Design, processing, and characterization of nickel titanium micro-actuators for medical implants

被引:8
|
作者
Hagemann, Ronny [1 ]
Noelke, Christian [1 ]
Rau, Thomas [2 ]
Kaierle, Stefan [1 ]
Overmeyer, Ludger [1 ]
Wesling, Volker [3 ]
Wolkers, Wim [4 ]
机构
[1] Laser Zentrum Hannover eV, D-30419 Hannover, Germany
[2] Hannover Med Sch, Inst Audioneurotechnol, D-30625 Hannover, Germany
[3] Tech Univ Clausthal, Inst Schweisstech & Trennende Fertigungsverfahren, D-38678 Clausthal Zellerfeld, Germany
[4] Leibniz Univ Hannover, Inst Mehrphasenprozesse, D-30167 Hannover, Germany
关键词
selective laser melting; shape memory alloy; medical implants; NiTi; DSC;
D O I
10.2351/1.4906381
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Cochlear implants (CI) are complex medical implants used as a common therapeutic measure for deaf people who suffer from damage to the inner ear. The success of CI insertion, a manual surgery procedure, is highly dependent on the surgeon's experience. Additionally, more precise positioning of the electrode close to the membrane structures could increase the effectiveness of frequency selectivity and stimulus conduction. To overcome these limitations, the degree of deformation of the electrode during its insertion has to be controllable. This ability can be achieved by integrating micro-actuator elements of a nickel titanium (NiTi) shape memory alloy (SMA) inside the electrode. These elements are manufactured using selective laser micromelting (SL mu M). Initially, different concepts of activation mechanisms for SMA actuators for CI electrodes are discussed. Following the rules of additive manufacturing on a microscale, the corresponding actuator design and manufacturing strategies are presented. Suitable SL mu M process parameters to achieve high spatial resolution are identified. Due to the high process temperatures, material chemical properties, respectively, its functional behavior, may be affected using SL mu M. Therefore, analyses of SL mu M NiTi parts manufactured using carrier gas hot extraction as well as differential scanning calorimetry (DSC) are carried out. Force measurements verify the available recovery forces of the produced micro-actuators activated thermally by one way effect. A suitable additive manufacturing strategy that allows the repeatable production of micro-actuators at a resolution of less than 100 mu m could be evaluated. Different anatomical geometries could be transferred from clinical data model to the manufacturing process. The processed NiTi parts meet the requirements of the ASTM F2063 concerning oxygen inclusion, which is an important condition to preserve shape memory functionality. DSC analyses reflect stable functional properties of the processed NiTi alloy independent of the adjusted laser parameters. Phase transformation of actuators could be actively proved using electrical current and passively using an external heat source. (C) 2015 Laser Institute of America.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] Processing and characterization of porous titanium for implants by using NaCl as space holder
    Torres, Y.
    Pavon, J. J.
    Rodriguez, J. A.
    JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2012, 212 (05) : 1061 - 1069
  • [32] Processing, characterization and biological testing of electrochemically modified titanium for dental implants
    Arias Suarez, Sandra Liliana
    Echeverry Rendon, Monica
    Robledo Restrepo, Sara Maria
    Arismendi Echavarria, Jorge Alberto
    Escobar Sierra, Diana Marcela
    Ossa Orozco, Claudia Patricia
    Pavon Palacio, Juan Jose
    REVISTA FACULTAD DE INGENIERIA-UNIVERSIDAD DE ANTIOQUIA, 2013, (69): : 96 - 108
  • [33] Processing and Characterization of Porous Titanium Powder Metallurgy for Implants and Joint Replacements
    Pavon, J.
    Torres, Y.
    Rodriguez, J. A.
    5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 168 - 171
  • [34] CHARACTERIZATION OF SPUTTERED NICKEL-TITANIUM SHAPE MEMORY ALLOY AND MICROFABRICATED THERMAL ACTUATORS
    Knick, Cory R.
    Srour, Merric D.
    Morris, Christopher J.
    2016 IEEE 29TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2016, : 524 - 527
  • [35] Surface modification to develop hierarchical micro/nano topography on titanium based medical implants
    Strnad, G.
    Jakab-Farkas, L.
    Cazacu, R.
    Russu, O.
    Petrovan, C.
    INNOVATIVE MANUFACTURING ENGINEERING AND ENERGY (IMANEE 2019) - 50 YEARS OF HIGHER TECHNICAL EDUCATION AT THE UNIVERSITY OF PITESTI, 2019, 564
  • [36] Anodic oxidation of titanium for implants and prosthesis: processing, characterization and potential improvement of osteointegration
    Pavon, J.
    Galvis, O.
    Echeverria, F.
    Castano, J. G.
    Echeverry, M.
    Robledo, S.
    Jimenez-Pique, E.
    Mestra, A.
    Anglada, M.
    5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 176 - +
  • [38] Characterization of bioactive ceramic coatings prepared on titanium implants by micro-arc oxidation
    Wang, YJ
    Nan, KH
    Chen, XF
    Ning, CY
    Wang, LY
    Zhao, NR
    RARE METALS, 2006, 25 (01) : 84 - 89
  • [39] Screening the optimal hierarchical micro/nano pattern design for the neck and body surface of titanium implants
    Zhou, Ping
    Mao, Feifei
    He, Fei
    Han, Yu
    Li, Hongjiao
    Chen, Jing
    Wei, Shicheng
    COLLOIDS AND SURFACES B-BIOINTERFACES, 2019, 178 : 515 - 524
  • [40] Design, Micro-Fabrication, and Characterization of a 3-DoF Micro-Conveyor Based on Digital Actuators
    Huyan, Pengfei
    Li, Pengchao
    Huang, Yulin
    Cui, Ximing
    ACTUATORS, 2022, 11 (10)