Usability Study of the User-Interface of Intensive Care Ventilators Based on User Test and Eye-Tracking Signals

被引:9
|
作者
Jiang, Mingyin [1 ,2 ]
Liu, Shenglin [1 ,2 ]
Feng, Qingmin [1 ,2 ]
Gao, Jiaqi [1 ,2 ]
Zhang, Qiang [1 ,2 ]
机构
[1] Huazhong Univ Sci & Technol, Dept Med Engn, Tongji Med Coll, Union Hosp, Wuhan, Hubei, Peoples R China
[2] Huazhong Univ Sci & Technol, Tongji Med Coll, Healthcare Ergon Lab, Union Hosp, Wuhan, Hubei, Peoples R China
来源
MEDICAL SCIENCE MONITOR | 2018年 / 24卷
基金
国家重点研发计划;
关键词
Critical Care; Equipment Design; Equipment Safety; Eye Movements; User-Computer Interface; Ventilators; Mechanical; CRITICAL INCIDENTS; ICU;
D O I
10.12659/MSM.909933
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: The aim of this study was to evaluate the ergonomics of the user-interface for 3 intensive care ventilators, and identify usability problems leading to user errors. Material/Methods: Sixteen respiratory therapists were recruited to perform 6 specific user-interface operational tasks on ventilators. Data (task completion time, pupil diameter, average slope of pupil diameter change, and subjective evaluation) were collected through objective measurement, questionnaires, and an eye-tracking instrument. Results: For task completion time, there were significant differences among ventilators in recognition tasks of ventilator mode and settings (P<0.05), modification of ventilator modes and recognizing (P<0.05) and changing alarm settings (P<0.05). A mean of 15 +/- 2 task failures was observed for each ventilator. For the change of pupil diameter, a significant difference was observed between ventilators (except task 2, P<0.05). For average slope of pupil diameter change, a significant difference was also observed between ventilators (except task 2, P<0.05). The Servo I showed a better correlation between task completion time and pupil diameter change. The subjective evaluation results were clear: Evital 4 received worst scores in terms of friendliness of user-interface, information display and safety (respectively, P<0.05). Conclusions: The present study provided valuable evidence to indicate the ergonomic of ventilators now used in China. With the result of this study, we can infer that the Evital 4 were poorly ergonomic designed. Furthermore, the study also demonstrated that eye-tracking can be a promising tool to evaluate the ergonomics of the user-interface.
引用
收藏
页码:6617 / 6629
页数:13
相关论文
共 50 条
  • [31] Where the User Does Look When Reading Phishing Mails - An Eye-Tracking Study
    Pfeffel, Kevin
    Ulsamer, Philipp
    Mueller, Nicholas H.
    LEARNING AND COLLABORATION TECHNOLOGIES. DESIGNING LEARNING EXPERIENCES, LCT 2019, PT I, 2019, 11590 : 277 - 287
  • [32] Usability Model Based on Traditional Metrics and Physiological and Eye Tracking Variables Depending on User Profile
    Laparra-Hernandez, Jose
    Belda-Lois, Juan-Manuel
    Page, Alvaro
    ADVANCES IN USABILITY AND USER EXPERIENCE, AHFE 2017, 2018, 607 : 291 - 302
  • [34] Entropy-based Eye-Tracking Analysis when a User Watches a PRVA's Recommendations
    Matsui, Tetsuya
    Yamada, Seiji
    2017 26TH IEEE INTERNATIONAL SYMPOSIUM ON ROBOT AND HUMAN INTERACTIVE COMMUNICATION (RO-MAN), 2017, : 23 - 28
  • [35] Design of User-Friendly Virtual Thai Keyboard Based on Eye-Tracking Controlled System
    Tantisatirapong, Suchada
    Phothisonothai, Montri
    2018 18TH INTERNATIONAL SYMPOSIUM ON COMMUNICATIONS AND INFORMATION TECHNOLOGIES (ISCIT), 2018, : 359 - 362
  • [36] The influence of snippet length on user behavior in mobile web search An experimental eye-tracking study
    Sachse, Jacqueline
    ASLIB JOURNAL OF INFORMATION MANAGEMENT, 2019, 71 (03) : 325 - 343
  • [37] The Impact of Advertising Location and User Task on the Emergence of Banner Ad Blindness: An Eye-Tracking Study
    Resnick, Marc
    Albert, William
    INTERNATIONAL JOURNAL OF HUMAN-COMPUTER INTERACTION, 2014, 30 (03) : 206 - 219
  • [38] Akori: A Tool Based in Eye-Tracking Techniques for Analyzing Web User Behaviour on a Web Site
    Vera, Felipe
    Cortes, Victor D.
    Iturra, Gabriel
    Velasquez, Juan D.
    Maldonado, Pedro
    Couve, Andres
    2017 17TH IEEE INTERNATIONAL CONFERENCE ON DATA MINING WORKSHOPS (ICDMW 2017), 2017, : 635 - 640
  • [39] Evaluation of Natural User Interface: A Usability Study Based on The Leap Motion Device
    Falcao, Christianne
    Lemos, Ana Catarina
    Soares, Marcelo
    6TH INTERNATIONAL CONFERENCE ON APPLIED HUMAN FACTORS AND ERGONOMICS (AHFE 2015) AND THE AFFILIATED CONFERENCES, AHFE 2015, 2015, 3 : 5490 - 5495
  • [40] An affective user interface based on facial expression recognition and eye-gaze tracking
    Choi, SM
    Kim, YG
    AFFECTIVE COMPUTING AND INTELLIGENT INTERACTION, PROCEEDINGS, 2005, 3784 : 907 - 914