Design and development of portable fluorescence reader using Silicon photomultiplier (SiPM) sensor

被引:4
|
作者
Makkar, Roshan Lal [1 ,2 ]
Aliya, Sana Syeda [1 ]
Borse, Vivek [1 ]
Srivastava, Rohit [1 ]
机构
[1] Indian Inst Technol, Dept Biosci & Bioengn, NanoBios Lab, Bombay 400076, Maharashtra, India
[2] SAMEER, Photon Div, IIT Bombay Campus, Bombay 400076, Maharashtra, India
来源
关键词
fluorescence intensity; lateral flow assay; silicon photomultiplier; LATERAL-FLOW IMMUNOASSAY; RAPID DETECTION; ANTIGEN; ASSAY; DIAGNOSIS; SENSITIVITY; OXIDE;
D O I
10.1117/12.2305846
中图分类号
TP7 [遥感技术];
学科分类号
081102 ; 0816 ; 081602 ; 083002 ; 1404 ;
摘要
Fluorescent lateral flow assays (LFA) strips have gained popularity for medical diagnostics application by offering fast and reliable response in both qualitative & quantitative readout formats. The fluorescence emission is generated when excited with appropriate optical source, and is dependent on the analyte concentration in the sample spotted on the LFA strips. Quantitative detection requires a LFA reader to carry out accurate and precise measurements of the fluorescence emission. These readers can be of either benchtop or handheld type, and conventionally use either photo multiplier tube (PMT) or avalanche photodiodes (APD) or customized photo diodes. In addition to their proven benefits like high sensitivity, speed and gain, the availability of silicon photomultipliers (SiPM) in micro form factor makes them good choice for developing miniaturized LFA strip readers. In this study, a portable fluorescence reader using SiPM sensor has been designed, which records the fluorescence intensity of the spot on planar surfaces e.g. nitrocellulose membrane (NCM) packaged in the plastic cases. The design and operation details of the benchtop reader using SiPM sensor along with excitation source, focusing and collimating optics and power supplies integrated with general purpose microcontroller board in a mechanical housing are reported in this paper. The testing of the developed reader is done by using mercaptopropionic acid capped cadmium telluride quantum dots (MPA-CdTe QDs) as the fluorescent analytes spotted on NCM packaged plastic strips. The results obtained from the developed portable reader are compared with the standard fluorescence plate reader for QD concentration varying from 60 mu g/mL to 420 mu g/mL and are found to be in good accordance with the response & resolution of the conventional fluorescent plate reader. Further work is under development for testing the developed reader for disease diagnostic applications.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Development of a wireless radiation detection backpack using array silicon-photomultiplier (SiPM)
    Kim, Jeong Ho
    Back, Hee Kyun
    Joo, Koan Sik
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2020, 52 (02) : 456 - 460
  • [2] Feasibility of underwater radiation detector using a silicon photomultiplier (SiPM)
    Kim, Jeong Ho
    Joo, Koan Sik
    JOURNAL OF INSTRUMENTATION, 2020, 15 (04)
  • [3] Development of the first prototypes of Silicon PhotoMultiplier (SiPM) at ITC-irst
    Dinu, N.
    Battiston, R.
    Boscardin, M.
    Collazuol, G.
    Corsi, F.
    Dalla Betta, G. F.
    Del Guerra, A.
    Llosa, G.
    Ionica, M.
    Levi, G.
    Marcatili, S.
    Marzocca, C.
    Piemonte, C.
    Pignatel, G.
    Pozza, A.
    Quadrani, L.
    Sbarra, C.
    Zorzi, N.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 572 (01): : 422 - 426
  • [4] Development and Testing of Portable Fluorescence Reader (PorFloR™)
    Borse, Vivek
    Patil, Abhaysinha Satish
    Srivastava, Rohit
    2017 9TH INTERNATIONAL CONFERENCE ON COMMUNICATION SYSTEMS AND NETWORKS (COMSNETS), 2017, : 498 - 501
  • [5] Modelling a silicon photomultiplier (SiPM) as a signal source for optimum front-end design
    Corsi, F.
    Dragone, A.
    Marzocca, C.
    Del Guerra, A.
    Delizia, P.
    Dinu, N.
    Piemonte, C.
    Boscardin, M.
    Dalla Betta, G. F.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 572 (01): : 416 - 418
  • [6] DEVELOPMENT OF A PORTABLE READER FOR AN OPTICAL INTRAOCULAR PRESSURE SENSOR
    Phuong Truong
    Phan, Alex
    Williams, Nicolas
    Talke, Frank E.
    PROCEEDINGS OF THE ASME 28TH CONFERENCE ON INFORMATION STORAGE AND PROCESSING SYSTEMS, 2019, 2019,
  • [7] Design and Development of Silicon Photomultiplier for Photon Counting Applications
    Yasser, Ammar
    Gupta, Monika
    Gupta, Roshish
    Wadhwa, Manoj
    Yasser, Ammar (ammar@scl.gov.in), 1600, ISRO Satellite Centre (32): : 13 - 21
  • [8] Development of Small-Animal PET Prototype Using Silicon Photomultiplier (SiPM): Initial Results of Phantom and Animal Imaging Studies
    Kwon, Sun Il
    Lee, Jae Sung
    Yoon, Hyun Suk
    Ito, Mikiko
    Ko, Guen Bae
    Choi, Jae Yeon
    Lee, Sung-Hyuk
    Song, In Chan
    Jeong, Jae Min
    Lee, Dong Soo
    Hong, Seong Jong
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (04) : 572 - 579
  • [9] DEVELOPMENT OF A FLUORESCENCE DETECTION SYSTEM FOR HIGH-RESOLUTION MELTING ANALYSIS USING SILICON PHOTOMULTIPLIER
    Chong, K. S.
    Devi, N. A.
    Then, S. M.
    Gan, K. B.
    JOURNAL OF ENGINEERING SCIENCE AND TECHNOLOGY, 2019, 14 (02): : 951 - 964
  • [10] A Dedicated Readout ASIC for Time-of-Flight Positron Emission Tomography using Silicon Photomultiplier (SiPM)
    Chen, Huangshan
    Briggl, Konrad
    Fischer, Peter
    Gil, Alejandro
    Harion, Tobias
    Munwes, Yonathan
    Ritzert, Michael
    Schimansky, David
    Schultz-Coulon, Hans-Christian
    Shen, Wei
    Stankova, Vera
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,