet Electrode for Blood Pressure Estimation
    Sakajiri, Yuichiro
    Ueno, Akinori
    42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 4543 - 4546
  • [15] Sensor Pillow and Bed Sheet System: Unconstrained Monitoring of Respiration Rate and Posture Movements During Sleep
    Lokavee, Shongpun
    Puntheeranurak, Theeraporn
    Kerdcharoen, Teerakiat
    Watthanwisuth, Natthapol
    Tuantranont, Adisorn
    PROCEEDINGS 2012 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN, AND CYBERNETICS (SMC), 2012, : 1564 - 1568
  • [16] Non-Contact Capacitive Ballistocardiogram Measurements Using In-Bed Fabric-Sheet Electrode for Blood Pressure Estimation
    Sakajiri, Yuichiro
    Nakamura, Hajime
    Ueno, Akinori
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 1155 - 1158
  • [17] Bed occupancy measurements using under mattress pressure sensors for long term monitoring of community-dwelling older adults
    Taylor, Matthew
    Grant, Theresa
    Knoefel, Frank
    Goubran, Rafik
    2013 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS PROCEEDINGS (MEMEA), 2013, : 130 - 134